Characterizing Electromagnetic Fields of
Common Electronic Article Surveillance Systems
During the past decade, the U.S. Food
and Drug Administration (FDA) has received
more than 28 medical device reporting incidents
of adverse interactions between medical devices
and electronic article surveillance (EAS)
systems, metal detectors, and security systems.1
Several case reports and four peer-reviewed
studies document adverse interactions between
EAS systems and implanted pacemakers, implanted
automatic cardiac defibrillators, implanted
neurostimulators, and other ambulatory medical
devices.2,3,47 Anecdotal
reports and many newspaper articles suggest
that many more device interactions have occurred
and gone unreported.
Each year millions of people enter
establishments protected by EAS systems. Because
more people are using electronic implants
and ambulatory medical devices, adverse interactions
with EAS systems are of increasing concern.
FDA conducted a study to provide data to characterize
electromagnetic fields generated by EAS systems.
The data presented in the study are being
used for susceptibility testing of various
implanted cardiac devices and other ambulatory
medical devices to magnetic fields emitted
from EAS systems.
FDA's Center for Devices and Radiological
Health (CDRH) identified EAS systems in common
use and collected samples of the most popular
technologies used in the United States. These
included extremely low frequency and voice-frequency
(both continuous-wave magnetic), low-frequency
pulsed magnetic, and medium-frequency and
high-frequency swept radio-frequency systems.
An EAS system installer (Sentec EAS Corp.;
Deerfield Beach, FL) and an EAS system manufacturer
(Checkpoint Systems Inc.; Thorofare, NJ) loaned
sample systems to CDRH for the study. The
seven sample systems, including one duplicate,
were from three different manufacturers: Sensormatic
Electronics Corp. (Boca Raton, FL); Knogo
North America (Hauppauge, NY); and Checkpoint
Systems Inc.
Testing Instruments and Methods
Each EAS system was mounted on a simple
wooden platform. The purpose of the platform
was to fix the separation distances between
the transmitter and receiver pylons to separation
distances found in the typical installation
for each type of system. A magnetic loop antenna
connected to a Hewlett-Packard (Palo Alto,
CA) Model 8560E spectrum analyzer and a Tektronix
Inc. (Portland, OR) Model TDS 380 oscilloscope
measured the frequency of operation, modulation
type, and duty cycle of each EAS system.
The spatial magnetic flux density distributions
were mapped using a scanning system capable
of positioning a probe anywhere in a volume
measuring 2 m wide x 2 m deep x 1 m high.
The software to control the scanning system
was developed by Sonix Inc. (Springfield,
VA). Data were recorded via a Datel Inc. (Mansfield,
MA) PC414A high-speed analog input board with
12/14-bit analog-to-digital resolution. The
three-axis scanning system was bolted to the
ceiling of a shielded room. The z-axis structure
(representing up and down scanning) was constructed
of nonconducting materials that are minimally
perturbing to electromagnetic fields.
Different electromagnetic field measurement
systems were used for different frequency
ranges (see Table I). All except one were
three-axis (isotropic) probes that measured
the total magnetic field at a given point.
A single-axis Deno electric field measurements
(EFM) (West Stockbridge, MA) Model 116-3-60-0367
magnetic field probe was used for the extremely
low frequency field mapping (219535
Hz). The extremely low frequency magnetic
fields were mapped in three separate scans
for each plane. In each scan, the Deno magnetic
field probe was oriented along a different
orthogonal axis. At each point, the data were
then combined using the square root of the
sum of the squares in each of the three orthogonal
field components. Extremely low frequency
and very low frequency magnetic fields were
also measured using a Holaday Industuries
Inc. (Eden Prairie, MN) Model HI-3627 extremely
low frequency magnetic field meter (5 Hz2
kHz).
|
Instrument
|
Magnetic/Electric
Field Type
|
Frequency
Range
|
Minimum
Sensitivity
|
Deno
EFM 116-3-60-367 |
Single-axis
magnetic
|
219 Hz* and
535 Hz*
|
0.13 µT
|
|
Wandell and
Goltermann
EFA-2 field analyzer
|
Three-axis
magnetic
|
5 Hz30 kHz*
|
0.04 µT
|
|
Holaday Industries
HI-3627 three-axis
magnetic field meter
|
Three-axis
magnetic
|
5 Hz2 kHz*
|
0.02 µT
|
|
Holaday Industries
HI-3637 three-axis
magnetic field meter
|
Three-axis
magnetic
|
2400 kHz*
|
0.004 µT
|
|
Holaday Industries
HI-4433-LFH
broadband isotropic
magnetic field probe
|
Three-axis
magnetic
|
0.310.0 MHz
|
1.26 µT
|
|
Holaday Industries
HI-4433-HSE
broadband isotropic
electric field probe
|
Three-axis
electric
|
0.5 MHz1.5 GHz
|
0.3 V/m
|
| *Calibrated
using an EMCO 6402 Helmholtz coil at the
specific frequencies measured.
|
| Table I. Measurement
instruments used for EM field mapping
of EAS systems. |
The voice-frequency magnetic fields
were mapped using a Wandell and Goltermann
(Research Triangle Park, NC) Model EFA-2 (5
Hz30 kHz) field analyzer that can make
isotropic measurements. Low-frequency pulsed
magnetic fields were mapped using a Holaday
Industries Inc. Model HI-3637 very-low frequency
(2400 kHz) isotropic magnetic field
meter. The medium- and high-frequency swept
radio-frequency magnetic fields were mapped
using an HI-4433-LFH broadband (0.310
MHz) magnetic field probe. The electric fields
were mapped using an HI-4433-HSE (0.5 MHz1.5
GHz) isotropic electric field probe. Prior
to making these magnetic field measurements,
the probes were calibrated using an EMCO (Austin,
TX) Helmholtz coil, driven by a Hewlett-Packard
Model 33120A function/arbitrary waveform generator
connected to an ADCOM (East Brunswick, NJ)
GFA-555II high-current power amplifier. For
the low-frequency range (30300 kHz)
and higher-frequency ranges, the arbitrary
waveform generator was connected directly
to the Helmholtz coil. Current to the Helmholtz
coil was monitored across a precision 1% 20-W
0.866-
resistor with a Fluke (Everett, WA) Model
87 True rms multimeter, a Keithley Model 197
autoranging microvolt digital multimeter,
or a Tektronix Model TDS 380 digital real-time
oscilloscope, depending upon the frequencies
being calibrated.
A scanning protocol specified that
electromagnetic measurements would be made
in two horizontal planes and three vertical
planes around the transmitter pylon resulting
in five data sets. Measurements were made
at two heights in the horizontal plane (see
Figure 1) for each EAS system. An electromagnetic
field map was made at a height of 130 cm and
at the height of the maximum flux density
for each EAS system (as determined by a scan
of the vertical normal plane). Vertical measurements
were made in three different planes. Two vertical-plane
measures were performed, each parallel with
the face of the EAS system transmitter pylon.
The planes were located 6 and 36 cm from the
transmitter pylon face. The vertical planes
were 100 cm high starting 55 cm from the base
of the transmitter pylon. The third vertical
plane was normal to and centered on the face
of the transmitter pylon (see Figure 3). The
nearest edge of this plane was 6 cm away from
the pylon of the EAS system.
Laboratory Research Results
Figure 2 shows an example plot for
the horizontal plane 130 cm from the floor.
Table II provides a summary for the eight
systems of the maximum magnetic flux density
measured at a single point 36 cm from the
EAS system transmitter pylon face at a height
of 130 cm from the ground, along the centerline
of the pylon. The distance from the floor
roughly approximates that of an implanted
pacemaker in a standing adult. The horizontal
distance from the transmitter pylon was chosen
to minimize mutual inductance between the
magnetic field probe and the transmitter pylon.
Also, at distances greater than three probe
diameters, the error between the idealized
point measurement reported and the isotropic
volume measurement of an ideal magnetic field
probe is less than 0.04 dB (1%). This determination
was made by a simple spreadsheet model of
a field gradient comparing point measurements
to an idealized measurement probe surface.
| EAS
System |
Frequency |
Modulation |
Magnetic
Field Strength
(µT)* |
Pylon
Separation (cm) |
|
|
|
|
|
Magnetic
#1 (ELF)
|
219 Hz
|
CW
|
122
|
81.3
|
|
Magnetic
#2 (VF)
|
535.7 Hz
|
CW
|
72
|
73.7
|
|
Pulsed magnetic
#1 (LF)
|
58 kHz
|
Pulsed**
|
64.9
|
182.9
|
|
Pulsed magnetic
#2 (LF)
|
58 kHz
|
Pulsed **
|
62.2
|
182.9
|
|
Pulsed magnetic
#3 (LF)
|
58 kHz
|
Pulsed**
|
61.7
|
274.3
|
|
Swept RF
#1 (HF)
|
7.29 MHz
|
FM***
|
<1.0
|
91.4
|
|
Swept RF
#2 (HF)
|
7.68.9 MHz
|
FM ***
|
<1.0
|
91.4
|
|
Swept RF
#3 (MF)
|
1.82.1 MHz
|
FM ***
|
1.0
|
182.9
|
*36 cm from transmitter
pylon centered 130 cm from floor.
**10% duty cycle: 1.66 ms ON, 16.6 ms
interpulse interval.
***12 ms/sweep ±10% of center frequency.
ELF = extremely low frequency, VF = voice
frequency, LF = low frequency, HF = high
frequency, and MF = medium frequency.
|
| Table II.
Summary of EAS system and field strengths.
|
Analysis and Discussion
Data are presented as peak magnetic
flux density. This format was chosen so that
the magnitude of the magnetic flux density
emitted from these EAS systems could be compared.
Time-varying magnetic fields (dB/dt) induce
voltages in tissues and medical device leads
that can be readily calculated from peak magnetic
flux density, frequency, and modulation. Using
these calculations, medical device designers
can determine if interactions are likely when
the device is exposed to given magnetic flux
densities and waveforms.
Converting continuous-wave modulation
measurements from instruments reporting true
root-mean-squared (rms) values is straightforward.
Data for the pulsed EAS systems required special
analysis because of the nature of the instruments
used. The magnetic flux densities displayed
by the Holaday Industries isotropic field
instruments for extremely low frequency, voice-frequency,
very-low frequency, and low-frequency magnetic
fields are generated by true rms converter-integrated
circuits. The true rms values from the circuits
are combined from each of the three orthogonal
magnetic field sensors through vector addition
to produce the resultant magnetic field magnitude
displayed by the instrument. In our study,
we converted the rms values (Brms)
for magnetic flux densities to peak measurements
for the gated low-frequency sinusoid as shown
in the following list using the definition
of rms as given in equation (1).
(1)
Substituting the equation of a sinusoid
for f(t) yields equation (2), where
= 2
f,
f = the frequency of the sinusoid, T = the
period of the gated sinusoid, and B0
= the peak magnetic flux density.
(2)
(3)
For
>>1/T, equation (3) can be approximated
as
(4)
Peak Magnetic Flux Density. Figure 3 is a
summary of the peak magnetic flux densities
measured in the horizontal plane 130 cm from
the floor (as shown in Figure 1). The closest
measurements shown in Figure 3 are 15 cm from
the transmitter face. Data measured closer
than this distance are likely to contain large
errors due to mutual inductance between the
coils of the EAS system and the magnetic field
instrument. The EAS system technologies measured
indicate that the magnetic flux densities
are highest for the extremely low frequency,
voice-frequency, and low-frequency systems.
The radio-frequency systems (medium frequency
and high frequency) had weaker magnetic flux
densities. The extremely low frequency system
that we measured had two transmitting pylons.
In this system, the exposure to the magnetic
fields remains relatively higher anywhere
between the pylons than was measured between
the single-transmitter pylon systems. Table
II shows a numerical summary of the flux densities
measured 36 cm from each of the transmitter
pylons at a single point on the centerline
of the EAS systems.
System Installation Specifications. Each
EAS system manufacturer specifies the nominal
environmental conditions in which their units
should be installed. Because most of the systems
studied were installed by a third party and
were acquired as used systems, some of the
units may not have flux densities representative
of a system installed by the manufacturer's
representatives. However, since these EAS
systems are available on the market, the magnetic
flux densities measured would represent exposures
that may be encountered by the public. Subsequent
transmitter coil current measurements indicated
that two of the low-frequency systems (PM
#1 and PM #2) produced magnetic fields more
than 10% stronger than the manufacturer reported
as nominal at the specified supply voltage.
These units also have unregulated power supplies.
The building power supplied to these units
was not regulated during testing. The typical
supply voltage to the systems during our testing
was 126 V ac, which was 16 V ac (15%) more
than the manufacturer's designed nominal ac
voltage of 110 V ac. These two factors combine
to produce reported magnetic flux densities
that may be stronger than nominal for many
units installed by the manufacturer and operated
at the nominal power line voltages (110 V
ac).
Field Tests of EAS Systems
FDA's Winchester Engineering and Analytical
Center (WEAC) measured a number of EAS systems
in use in retail stores and libraries in and
around the greater Boston area.8
Results of this study were compared with measurements
reported in this paper for similar systems
based on the EAS system emission frequency
(Table III). WEAC measured two extremely low
frequency systems in music stores. The separation
distances between EAS system pylons for these
units were greater than the pylon separation
distances that were recommended for the extremely
low frequency EAS system installed in the
CDRH laboratory. The flux densities measured
were generally higher for the field systems
than for the unit measured in the laboratory
by 3 and 1.6 dB respectively at a horizontal
distance of 40 cm from the transmitter pylon
and 130 cm above the floor. When the maximum
values of the measurements made in the field
and in the laboratory are normalized to one
and compared, the data show differences of
1.2 and 0.4 dB respectively at
40 cm from the transmitter pylon and 130 cm
from the floor.
|
EAS
System
|
Frequency
|
Modulation
|
Flux
Density (S = 40 cm, H = 130 cm (µT)
|
Difference
(dB)
|
|
CDRH
|
WEAC
|
| Magnetic
#1 (ELF) |
219
Hz
|
CW
|
117.9
|
167.4
(H1)***
|
3.0
|
| Magnetic
#1 (ELF) |
219
Hz
|
CW
|
117.9
|
141.6(H2)***
|
1.6
|
| Magnetic
#2 (VF) |
535
Hz
|
CW
|
136
|
130.2
|
0.4
|
| Pulsed
magnetic #3 (LF) |
58
kHz
|
Pulsed
|
38.5
|
27.1
|
3.0
|
| Swept
RF #2 (HF)* |
8
kHz
|
Swept
|
0.3
|
8.5
(R1)***
|
29.9
|
| Swept
RF #2 (HF) |
8
kHz
|
Swept
|
0.3
|
1.25(R2)***
|
13.2
|
| Swept
RF #2 (HF) |
8
kHz
|
Swept
|
0.3
|
0.4
(R3)***
|
3.3
|
| Electric
Field |
V/m
|
V/m
|
|
| Swept
RF #1 (HF) |
8
kHz
|
Swept
|
0.16
|
**
|
16.0
|
| Swept
RF #1 (HF) |
8
kHz
|
Swept
|
0.16
|
0.14(R3)***
|
0.97
|
| Swept
RF #2 (HF) |
8
kHz
|
Swept
|
1.0
|
0.14(R3)***
|
17.0
|
*CDRH RF #1 magnetic flux
density could not be measured at this
distance with the HI-4433-LFH.
*Comparison between CDRH RF #1 and
CDRH RF #2.
***WEAC system identifiers for units operating
at similar frequencies. R1, R2, and R3
are library systems. H1 and
H2 are retail systems. |
| Table III. Comparison
of CDRH and WEAC data. |
Comparison of flux densities for voice-frequency
EAS systems at 40 cm from the transmitter
pylon and 130 cm from the floor showed measurement
differences of 0.2 dB. Flux densities for
low-frequency EAS systems at 40 cm from the
transmitter pylon and 130 cm from the floor
showed measurement differences of 3 dB; normalized
data show a difference of 2.1 dB. The radio-frequency
EAS systems have relatively weak magnetic
fields, and, therefore, the instruments used
by both groups were not able to measure the
magnetic fields for the laboratory EAS systems
at distances of 40 cm or greater. The radio-frequency
systems measured by WEAC were library systems.
Differences between electromagnetic
flux densities measured in the laboratory
by CDRH and field measurements by WEAC would
be expected. The make and model numbers of
some of the systems measured by WEAC were
not the same as the models measured in the
laboratory. Differences in commercial ac power
levels supplied to the systems and the presence
of conductive and magnetic materials in proximity
to the systems or the fields being measured
could change the electromagnetic field levels
and patterns generated by the systems (Table
III). Measurements made where the flux densities
are weak and where flux densities are close
to the minimum sensitivity of the measurement
instrument are subject to greater measurement
error. Under these conditions, small perturbations
will indicate larger differences in electromagnetic
field measurements. Probe positional differences
can also significantly affect the repeatability
of measurements, particularly for those made
close to the source. Here the flux densities
are changing rapidly with distance. While
efforts can be made to control these variables
in the laboratory, they are very difficult
to control during on-site measurements.
FDA studied eight EAS systems representing
seven different models from three manufacturers.
The operating frequencies of these units varied
from about 200 Hz to about 10 MHz. Spatial
maps of electromagnetic fields indicate that
lower frequency systems generally have stronger
magnetic fields than the high frequency systems.
Magnetic fields also fall off rapidly as distance
increases from the transmitting coil. Measurements
indicate that systems installed in retail
stores and libraries may have flux densities
that vary within ±3 dB of the laboratory
measurements. Magnetic fields emitted from
individual EAS units of the same model may
vary.
There have been a number of reports
of implanted and other ambulatory electronic
medical devices interacting with the electromagnetic
fields emitted from EAS systems. The information
provided in this paper may be useful in studying
such EMI and in helping engineers employ design
techniques to minimize the vulnerability of
implanted equipment.
The author would like to thank Leah
M. Shrupp, a senior in biomechanical engineering
minoring in biology and chemistry, and Michael
Cobb, a senior in biomedical/electrical engineering,
both at Marquette University (Milwaukee, WI),
for contributing to this project. They participated
via an internship arranged through a cooperative
program between Marquette University, the
Les Aspen Center (Washington, DC), and FDA.
1. MDR reports 1987 through 1997.
2. P Mathew et al., "Interaction between
Electronic Article Surveillance Systems and
Implantable Defibrillators: Insights from
a Fourth Generation ICD," PACE 20 (1997):
28572859.
3. M McIvor, "Environmental Electromagnetic
Interference from Electronic Article Surveillance
Devices: Interactions with an ICD," PACE 18
(1995): 22292230.
4. E Lucas, D Johnson, and B McElroy,
"The Effects of Electronic Article Surveillance
Systems on Permanent Cardiac Pacemakers: An
In Vitro Study," PACE 17, part 2 (1994): 20212026.
5. B Dodinot, J Godenir, and A Costa,
"Electronic Article Surveillance: A Possible
Danger for Pacemaker Patients," PACE 16, part
1 (1993): 4653.
6. A Wilke et al., "Interactions between
Pacemakers and Security Systems," Herz-schrittmacher
16 (1996): 255260.
7. D Beaugeard et al., "Interference
between Cardiac Pacemakers and Electromagnetic
Antitheft Devices," Arch. Mal Coeur 85(1992):
14571461.
8. C Harris et al., "Electromagnetic
Field Strength Levels Surrounding Electronic
Article Surveillance (EAS) Systems." Submitted
for publication.
Jon P. Casamento is working at FDA as
a senior regulatory research officer and electronics
engineer with the Center for Devices and Radiological
Health. He received his BS in electrical engineering
from the University of Maryland, College Park,
in 1979. He can reached by phone at 301/827-4959
or by e-mail at JPC@cdrh.fda.gov.