ACA
EMR Standard 2003: Compliance with Australian Regulations
Chris
Zombolas
Australia
has expanded the scope of its regulations concerning exposure to
electromagnetic radiation to capture a wider range of mobile and
portable transmitting equipment used close to the body.
The
use of mobile and portable transmitting equipment (MPTE) has grown
exponentially recently, with consumer MPTE that is operated in close
proximity to the human body proliferating. To ensure that workers
and the general public are protected from exposure to radio-frequency
(RF) electromagnetic radiation (EMR), most countries have regulations
that limit personal exposure to RF fields generated by RF transmitters
operated very near the body. Human RF-exposure guidelines and evaluation
methods for portable and mobile devices differ according to the
proximity of the device to exposed individuals.
Devices
used very close to the body are evaluated against limits for specific
absorption rate (SAR). Devices not used in such proximity may be
evaluated with respect to reference levels or maximum permissible
exposure (MPE) limits for power density.
To
market MPTE legally in Australia, suppliers must comply with the
provisions of the Australian Communications Authority (ACA) regulations
on EMR and SAR. The regulatory arrangements include RF EMR human
exposure limits, evaluation criteria, three SAR measurement methods,
and labeling requirements.
ACA
has progressively extended the scope of its EMR regulations to capture
a broader range of MPTE. It now covers most devices used at the
ear or otherwise in close proximity to the body. Since March 1,
2003, the ACA EMR testing and approval requirements have been mandated
by the Radiocommunications (Electromagnetic Radiation–Human
Exposure) Standard 2003 (known as the ACA EMR standard or EMR Standard
2003) and the Radiocommunications (Compliance Labeling–Electromagnetic
Radiation) Notice 2003 (known as the ACA EMR Labeling Notice).1,2
The
ACA EMR standard applies to all devices that operate between 100
kHz and 100 GHz and that have an integral antenna. It mandates,
for specified radiocommunications transmitters, the EMR exposure
limits and the EMR evaluation criteria of the standard published
by the Australian Radiation Protection and Nuclear Safety Agency
(ARPANSA).3 The ARPANSA standard has adopted the recommendations
of the International Commission on Non-Ionizing Radiation Protection
(ICNIRP).4 It replaces the previous interim standard AS/NZS 2772.1.5
When
a product requires evaluation against the SAR limits, the evaluation
must be performed using the guidelines and procedures prescribed
by EMR Standard 2003.
This
article details the steps necessary for MPTE to be approved to the
ACA EMR requirements for marketing in Australia. The EMR regulations
and SAR measurement methodologies also are discussed.
Exposure
Categories
Referring
to field strength, power density, and SAR requirements, the ARPANSA
exposure standards define two categories of exposure upon which
the guidelines are based. They recommend exposure limits for occupational/controlled/aware
users and for general-population/uncontrolled/unaware users. The
compliance requirements for each category are based on a person's
awareness of his or her exposure and ability to exercise control
over it.
Occupational/Controlled/Aware
Exposure. In general, occupational/controlled exposure
limits are applicable to situations in which people are exposed
to RF EMR as a consequence of their employment and where they have
been made fully aware of the potential for exposure and can exercise
control over their exposure. This category also comprises situations
when the exposure is transient, caused by incidental passage of
radiation through a location so as to create exposure levels that
may be higher than the general-population/uncontrolled limits, but
where again the exposed person is fully aware of the potential for
exposure and can exercise control over the situation by leaving
the area or by some other effective means. If appropriate, warning
signs and labels can be used to establish such awareness by providing
prominently displayed information on the risk of potential exposure
and instructions on methods to minimize exposure risks.
General-Population/Uncontrolled/Unaware
Exposure.
Limits in this category apply in situations in which the general
public may be exposed or in which people who are exposed as a consequence
of employment may not be made fully aware of the potential for exposure
or cannot exercise control over their exposure. Members of the general
public would come under this category when exposure is not employment
related, for example, in the case of a wireless transmitter that
exposes any people in its vicinity. The category also includes pregnant
workers.
Warning
labels placed on low-power consumer devices such as cellular telephones
and walkie-talkies are not considered sufficient to allow the device
to be included in the occupational/controlled category. The general-population/uncontrolled
exposure limits apply to those devices.
Exposure
Limits in Australia
The
ARPANSA standard specifies limits of human exposure to RF fields
in the frequency range of 3 kHz to 300 GHz, which are intended to
prevent adverse health effects. These limits are defined in terms
of basic restrictions on exposure of all or a part of the human
body. However, these mandatory restrictions are specified as quantities
that are impractical to measure. Therefore, reference levels–unperturbed
ambient electric and magnetic fields, induced limb currents, and
contact currents–that involve quantities that are more practical
to measure are provided as an alternative means of showing compliance
with the basic restrictions. Reference-level measurements are usually
performed with a power-density or field-strength meter. In particular,
the ARPANSA standard specifies:
- Basic restrictions
for occupational exposure (see Table I below) with corresponding
derived reference levels as a function of frequency (see
Table II).
- Basic restrictions
for general public exposure (see Table I) with corresponding derived
reference levels as a function of frequency (see
Table III).
- Equipment and usage
parameters to assist in the determination of compliance with the
standard.
 |
| Table
I. ICNIRP/ARPANSA guidelines: basic restrictions for SAR. |
Limits
for Localized Exposure or Spatial Peaks. When a transmitting
device is operated in close proximity to the body, whole-body exposure
does not occur. The RF exposure causes localized heating of the
part of the body that is very near the transmitting device. EMR
Standard 2003 increased the SAR limit for localized exposure from
1.6 W/kg (measured in a 1-g tissue mass) to 2.0 W/kg (measured in
a 10-g tissue mass). The basic restrictions for whole-body exposure
remain unchanged.Certain MPTE used under controlled circumstances
by employees of such organizations as the police department, who
have been trained to be aware of and to limit potential exposure,
is assessed under the occupational limits or the aware-user provisions
of the ARPANSA standard. Such MPTE devices are allowed exposure
limits five times higher than the MPTE intended for use by the general
public.
International
SAR Limits. As can be seen in Table
IV, it is not possible to determine worldwide SAR compliance
by testing to one limit or one standard. Manufacturers and suppliers
in Australia must ensure that MPTE meets the SAR limits by testing
to one of the ACA test methods.
The
ACA EMR Standard
The
ACA EMR standard applies to mobile stations. The term mobile
station has replaced the term mobile and portable transmitting
equipment that was used in the old standard, which has been
superseded by the 2003 standard. A mobile station is defined
as a "radiocommunications transmitter that is established for
use: (a) in motion, on land, water, or in the air; or (b) in a stationary
position at unspecified points on land, water, or in the air."
Examples given are a wireless modem operating in a laptop computer
and a handheld cellular or personal communication system (PCS) telephone
with a radiating antenna in the handpiece.
Scope.
The scope of the ACA EMR standard encompasses virtually all devices
operating within 300 kHz to 100 GHz. ACA has broadened the scope
of the SAR requirements to capture more handheld and portable devices
such as push-to-talk (PTT) radios, wireless local-area network (WLAN)
cards for laptop computers, wireless EFTPOS (electronic funds transfer
point-of-sale) terminals, and body-worn and similar devices having
integral antennas and used within 20 cm of the body. (EMR requirements
for licensed transmitters are regulated under a different legislative
instrument, the ACA Licence Condition Determination 2003.)
Reference-Level
Measurements Acceptable. Devices operating further than
20 cm from the human body do not require SAR evaluation. However,
they must comply with the ARPANSA reference levels (analogous to
the FCC MPE limits) over the modified frequency range of 0.3 MHz
to 100 GHz (see Tables II and III). Evaluation against the reference
levels is simpler and more cost-effective than SAR measurement because
a conventional power-density meter is used to perform the measurements.
SAR
Evaluation Necessary. Devices operating at frequencies
between 300 kHz and 100 GHz, transmitting an average of more than
20 mW of RF power, and used within 2.5 cm of the body require SAR
evaluation to demonstrate compliance. Devices used at distances
up to 20 cm from the human body and exceeding the power threshold
specified in Table III also require SAR evaluation.
While
ACA has harmonized the SAR limits with those of Europe for devices
used at the ear, the FCC measurement method is used for all other
devices. The ACA EMR standard contains three SAR measurement methods
- Schedule 1, known
as the old ACA SAR measurement standard.6
- EN 50361, the European
SAR measurement standard for mobile phones or devices used at
the ear.7 The limits are specified in the product standard EN
50360.8
- Schedule 2, based
on FCC SAR guidelines for devices used at the ear or within 20
cm of the human body.9,10
Schedule
1
Schedule
1 is limited to devices used at the ear and operating in the range
of 800-2500 MHz. The ARPANSA standard sets the basic restrictions
for human exposure and references the measurement and calculation
procedures of the standard AS/NZS 2772.2;11 however, actual SAR
measurement procedures are not specified in these standards.
Prior
to the publication of internationally harmonized SAR standards,
ACA published its own SAR measurement standard. This so-called old
standard was based on the work of the International Electrotechnical
Commission (IEC) and the Institute of Electrical and Electronics
Engineers (IEEE), and represented current best practice for SAR
measurement methodology. It was updated and included as Schedule
1 in the latest revision of the ACA EMR standard in order to facilitate
a smooth transition to the new standards. It expires on March 1,
2005, after which only Schedule 2 or EN 50361 may be used.
Schedule
1 specifies the SAR evaluation procedures, though not precisely.
It covers test system specifications, SAR measurement probes, test
sample positioning, test conditions, properties of phantoms, and
the assessment of measurement uncertainty and report requirements.
Compliance
Criteria. The time- and spatially averaged SAR value must
not exceed the SAR limits specified in the ARPANSA standard for
each of the test positions and possible operating conditions of
the mobile phone. The phone is tested in the touch position only,
for the right and left sides of the head. Each test position and
configuration must be tested at three frequencies, high, low, and
middle, for each band and mode, with antenna retracted and extended.
This means that a dual-band global system for mobile communications
(GSM) phone with a fixed antenna will require at least 12 SAR scans,
while a triple-band (dual-band GSM/code-division multiple-access
[CDMA]) phone with extendable antenna will require 36 scans.
The
test report must provide SAR distribution plots of localized SAR
for all test positions, conditions, and frequencies. The sample
complies if no SAR result exceeds 2 W/kg. ACA does not require an
allowance to be made for measurement uncertainty when determining
compliance with the SAR limit.
Phantom
Properties. Two types of phantom are specified in the ACA
standard, though other suitable phantoms may be used. The thickness
of the phantom shell must be less than 3 mm, and the thickness at
the ear position must be 6 mm. The more recently developed specific
anthropomorphic mannequin (SAM) phantom specified in the latest
SAR measurement standards EN 50361 and IEEE 1528 may be used.12
Tissue-Simulating
Liquid. The dielectric properties of the tissue-simulating
liquid for the homogeneous phantom head are given in Table
V. These values are significantly different from those given
in the European and FCC SAR standards. The higher values for permittivity
and lower values for conductivity will result in lower SAR values
than those that would be obtained using the dielectric values specified
by other SAR measurement methods. Also, note that the tolerance
for conductivity is 10% versus 5% for the other standards.
Measurement
Variability. As mentioned, the old ACA SAR limit of 1.6
W/kg measured in a 1-g cube of tissue has now been increased to
2 W/kg measured in a 10-g cube of tissue. Evaluation of the 10-g
SAR produces SAR values lower than those obtained with a 1-g tissue
mass. The net effect of the latest changes is that, because measured
SAR values will now be significantly lower, mobile phones that were
previously noncompliant could now be compliant.
EN
50361 SAR Measurement Method
EN
50361 is a European harmonized basic standard for SAR measurements
on mobile phones and similar devices used near the ear and operating
in the frequency range of 300 MHz to 3 GHz. In Europe, compliance
with the product standard EN 50360 may be used to demonstrate compliance
with the essential safety requirements (relevant to human exposure)
of the European Radio & Telecommunications Terminal Equipment
Directive and Council recommendation 1999/519/EC.13,14
ACA
also will accept test method EN 50361. The test report, produced
by an accredited test laboratory, must include the specific requirements
detailed in Section 8 of EN 50361. The method may not be used to
test devices not used near the ear, for example, PTT radios and
WLAN cards. (Schedule 2 must be used for such devices.)
Test
Sample Configuration. The mobile phone is tested against
the phantom in the cheek and tilted positions that are precisely
defined in the standard and identical to FCC requirements. The phone
is set for center-frequency channel at maximum transmitting power,
and the SAR is measured in both tilted and cheek positions on the
left and right sides of the head. The position of highest SAR on
each side of the head is then tested for the upper and lower frequencies
of the band. If the phone has a retractable antenna, all of these
tests are performed both with the antenna extended and retracted.
The entire process is repeated for each transmitting band and each
operating mode.
The
SAR plot for each test condition must be included in the report
along with the other report requirements specified in Section 8
of the standard.
SAM
Phantom. The specific anthropomorphic mannequin is based
on dimensions selected from a large anthropomorphic database of
male humans. The thickness of the phantom shell is specified to
be 2.0 mm ± 0.2 mm except at the ear reference point, which
shall be 6.0 mm ± 0.2 mm. The SAM material must have low-loss
and low-permittivity characteristics. The SAM phantom is also specified
by the FCC and ACA Schedule 2 measurement methods.
Tissue-Simulating
Liquid. The dielectric properties of the tissue-simulating
liquid are shown in Table V. Linear interpolation is used to obtain
dielectric properties at other, intermediate, frequencies. Examples
of recipes to obtain the stated values are given in the standard.
The deviation of the permittivity and conductivity must be within
±5% of the target values.
Measurement
System Validation. The SAR measurement system calibration
is checked in two ways: a simplified performance check is made before
each SAR measurement, and a system validation check is conducted
at least once annually.
Simplified
Performance Check. A simplified performance check is carried
out to verify that the system operates within its specifications.
It is a basic test that checks measurement repeatability to ensure
that the compliance test is performed correctly. It checks for possible
drift due to changes in liquid parameters caused by water evaporation
or temperature change, and for performance changes due to component
failures, drift, or operator errors in software parameter setup.
Power at a known level is fed into a standard dipole source antenna
positioned against a reference phantom filled with a reference liquid.
The
measured 10-g SAR must be within 10% of the actual calibrated value.
The results of the simplified performance check, including the SAR
plot, must be included in the test report.
System
Validation Test. The definition of system validation in
EN 50361 is different from the definition in IEEE 1528. EN 50361
system validation tests the SAR measurement system against the known
SAR values of selected standard reference phones. System validation
data must be gathered at least once per year.
Schedule
2
The
ACA EMR standard's Schedule 2 may be used to test for the compliance
of devices used at the ear and those used within 20 cm of any part
of the human body (body-worn devices) and that transmit in the frequency
range of 150 MHz to 5.8 GHz. This method captures virtually all
transmitting devices that were previously exempt from the regulations.
Schedule
2 was largely drawn from Appendices B, C, and D of FCC Office of
Engineering and Technology Bulletin 65C, Supplement C.10 Compliance
by means of SAR computation is excluded. An important difference
is that the ACA SAR evaluation must be performed in a 10-g tissue
mass while the FCC requirement calls for a 1-g tissue mass. Devices
covered by Schedule 2 include mobile phones, PTT transmitters, body-worn
transmitters, WLAN transmitters in laptop computers, and, generally,
all transmitting devices used within 20 cm of the human body that
transmit RF power at levels exceeding the ARPANSA threshold levels.
The evaluation criteria are displayed in Table
VI and Figures 1 and 2.
 |
| Figure
1. Evaluation criteria for products involving occupational exposure. |
The
measurement methodology is very similar to the EN 50361 method.
Schedule 2 will generally be equivalent to EN 50361 but not to Schedule
1. The test setup and configuration of mobile phones and devices
used at the ear are the same as in EN 50361; however, additional
setup information for testing body-worn devices is provided. Schedule
2 prescribes use of the SAM head phantom, along with additional
body (flat) phantoms for testing devices not used at the ear. The
head/brain dielectric properties are equivalent to those in EN 50361,
with minor deviations. The schedule gives the dielectric properties
of body- and muscle-tissue-simulating liquids as well.
Test
Sample Configuration. The cheek and tilted positions are
specified for devices used at the ear. Body-worn and PTT devices
are tested against a flat phantom. Each frequency band is tested
for the upper, middle, and lower frequencies. Devices having an
operating bandwidth of less than 10 MHz need only be tested at the
center frequency. For WLAN laptop PCs, a total of three SAR scans
is usually sufficient. The highest operating-duty factor should
be used for the SAR evaluation. The tests are repeated for each
operating mode and battery type. Battery droop or power drift must
be factored into the final SAR result. Accessories such as headsets
and microphones should be connected as in normal use and positioned
against the flat phantom.
The
setup for multimode devices is not the same as in the FCC method.
Schedule 2 describes special tests for devices where the difference
between the highest output of a low-output mode and the lowest output
of the highest-output mode is more than 2 dB.
 |
| Figure
2. Evaluation criteria for products involving nonoccupational
exposure.
|
Head
and Body Phantoms. The SAM head phantom is specified for
devices to be tested at the ear. Body-worn devices are tested against
a flat phantom made of low-loss dielectric material of 2.0-mm thickness
with a uniformity of ±0.2 mm. The length and width of the
flat phantom should be at least twice the corresponding dimensions
of the device under test, including its antenna. The depth of the
liquid should be 15.0 cm ± 0.5 cm.
Tissue
Dielectric Parameters. The head and body tissue dielectric
parameters used by Schedule 2 are shown in Table VI. These dielectric
parameters are slightly different from the EN 50361 values. Fine-tuning
of the liquid parameters will allow the same liquid to be used for
both EN 50361 and Schedule 2. The tissue dielectric values for the
center frequency of the transmission band should be within 5% of
the target value. The values in the table should be linearly interpolated
when testing at other frequencies within the band. Because it is
difficult to achieve the 5% tolerance at some frequencies, the use
of 10% tolerance is allowed as an interim measure. (EN 50361 does
not offer this relaxation.)
Measurement
System Verification. Schedule 2 system verification tests
are identical to those of the FCC method. System verification is
necessary to check and track the performance of the SAR measurement
system. A flat phantom and precision dipole-radiating source are
specified to determine whether the measurement system meets its
performance requirements. System verification checks should be performed
daily within each frequency band of the test sample. The verification
frequency must be within 100 MHz of the device transmitting frequency.
The measured 10-g SAR must be within 10% of the target values for
the specific phantom and transmitting antenna used.
Reference
Phantom for System Verification. A special reference phantom
of low dielectric loss and minimum dimensions is specified. The
reference phantom should be at least 0.75 * long and 0.5 * wide,
where * is the wavelength of the reference dipole frequency. The
measured 10-g SAR must be within ±10% of the actual target
calibrated reference value. Target SAR values are normally applicable
only to specific matched combinations of reference dipoles and reference
phantoms. A summary of these specifications is given in Schedule
2.
Measurement
Uncertainty. Schedule 2 requires the documentation of measurement
uncertainty for both SAR evaluations and SAR system verification.
Full details must be provided in the test report. The measurement
uncertainty does not have to be considered in determining compliance
with the SAR limit. However, it must not exceed 30%.
Report
Requirements. Schedule 2 lays down exhaustive requirements
regarding data to be included in the test report. Unlike electromagnetic
compatibility (EMC) regulations, the data to be provided within
the SAR report are specified, and they are required for compliance.
The ACA Labeling Notice
The
ACA EMR Labeling Notice contains administrative requirements for
manufacturers and importers of equipment covered by the ACA EMR
standard. Suppliers, whether manufacturers or importers, must label
devices to indicate that they meet the requirements of the mandated
standards. The labeling notice specifies three levels of compliance
and two device categories.
In
order to determine whether a device falls into category A or category
B, its output power level, duty factor, and separation distance
from the body must be identified and matched against the compliance
provisions of Schedule 5, Table S1, of the ARPANSA standard.
Category
A Devices. A device in this category is an aware- or unaware-user
device that does not require evaluation under Section 5.2 or Section
5.3 of Schedule 5 of the ARPANSA standard because it meets the criteria
of the test exemption column of Table S1. A category A device is
deemed by the ARPANSA standard to comply without testing. Compliance
is nevertheless mandatory, and a declaration of conformity (DoC)
must be completed.
Category
B Devices. A category B device is an aware- or unaware-user
device that is not a category A device. It does not fit within the
test exemption column of Table S1 of Schedule 5 of the ARPANSA standard.
In order to establish compliance for category B devices, evaluation
by means of power-density meter or spatial-peak SAR measurement
must be performed.
Devices
Excluded from the ACA EMR Standard. RF welders and similar
apparatus, otherwise known as industrial, scientific, and medical
(ISM) RF equipment, are not included within the scope of the ACA
EMR regulations. In Australia, human exposure to ISM RF equipment
falls under the jurisdiction of the state governments and is mandated
through National Occupational Health and Safety Commission regulations.
(ACA regulates RF interference from ISM equipment through the C-Tick
regulations.)
Also
excluded from the scope of the ACA EMR standard are RF identification
device fixed stations, field-disturbance transmitters, and similar
devices not transmitting RF for radiocommunications purposes. MPTE
used in law enforcement, by state and federal agencies, or by the
military are exempt from the ACA EMR regulations. These agencies,
however, often impose their own specific EMR SAR requirements.
Compliance
Level 1. Compliance level 1 applies to category A devices.
Examples of compliance level 1 MPTE are analog cordless phones,
remote-control toys, electronic garage- door openers, low-power
WLANs, laptop computers, and WLAN PCMCIA cards, among other devices.
Category A devices must transmit average RF power of no more than
20 mW or meet the nonevaluation criteria of Schedule 5, Table S1.
No testing is required before labeling compliance level 1 MPTE because
such equipment cannot by definition exceed the ARPANSA limits and
hence does not require evaluation. A category A device may be marketed
after a description of it has been prepared, a DoC has been completed,
and the device has been labeled with the C-Tick or A-Tick as appropriate.
It is important to note that all devices falling within the scope
of the ACA EMR standard must now be labeled with a C-Tick and that
a DoC must be completed.
Compliance
Level 2. Compliance level 2 applies to category B devices
for which the normal position of use is more than 20 cm from the
body. To satisfy compliance level 2, the supplier must comply with
the level 1 requirements and show conformity with the ACA EMR standard
through a test report of the results of reference-level measurements
conducted in accordance with AS/NZS2772.2, usually with a power-density
meter. The test report must indicate compliance with the applicable
exposure limits given as reference levels in tables 7 and 8 in Section
2.4 of the ARPANSA standard. Devices that fall within the boundaries
of compliance level 2 are, for example, WLAN access points, cordless-phone
base stations, computer WLAN cards, certain laptop computers, and
similar unlicensed transmitters in which the radiating antenna is
more than 20 cm from the body.
Compliance
Level 3. This compliance level covers category B devices
for which the normal position of use is not more than 20 cm from
the body. Compliance for a Level 3 device is established by means
of SAR evaluation. To fulfill compliance level 3, the supplier must
comply with level 1 requirements and show conformity with the ACA
EMR standard through a test report of the results of tests conducted
in accordance with the appropriate SAR measurement method, whether
Schedule 1, Schedule 2, or EN 50361. Compliance must be established
in accordance with the spatial-peak SAR limits given in Table I,
and the test report must be issued by a test laboratory specifically
accredited for SAR measurements according to Schedule 1, Schedule
2, or EN 50361. Level 3 devices fall into two groups: devices used
not more than 2.5 cm from the body and transmitting an average power
level of more than 20 mW (in cases where the user is unaware) or
more than 100 mW (for aware users), and devices used not more than
20 cm from the body and transmitting an average power level exceeding
the threshold levels of Table VI.
Assessment Test Report
The
ACA EMR Labeling Notice specifies the data to be included in the
test report. The test report must contain a description of the measurements
or evaluation methods used; results of the measurements or evaluations,
including the data; and an unequivocal statement of whether or not
the device tested complied. Compliance level 3 devices must be tested
by a laboratory that is accredited to conduct the appropriate SAR
tests. This accreditation must be granted by the National Association
of Testing Authorities (NATA) or by an international body that has
a mutual recognition agreement (MRA) with NATA. MRA partner accreditation
bodies include the (U.S.) National Voluntary Laboratory Accreditation
Program, American Association for Laboratory Accreditation, United
Kingdom Accreditation Service, German Accreditation Council, and
more than 30 other international bodies.
Compliance Recordkeeping
The
compliance records must be kept for five years from the date of
last sale of the product. They must be in English, may be kept in
electronic form, and must be kept at the principal business address
of the supplier in Australia. The compliance record must include:
-
The description of the device.
-
The test report, including an unequivocal statement of compliance
of the device and its variants.
-
Identification of variants and evidence that the variants do not
exceed the exposure levels of the device.
Device Variants
Variants
of an MPTE device may be included on the same DoC and compliance
record. Technical justification for compliance of the variant must
be made in writing, must identify the variant, and must describe
the difference between device and variant in sufficient detail to
show that the variant is essentially the same device. While cosmetic
changes involving different dielectric materials can be presumed
not to degrade EMC performance, their effect on SAR is less predictable
and not intuitively obvious. In general, any of the following may
require the SAR to be reevaluated:
-
Changes to the geometry of the device.
-
Changes to the types of material used in the construction of the
phone, keyboard, display, etc.
-
The use of different dielectric materials in the vicinity of the
antenna/transmitter circuitry, particularly when the permittivity
of the material is higher.
-
Changes in the operating frequency or increases in transmitted
power.
-
Changes in the location or alignment of the antenna.
-
The use of alternative battery types, particularly when voltage
is increased, even slightly.
-
Changes to the antenna, including the fitting of so-called radiation
reduction devices or performance enhancement devices.
This list of circumstantial changes is not comprehensive.
Product Labeling
The
labeling of compliance level 1 products is voluntary under the EMC
regulations but mandatory under the EMR labeling notice. (Compliance
in any case is mandatory, however.) All level 2 and level 3 products
must be labeled appropriately before they can be marketed. The required
label consists of either the C-Tick mark or its alternative, the
regulatory compliance mark (RCM), or else the A-Tick mark for telecommunications
devices. In New Zealand, the A-Tick mark does not signify compliance
with telecom standards as it does in Australia.
In
addition to the relevant, applicable compliance mark, each label
must also contain acceptable supplier identification. If it is not
practical to attach a label to the exterior of a device, the label
may be attached alternatively in the following order of preference:
1. To the outer surface of the packaging; or, if impractical,
2.
To instructions for use; or, if impractical,
3.
To the guarantee or certificates.
These labeling options may be used only with the written permission
of ACA or, in New Zealand, the economic development ministry's Radio
Spectrum Management Group (RSM). A valid reason must be given to
explain why it is not practical to label the product itself.
C-Tick.
The C-Tick is a certification trademark registered to ACA
under the Trade Marks Act of 1995 and to RSM under Section 47 of
the NZ Trade Marks Act. The mark must be used in accordance with
its governing regulations and is therefore issued only to an Australian
or New Zealand-based supplier. The label is expected to be durable
and limited in its size, scale, and color according to the regulations.
A-Tick
(Telecommunications Standards). All products subject to
the Australian Telecommunications Labeling Notice must also meet
the requirements of an EMC standard such as AS/NZS CISPR 22. The
A-Tick mark indicates compliance with both the EMC and telecommunications
requirements in Australia, but with only EMC requirements in New
Zealand. The requirements for conducted emissions on telecom ports
have been deferred by the revised EMC regulations until 2005 to
be consistent with the Official Journal of the European Union.
Regulatory
Compliance Mark. The RCM is a trademark owned by Australian and
New Zealand regulators and is an alternative to the C-Tick. This
mark signifies compliance with all the applicable regulatory standards,
including EMC. Suppliers intending to use the RCM should register
with Standards Australia in accordance with AS/NZS 4417.1 and complete
the application form in AS/NZS 4417.3. The RCM is not an alternative
mark to the A-Tick telecommunications compliance mark.
Supplier
Identification. In addition to one of the marks, product
labels must include the identification of the manufacturer, importer,
or agent. In Australia, one of the following must be on the label:
-
A business name and address in Australia.
-
A business name registered on the national business register.
-
A personal name and address in Australia of the place of business.
-
An Australian company number (ACN).
-
An Australian registered body number (ARBN).
-
An Australian business number (ABN).
-
An Australian registered trademark.
The supplier code number issued by ACA on application.
In
New Zealand, one of the following options must appear on the label:
-
The registered name and address of the licensee.
-
A New Zealand company number of the licensee.
-
A New Zealand registered trademark of the licensee.
-
A registered goods and services tax (GST) number.
-
The supplier code number issued by RSM on application.
Use of Marks. The C-Tick and A-Tick marks are protected
symbols to be used only with the written permission of ACA and in
accordance with conditions specified by that authority. The EMR
labeling requirements are the same as for C-Tick labeling under
the ACA EMC Framework. EMR devices connecting to the public switched
telephone network must be labeled with the A-Tick. The C-Tick or
A-Tick label, as appropriate, may be applied by the manufacturer,
importer, or authorized representative either in Australia or overseas
after obtaining written permission from ACA.
Audit Provisions
ACA
conducts audits on a random basis and also in response to complaints.
On written request from ACA, a company must produce compliance records
within 10 days. If the records do not provide sufficient evidence
that the device is compliant, the authority may request that the
supplier provide a test report from an accredited test laboratory.
When the compliance of a product is called into question, the ACA
auditor may ask the registered supplier to provide, within 10 days,
three samples of the product for testing by a NATA-accredited laboratory
nominated by ACA.
Conclusion
In
Australia, virtually all transmitting devices used in close proximity
to the human body must comply with EMR Standard 2003 and the ACA
EMR Labeling Notice. They must be evaluated against the SAR or reference-level
exposure limits of the ARPANSA standard in accordance with the ACA
SAR measurement method or by measurements conducted in accordance
with AS/NZS 2772.2. MPTE operating within 20 cm of the body and
meeting the SAR evaluation criteria must be tested for compliance
against the ARPANSA aware- or unaware-user SAR limit.
The
new regulatory arrangements allow manufacturers flexibility in choosing
a method for SAR evaluation of MPTE destined for Australian markets.
By using accredited testing, the self-declaration approval path
ensures rapid time to market. Suppliers in Australia must ensure
that evidence of fulfillment of administrative requirements and
the prescribed documentation are available for examination by an
ACA auditor.
References
1.
"Radiocommunications (Electromagnetic Radiation--Human Exposure)
Standard 2003" (ACA EMR Standard) (Melbourne: Australian Communications
Authority, March 2003).[Au: Is city of publication correct? Ditto
for Ref. 2.]
2.
"Radiocommunications (Compliance Labeling--Electromagnetic
Radiation) Notice 2003" (Melbourne: Australian Communications
Authority, March 2003).
3.
"ARPANSA Radiation Protection Standard No. 3: Maximum Exposure
Levels to Radio-Frequency Fields--3 kHz to 300 GHz" (Sydney:
Australian Radiation Protection and Nuclear Safety Agency, 2003).
4.
International Commission on Non-Ionizing Radiation Protection (ICNIRP),
"Guidelines for Limiting Exposure in Time-Varying Electric,
Magnetic, and Electromagnetic Fields (up to 300 GHz),"
Health Physics 74 (1998): 494-522.
5.
AS/NZS 2772.1 (interim), "Radio-frequency fields, part 1, Maximum
exposure levels 3 kHz to 300 GHz," (Sydney: Standards Australia,
1998).
6.
ACA EMR Standard Schedule 1. ACA test method: "Specific Absorption
Rate Test Method Using Phantom Model of Human Head (1): 2001."
7.
EN 50361:2001, "Basic standard for the measurement of specific
absorption rate related to human exposure to electromagnetic fields
from mobile phones (300 MHz to 3 GHz)" (Brussels, CENELEC,
2001).
8.
EN 50360:2001, "Product standard to demonstrate the compliance
of mobile phones with the basic restrictions related to human exposure
to electromagnetic fields (300 MHz-3 GHz)" (Brussels: CENELEC,
2001).
9.
ACA EMR Standard Schedule 2, "[SAR] Measurement method for
devices 20 cm or less from the human body; Information for documenting
compliance."
10.
FCC, Evaluating Compliance with FCC Guidelines for Human Exposure
to Radiofrequency Electromagnetic Fields, OET Bulletin 65,
ed. 97-01; Supplement C to OET Bulletin 65, ed. 01-01 (Washington,
DC: Office of Engineering and Technology, FCC, 2001).
11.
AS/NZS 2772.2:1988, "Radio-frequency radiation, Part 2: Principles
and methods of measurement–300 kHz to 100 GHz," (Sydney:
Standards Australia, 1988).
12.
IEEE 1528:2003, "Recommended Practice for Determining the Peak
Spatial-Average Specific Absorption Rate (SAR) in the Human Head
Due to Wireless Communications Devices: Measurement Techniques,"
(New York: Institute of Electrical and Electronics Engineers, 2003).
13.
"Directive 1999/5/EC of the European Parliament and of the
Council of 9 March 1999 on radio equipment and telecommunications
terminal equipment and the mutual recognition of their conformity,"
Official Journal of the European Communities no. L 91/10,
April 7, 1999.
14.
"Council recommendation 1999/519/EC of 12th July 1999: Limitation
of exposure of the general public to electromagnetic fields (0 Hz
to 300 GHz)," Official Journal of the European Communities
no. L199, July 30, 1999.
Chris Zombolas is technical director of EMC Technologies Pty
Ltd. (Melbourne, Australia). He can be reached at chris@emctech.com.au.
|