|
 |
| A free-space (fully anechoic) test chamber.
|
Free-Space Radiated Emissions Certification Testing for IT Equipment
Clark Vitek
Free-space chambers may be a viable, cost-effective alternative
to open area test sites for testing tabletop information technology
equipment.
The standard for radiated immunity testing in accordance with
IEC 61000-4-3 is free space. Free space is also accepted for radiated
emissions certification testing of wireless equipment in accordance
with standards published by the European Telecommunications Standards
Institute. With free space an accepted reference environment for
many other product certifications, it may seem a bit strange that
self-certification for information technology equipment (ITE) based
on free-space emissions measurements is not already widely accepted.
Will free space soon be recognized as an acceptable environment
for international radiated emissions certifications of ITE? The
answer to this question depends largely on whether the concepts
proposed by a CENELEC working group in European standard prEN 50147
Part 31 will gain acceptance within IEC/CISPR as amendments
to either CISPR 16 or 22.
One thing is certain: changing the certification system
for ITE affects the largest number of products certified internationally,
probably more than all the other product types combined. ITE manufacturers
have a legitimate concern that any new test method must not be a
tightening of specification limits in disguise. A test method that
loosens specification limits could equally concern regulators. For
now, it seems that both regulators and industry are content with
the existing open area test site (OATS) standards. This article
examines whether any benefits would be gained by adopting an alternative
radiated emissions approval for ITE based on measurements in free
space.
Understanding the potential advantages of free space requires
an appraisal of the successes and failures of the current OATS standards.
The best argument in favor of maintaining OATS standards is that
they have been successful in controlling interference for more than
30 years. Therefore, it is widely agreed that any alternative test
methods must preserve a link in principle to this historically successful
model. In fact, some have argued that new methods based on free
space cannot be developed legitimately, simply because the specification
limits enforced by regulators such as FCC are so deeply rooted in
OATS methods.
But the arguments in favor of identifying alternatives to
OATS methods are strong too. Currently, an OATS is the theoretical
reference environment upon which manufacturers must rely to make
critical product decisions. However, such OATS are subject to a
wide range of uncontrollable weather as well as local RF ambient
conditions that vary from site to site and from day to day.
OATS testing is so fundamentally complex that some operators
believe the OATS-based test standards are not being applied uniformly
within reasonable bounds of measurement uncertainty. Consider that
proper OATS testing requires an operator to investigate and maximize
signal amplitude as a function of at least four simultaneous parameters:
frequency, turntable angle, antenna scan height, and equipment under
test (EUT) placement, including placement of interconnect cables.
Also consider that the received signal amplitude varies greatly
on an OATS as a function of some parameters even while other parameters
remain fixed. For outdoor OATS facilities, testing must take place
in the presence of a local RF ambient that can jam the frequency
spectrum with continuous transmissions from sources such as television
and radio broadcasts, as well as intermittent interference from
mobile telecommunications, airports, and military installations.
Because of these complications, manufacturers have found
that obtaining consistent problem frequencies from two different
accredited OATS laboratories is often nearly impossible. Deciding
whether a product should be modified or just retested at another
laboratory based on a report of noncompliance is a potentially costly
decision for manufacturers. Measurement uncertainty is simply too
often the most significant determinant of compliance or noncompliance
for products that are close to the CISPR 22 or FCC specification
limits.
Viewed against these conditions, free-space chambers have
a lot to offer. The technical advantage of testing indoors is straightforward:
maximized emissions can be obtained in an environment that eliminates
external ambient influence. In addition, once indoors, a typical
3-m free-space chamber can be constructed for approximately 25%
of the cost of a 10-m, full antenna scan height semianechoic chamber
(the indoor equivalent of an OATS). Free-space chambers minimize
variations caused by EUT and cabling placement and eliminate the
fundamental need for a receive antenna height scan. This effectively
reduces the required operator considerations in a free-space test
to just two major components: frequency and turntable angle.
Free Space versus OATS: Field Uniformity
The improved field uniformity yielded by the free-space
environment was a major factor in the adoption of free-space concepts
for IEC 61000-4-3 radiated immunity testing.2,3 An important
consideration is whether field uniformity is an issue that is relevant
for emissions testing. To obtain the maximized levels on an OATS,
an operator (or test software) must properly vary frequency, receive
antenna height, turntable angle, and equipment and cabling placement.
Achieving this on an OATS would require incrementally adjusting
the turntable angle and performing a full antenna height scan over
the entire frequency range for each incremental turntable angle.
Such a procedure would require a significant amount of test time.
Consequently, current industry practice is to use a fixed antenna
height for initial measurements, obtain a frequency list, and then
vary the antenna height at the frequencies obtained from the fixed-height
measurements. This method is specified in ANSI C63.4.4 Unfortunately,
this abbreviated procedure uses a fixed antenna height on an OATS,
which results in large variations in the received signal as a function
of frequency and the equipment's emissions source location within
the test volume. Figures 1 and 2 illustrate this problem for a fixed-height
3-m OATS or semianechoic chamber by analyzing the relative signal
level received from three locations (front, center, side) at three
source heights (0.4, 1, and 1.5 m), as shown in Figure 3.
 |
| Figure 1. Variation in received signal, fixed
1.5-m-height receive antenna in a semianechoic chamber or OATS
(3-m distance; vertical polarity). |
 |
| Figure 2. Variation in received signal, fixed
1.5-m-height receive antenna in a semianechoic chamber or OATS
(3-m distance; horizontal polarity). |
 |
| Figure 3. Positions included in analysis of
Figures 16. |
In many cases, varied field uniformity on an OATS invalidates
the assumption that the scan height can be fixed to obtain a preliminary
list of frequencies. This is because once an operator varies the
turntable angle or attempts to maximize the signal by moving the
EUT and cabling, the list of frequencies may change. Then, once
the antenna height is varied again, the angle at which the maximum
readings occur may change, requiring a new cable maximization and
height scan, and so on. Therefore, the theoretical quality test
that might be achieved by a full scan height OATS or semianechoic
chamber is, in fact, rarely achieved because much of the critical
information for the test--the preliminary frequency list, maximized
equipment placement and cables, and maximized turntable angle--is
usually performed with a fixed antenna height even in full scan
heightcapable facilities. This is just one plausible explanation
for the difficulty of obtaining the same frequency list from even
two accredited, high-quality OATS or semianechoic chambers.
In free space, much less variation occurs in signal amplitude
using a fixed antenna height, in either vertical or horizontal polarity.
The free-space results shown in Figure 4 are obtainable using a
fixed antenna height, yet the field uniformity is closer to that
of a full scan height OATS, represented by Figures 5 and 6. This
comparison between facility types was developed using the model
for normalized site attenuation (NSA) as contained in CISPR 22 and
ANSI C63.4. Figures 16 were experimentally verified using
a sample source radiator and were published at the 1999 IEEE International
Symposium on EMC held in Seattle in August 1999.5
 |
| Figure 4. Variation in received signal, fixed
1.5-m receive antenna height, free-space chamber (3-m test distance;
horizontal or vertical polarity). |
 |
| Figure 5. Variation in received signal, receive
antenna height varied from 1 to 4 m, semianechoic chamber or
OATS (3-m test distance; vertical polarity). |
 |
| Figure 6. Variation in received signal, receive
antenna height varied from 1 to 4 m, semianechoic chamber or
OATS (3-m test distance; horizontal polarity). |
Improved field uniformity in free space effectively eliminates
the need to scan antenna height and reduces the dependence on EUT
placement and cable maximization. Signal amplitude at each frequency
in a free-space chamber, therefore, is primarily affected by only
one variable: the turntable angle.
Free Space versus OATS: Measurement Uncertainty
Setting aside the ambient and operator issues, it is possible
to compare basic measurement uncertainty from OATS to free-space
chambers by two methods: calculation of measurement uncertainty
by a combination of Type A and Type B factors as defined by the
International Standards Organization (ISO),6 or calculation
of measurement uncertainty by sample EUT measurements. In principle,
as long as the free-space chambertoOATS uncertainty
is within the typical OATS-to-OATS uncertainty, including free-space
chambers in the list of acceptable facility types based on measurement
uncertainty should pose no additional risk.
Uncertainty Comparison Using Instrumentation Considerations.
Tables I and II compare measurement uncertainty factors contained
in an annex to prEN 50147 Part 3. The tables illustrate that, in
theory, a free-space chamber and an OATS should provide approximately
the same measurement uncertainty based on the factors that can be
quantified for the measurement instrumentation, including the facility
and the antenna.
|
Component
|
Probability
Distribution
|
Uncertainty dB
|
|
.
|
Bicon
|
LPDA
|
| Antenna factor calibration |
Normal (k = 2)
|
±1.0
|
±1.0
|
| Cable loss calibration |
Normal (k = 2)
|
±0.5
|
±0.5
|
| Receiver specification |
Rectangular
|
±1.5
|
±1.5
|
| Antenna directivity |
Rectangular
|
+0.5,0
|
+2,0
|
| Antenna factor variation with height |
Rectangular
|
±2
|
±0.5
|
| Antenna phase center variation |
Rectangular
|
+0
|
±2
|
| Antenna factor frequency interpolation |
Rectangular
|
±0.25
|
±0.25
|
| Antenna balun imbalance |
Rectangular
|
±1.0
|
±0
|
| Measurement distance error ±2 cm |
Rectangular
|
±0.1
|
±0.3
|
Height of antenna above ground plane
Height error ±2 cm |
Rectangular
|
±0.1
|
+1.0,0
|
Height of EUT above ground plane
Height error ±2 cm |
Rectangular
|
±0.05
|
±0.05
|
| Site imperfections |
Rectangular
|
±1.0
|
±1.0
|
| Mismatch |
U-shaped
|
±1.1
|
±0.5
|
| System repeatability |
Normal
|
±0.5
|
±0.5
|
| Ambient interference |
--
|
Large
|
Large
|
| Reproducibility of EUT/cable layout |
--
|
Poor
|
Poor
|
| Combined standard uncertainty uc(y) |
Normal
|
+1.92,1.90
|
+2.1,1.8
|
| Expanded uncertainty U |
Normal (k = 2)
|
±3.8
|
+4.2,3.6
|
Table I. Uncertainty budget for emission measurements on a 3-m
open area test site.
|
Component
|
Probability
Distribution
|
Uncertainty dB
|
|
.
|
Bicon
|
LPDA
|
|
Antenna factor calibration
|
Normal (k = 2)
|
±1.0
|
±1.0
|
|
Cable loss calibration
|
Normal (k = 2)
|
±0.5
|
±0.5
|
|
Receiver specification
|
Rectangular
|
±1.5
|
±1.5
|
|
Antenna directivity
|
Rectangular
|
+0
|
+0.2,0
|
|
Antenna factor variation with
height
|
Rectangular
|
+0
|
+0
|
|
Antenna phase center variation
|
Rectangular
|
+0
|
±0.5
|
|
Antenna factor frequency interpolation
|
Rectangular
|
±0.25
|
±0.25
|
|
Measurement distance error ±3
cm
|
Rectangular
|
±0.1
|
±0.1
|
|
Site imperfections
|
Rectangular
|
±3.0
|
±2.5
|
|
Mismatch
|
U-shaped
|
±1.1
|
±0.5
|
|
System repeatability
|
Normal
|
±0.25
|
±0.25
|
|
Ambient interference
|
--
|
none
|
none
|
|
Reproducibility of EUT/cable
layout
|
--
|
good
|
good
|
|
Combined standard uncertainty
uc(y)
|
Normal
|
±2.18
|
±1.86
|
|
Expanded uncertainty U
|
Normal (k = 2)
|
±4.4
|
±3.7
|
Table II. Uncertainty budget for emission measurements in a
3-m free-space chamber.
Experimental Verification of Free-Space Chamber Measurement
Uncertainty by a Simultaneous Method of Emission/Immunity Facility
Calibration. Using a NISTtraceable isotropic field probe,
it is possible to use data typically collected during an IEC 61000-4-3
uniform plane calibration in a free-space chamber to obtain both
traceable error and uncertainty terms in accordance with ISO and
NIST guidelines.6,7 This method of simultaneous, traceable
calibration of free-space facilities can be summarized as follows:
8
With an isotropic field probe, directional
coupler, and power meter, data are collected at multiple locations
in the volume to be occupied by the EUT (the uniform plane in the
case of the IEC 61000-4-3 geometry). At each calibration frequency,
the E-field is recorded as measured by the isotropic field probe.
The forward power required to generate the field is also recorded.
After the E-field and forward power data are collected, the results
are normalized to represent the required forward power to generate
1 V/m at each sample location. The average power (in watts) required
to generate 1 V/m is computed as follows:
(1)
where Pfi is the normalized forward
power required to generate 1 V/m at each of the individual sample
points, and n is the number of sample points. The standard
deviation of the required normalized forward power is then also
computed:
(2)
The system gain over isotropic is then computed from
the average value of PF and the following equation:
(3)
Note that for Equation 3, a distance must be assumed.
It is suggested that by using the specification protection-zone
distance (i.e., 3 or 10 m), the resulting average gain over isotropic
is computed directly referenced to the ideal, free-space isotropic
result at that protection zone distance, regardless of the actual
geometry of the sample-collection points. The average system transducer
factor (CdB) is calculated based on the gain by the following
equation:
(4)
This is the average transducer factor for the free-space
system (including the antenna, cabling and chamber) to be used for
future emissions measurements and is traceable to a national reference
such as NIST through the instrumentation used.
It is recommended that gain, transducer factor,
and standard deviation be computed separately for horizontal and
vertical polarity data populations if a linearly polarized antenna
is used. Although a free-space measurement system should theoretically
have no difference in characteristics depending on polarity, real
(non-ideal) measurement systems will exhibit a difference. Furthermore,
during the measurement of an actual EUT the transmit or receive
antenna may be polarized with respect to structures on the EUT,
suggesting that both polarities should be investigated during a
certification test and, therefore, should be calibrated.
The Type A (standard, k = 1) uncertainty in
the average gain and system transducer factor is calculated from
the standard deviation. This value is computed as follows:
(5)
where s is the standard deviation of the sample-population
data, and n is the number of data points included in the
computation of s. Note that for a standard 16-point IEC 1000-4-3
plane, n = 16 for each polarity.
The resulting Type A uncertainty of Equation
5 is then combined with Type B factors for the instrumentation used
during the calibration. This includes the isotropic field probe
and equipment used to monitor forward power. The Type A and Type
B factors are combined by the root-sum-squared (rss) technique and
then expanded by an expansion factor of k = 2 to represent a 95%
confidence level. This represents the 95% confidence uncertainty
in the gain and transducer values determined by the calibration.
The measurement uncertainty for a sample free-space
facility, obtained using the rss method, is shown in Figures 7 and
8. Note that this uncertainty is slightly higher for emissions than
for immunity because an emissions preamplifier was included in the
uncertainty calculations. The results shown in Figures 7 and 8 have
been obtained for multiple free-space chambers and are also consistent
with the analysis shown in Table II.
 |
| Figure 7. Measured chamber uncertainty (k =
2, Gaussian) using simultaneous emissions/immunity calibration
based on IEC 61000-4-3 uniform plane method (horizontal polarity).
|
 |
| Figure 8. Measured chamber uncertainty (k =
2, Gaussian) using simultaneous emissions/immunity calibration
based on IEC 61000-4-3 uniform plane method (vertical polarity).
|
Other methods for evaluating free-space chamber
performance based on NSA measurements are contained in prEN 50147
Part 3 and have also been discussed in previous works.9
Although NSA is a more familiar and perfectly acceptable method
of evaluating chamber performance, simultaneous calibration presents
several advantages over the NSA method. Besides the obvious economic
benefit that only a single calibration is required for both emissions
and immunity, the simultaneous method produces traceable error and
uncertainty terms, in accordance with the ISO guidelines, for the
same equipment, in the same geometry used for subsequent EUT measurements.
By contrast, NSA is only a verification test, which does not directly
produce traceable error and uncertainty terms that apply to future
testing. This is in part because the NSA verification is often performed
with different antennas and geometries than will be used for subsequent
EUT tests.
Because the simultaneous method calibrates
the free-space chamber, antenna, and feed cabling together as a
system, there is no need to calibrate these components individually.
This eliminates the need to ship equipment (primarily the antenna)
to a different facility for calibration. The uncertainty of the
isotropic field probe used for the simultaneous method is substantially
larger than the theoretical uncertainty obtainable by an insertion
loss measurement such as NSA. However, this does not affect the
resulting overall, combined expanded uncertainty that must be stated
for the equipment and facility during EUT testing. In fact, the
uncertainty is comparable by either method because even though the
NSA uncertainty may itself be quite low, that uncertainty must then
be combined and expanded for the antenna and cables actually used
during subsequent testing. The similarity of the bottom-line uncertainty
obtained by either method can be observed by comparing the results
in Table II, developed assuming NSA verification, to the results
in Figures 7 and 8, which were obtained by the simultaneous calibration
method using an isotropic field probe. The agreement in results
obtained by these substantially different methods provides a significant
validation for both the proposed simultaneous calibration and the
traditional approach currently described in prEN 50147 Part 3.
OATS Interlaboratory Comparisons
Although it has been known for some time that OATS-to-OATS
deviations can be substantial, very few studies have been conducted
until recently to evaluate OATS uncertainty based on a controlled
comparison of measurement data from several laboratories. Furthermore,
laboratory accreditation according to ISO Guide 25 was not required
in the United States for FCC measurements until 1997, so previous
studies could not compare data based on quality ensured by similar
accreditations.
A recent study, conducted by Ghery Pettit of
the Intel Corp., gathered data for a sample EUT from seven laboratories
operated by five different companies. A representative traveled
with the sample to ensure that the test setup and device operation
were consistent between laboratories. However, to better ensure
the independence of the results, the representative provided no
additional assistance (such as frequency information) to the participants.
Figure 9 shows the deviation from mean value
for the EUT sample in the frequency range of 30 MHz through 1 GHz.
The results of this study illustrate that the measurement uncertainty
determined by interlaboratory comparison measurements for an OATS
is comparable to the result predicted by considering just the instrumentation
factors (Table I). The slightly higher uncertainty in the experimental
study could be partially attributed to transportation of the equipment
samples or some variation between laboratory procedures in obtaining
maximized emissions, although the representative observed no specific
differences in test methods.
 |
| Figure 9. OATS-to-OATS correlation study using
sample equipment under test (EUT) (results of seven OATS combined).
|
It is important to note, however, that the
results of Figure 9 showing decent (± 5.3 dB, 95% confidence)
overall OATS-to-OATS correlation is based only on those frequencies
identified by two or more of the participating laboratories. The
difficulty in obtaining identical frequencies was observed in the
OATS-to-OATS study: a majority of the laboratories (4 of 7) agreed
on only 17 of the 35 frequencies reported in the frequency range
of 30 MHz through 1 GHz; less than 50% of the time. All seven laboratories
agreed on just three frequencies, which was less than 9% of the
frequencies identified by two or more laboratories.
Free-SpacetoOATS Correlation
A similar study to evaluate free-spacetoOATS
correlation was performed by CKC Laboratories Inc. using data compiled
from three sample systems tested in two free-space chambers and
on three different OATS. The combined results from this study are
shown in Figure 10. To achieve the results of Figure 10, a theoretical
correction factor based on a 1-m source height above ground was
used to correct the 3-m free-space results for comparison to the
10-m OATS readings. The Figure 10 results were obtained using a
fixed antenna height in the chamber, and a full 14-m receive
antenna scan height on the OATS. All samples included in Figure
10 were tabletop equipment.
 |
| Figure 10. Free-spacetoOATS correlation
study using sample equipment under test (EUT) data (results
combined for three EUTs and two free-space chambers). |
The results of Figure 10 for the free-spacetoOATS
correlation demonstrate excellent agreement with the theoretical
uncertainty presented in Table II and the uncertainty obtained by
the method of simultaneous calibration illustrated in Figures 7
and 8. Furthermore, the free-spacetoOATS experimental
results of Figure 10 are within the typical OATS-to-OATS uncertainty
quantified analytically in Table I and experimentally in the OATS-to-OATS
study presented in Figure 9.
Conclusions
Free-space chambers for radiated emissions
certification testing should be viewed as an opportunity for ITE
manufacturers to gain uniform worldwide access to the same low-cost,
high-quality test facilities that are already used for radiated
emissions certifications of other equipment. The adoption of free-space
alternatives for ITE would allow manufacturers to perform all of
the required radiated EMC measurements in a single facility. Because
free-space chambers eliminate interfering local RF ambient, tests
are less complex, providing an obvious advantage over OATS. Compared
to 10-m semianechoic facilities, these chambers can be built for
about 25% of the cost and yet still provide improvements in field
uniformity and test repeatability. Compared to both OATS and semianechoic
chambers, free-space chambers significantly reduce test complexity.
The use of the free-space environment substantially reduces variations
due to equipment and cable placement and effectively eliminates
the need to scan the receive antenna height to obtain maximized
emissions in the frequency range of 30 MHz through 1 GHz. This claim
is supported theoretically by the comparisons of Figures 16
and experimentally by the free-spacetoOATS results of
Figure 10 for tabletop equipment. The results of Figure 10 were
obtained using a fixed antenna height in the 3-m free-space chambers
and compared favorably to the results obtained using a full 14-m
receive antenna scan height on the OATS at a 10-m test distance.
Because OATS limits have been successful in
controlling interference, manufacturers should be allowed to continue
to certify products based on the existing OATS standards. This is
particularly applicable to large, floor-standing equipment designed
to be installed on a ground plane or computer floor. For tabletop
equipment, free-space correlation to an OATS is well within the
typical OATS-to-OATS uncertainty as demonstrated through both theoretical
and experimental considerations. The cost, convenience, and technical
advantages that free-space testing offers can provide economic benefits
without sacrificing the success that OATS have provided in controlling
interference.
References
- "prEN 50147 Part 3: Electromagnetic Compatibility Basic Emission
Standard Part 3: Emission Measurements in Fully Anechoic Rooms,"
CENELEC, TC 210, WG 4-9905, January 1999.
- Heinrich Garn, Wolfgang Mullner, and R Kresmer, "A Theoretical
and Experimental Investigation of Field Homogeneity in Absorber-Lined
Chambers for Radiated Susceptibility Tests According to IEC
801-3 second edition," IEEE 1994 International Symposium on
Electromagnetic Compatibility (Chicago: IEEE EMC Society, 1994).
- "Electromagnetic Compatibility (EMC), Part 4: Testing and
Measurement Techniques, Section 3: Radiated, Radio-Frequency,
Electromagnetic Field Immunity Test," IEC 61000-4-3, International
Electrotechnical Commission, Geneva, 1995. This standard has
also been published with some modifications as a European Norm
entitled EN 61000-4-3 (1996).
- "American National Standard for Methods of Measurement of
Radio-Noise Emissions from Low-Voltage Electrical and Electronic
Equipment in the Range of 9 kHz to 40 GHz," ANSI C63.4 (1992).
- Clark Vitek and Ed Blankenship, "An Analytical and Experimental
Comparison of Radiated Emissions Test Facilities," IEEE 1999
International Symposium on Electromagnetic Compatibility (Seattle:
IEEE EMC Society, 1999).
- Guide to the Expression of Uncertainty in Measurement,
1st ed. (Geneva: International Organization for Standardization,
1993).
- "Guidelines for Evaluating and Expressing the Uncertainty
of NIST Measurement Results," NIST Technical Note 1297, U.S.
Department of Commerce, National Institute of Standards and
Technology, 1994.
- Clark Vitek, "A Method of Simultaneous, Traceable Calibration
of Free-Space Measurement Systems for Radiated Emissions and
Radiated Immunity Testing," IEEE EMC Society Newsletter no.
181 (1999): 21.
- Roger A McConnell and Clark Vitek, "Calibration of Fully Anechoic
Rooms and Correlation with OATS Measurements," IEEE 1996 International
Symposium on EMC (Santa Clara, CA: IEEE EMC Society, 1996).
Clark Vitek is EMC staff engineer for CKC Laboratories Inc.
(Hillsboro, OR).
Back to March/April
Table of Contents
|