|
A
Daily Verification Method for ESD Immunity Equipment
Varuzhan
Kocharyan and Dave Tolman
Measurement
of the near electric field of the horizontal coupling plane of the
test station can turn up problems with ESD immunity equipment quickly
and economically.
The
management of electrostatic-discharge (ESD) simulators and standardized
ESD test stations to ensure sound test results continues to be a
major concern for test houses. Those facilities' quality-management
systems should prescribe a daily equipment calibration check prior
to performing tests. Early detection of malfunctioning ESD equipment
is possible if a day-by-day check is performed.
ESD
simulators and ESD test stations receive a lot of abuse. The simulators
can malfunction, and the grounding cables and bleeder resistors
can deteriorate. Annual calibration of the simulators and weekly
checks of the bleeder resistors on the test stations generally are
not sufficient. If a problem with a simulator is discovered during
an annual calibration, the immunity tests that were done during
the previous year must be investigated. This could result in costly
retesting. A similar problem associated with the bleeder resistors
could cause
inaccurate test results.
 |
| Figure
1. Measured near fields on the HCP 0.4 m from the discharge
point. |
The
standard system for verification of ESD simulators is large, expensive,
and impractical for performing an every-day check. It is a 1.5 X1.5-m
metal sheet with a specially designed target, and requires a Faraday
cage big enough to hold a 1-GHz-bandwidth oscilloscope.1 The method
for verifying ESD simulators and standardized ESD test stations
that is discussed in this article represents an attempt to find
a simpler substitute for the standard technique. This new technique
of ESD immunity equipment verification is based on the measurement
of the near electric field of the horizontal coupling plane (HCP)
of the test station. It is not intended to replace the annual calibration,
but rather to supply an effortless method suitable for day-to-day
checks in test houses.
The
HCP of the test station was chosen as the target for discharges
in order to simplify the set of devices necessary to conduct this
express method of verifying ESD equipment. For the same reason,
a method based on measurement of a discharge current was not used.
A
100- to 500-MHz-bandwidth oscilloscope can be used to measure the
quasi-electrostatic field of the HCP after the package of discharges
has been applied. The oscillograms and original tools presented
in this article are believed to support claims that this method
can assist in detection of potential problems with ESD equipment.
The HCP Near Field through the ESD Event
 |
| Figure
2. Near E-field of HCP between 100 and 700 nanoseconds. |
The
system developers, in analyzing the near field of the HCP after
applied discharges, mainly investigated the first phase of development
before the quasi-electrostatic condition was established. Figure
1 provides a typical sample of such measurement. That phase concludes
when the near magnetic field fades and the electric field of the
HCP has attained its maximum value. This takes a few tens of nanoseconds.
Figure
2 represents the electric field of the 0.7 X1.4-m coupling plane
after the +4-kV discharge has been applied. It is evident that the
quasi-electrostatic condition had been established at about 100-200
nanoseconds and during the next 0.5 microseconds had changed insignificantly.
The
process of bleeding charges from the HCP takes much longer, tens
of microseconds after the discharge had been applied. Figure 3 shows
the final stage of development of the near electric field of the
test station HCP.
Indeed,
the oscillogram in Figure 3 indicates a capacitor's discharge process.
The flat capacitor has the HCP as one lead and the ground reference
plane of the test station as a second. The originality of the equipment
verification process is that the charge bleeds not only through
the bleeder resistor, but through space also.
 |
| Figure
3. Near E-field of HCP at final phase. |
This
last stage of development of the HCP electric field takes a comparatively
long time, that is, a few tens of milliseconds. It is obvious that
the HCP discharge rate depends upon the impedance of the bleeder
paths.
The
process discussed here has two phases: the rapid phase of discharge
of the ESD simulator to the HCP, and the relatively slow phase of
discharge of the HCP to the reference ground plane. The simple physics
of this process allows making the following assumptions:
-
The discharge rate from ESD simulator to HCP depends upon the
parameters of all elements of the discharge circuit, particularly
the discharge network of the ESD simulator; thus, the rise time
seen in the oscillograms such as Figure 3 may qualify the parameters
of the simulator discharge
network.
-
The quasi-electrostatic field of the HCP carries quantitative
information about the discharge that caused the field; thus, the
peak value of such a measurement depends upon the discharge voltage
of the ESD simulator. (A low-sensitivity well-shielded transducer
is able to measure the strong near-field event while rejecting
the comparatively weaker facility-generated electromagnetic noises.)
-
The discharge rate of HCP to ground plane depends upon the impedance
of the bleeder resistor between the HCP and ground plane; thus,
the minimum value of measurements similar to those depicted in
Figure 3 is able to qualify the bleeder resistor's impedance.
An account of experimental results that prove these assumptions
follows.
 |
| Figure
4. Experimental setup for measurements of near E-field of HCP. |
Setup
and Measurement Devices
The
measurement setup used in the equipment verification method is shown
in Figure 4. The relative positions of the oscilloscope, the discharge
point, the point where bleeder resistors are connected to the HCP,
and the grounding point of the ESD simulator return cable all affect
measurement results. Any change in them may change a result. Therefore,
for the method to succeed in its purpose, it is necessary to establish
a distinct position for all of the elements involved. The method
works well with any reasonable choices of position for these measuring
system components, as long as the elements stay within the boundaries
of the ESD test stations. Once the position of the oscilloscope
and the discharge location have been selected, they should remain
unchanged.
The
oscilloscope setup is aimed mainly to capture the second phase of
the event, the slowly changing quasi-electrostatic field of the
HCP during the first 25-45 microseconds. The oscilloscope's horizontal
scale is set to 4-5 microseconds per division. With this setup,
the oscillograms carry enough information to allow detection of
changes in the discharge network of the ESD simulator, changes in
the discharge voltage, and changes in the bleeder resistor impedance.
As shown below, three values are sufficient to draw diagnostic conclusions:
the maximum voltage, the voltage at a few tens of microseconds later
(the investigators selected 28 microseconds), and the rise time.
The experiments employed the average acquisition mode, using 16
acquisitions, in order to minimize random or uncorrelated noise.
 |
| Figure
5. Shielded sensor for near E-field measurements. 1 = plastic
case, 2 = inner shield, 3 = shield opening toward measurement
area, 4 = wire antenna, and 5 = BNC connector. |
Many
publications discuss electric-field measurement and commercially
available sensors or probes. One requirement is important in this
case: the probe should be protected as well as possible against
radiation from the ESD simulator. The investigators obtained good
results using a self-manufactured sensor attached directly to the
oscilloscope with no additional cables between them. Its design
is shown in Figure 5.
For
purposes of the research, it was not necessary to convert direct
readings from the oscilloscope–volts on the input of the scope–into
the actual volts-per-meter measurement of the quasi-electrostatic
field of the HCP.
Experimental Results
Detection
of Changes in the ESD Simulator Discharge Network. An ESD
simulator discharge network contains a discharge resistance and
a capacitance. Various ESD test standards require different values
for these. Some models have variable networks that allow for adjustment
of the ESD simulator for different standards. Because there is no
immediately visible indication of which resistor-capacitor network
is installed, with some models it is necessary to disassemble the
simulator, check the discharge network, and put the simulator back
together again. This is a potential source of mistakes.
It
is also logical to assume that network parameters might change dramatically
owing to physical strain on the equipment. Experimentation showed
that the rise time provides an indication of major changes to the
network parameters. Figure 6
compares two measurement results. Each test was taken at the same
station, with the same ESD simulator and the same discharge voltage.
However, a different discharge network was used in each measurement.
Detection
of Changes in Discharge Voltage. The accuracy of the indicated
value of the discharge voltage with respect to the actual discharge
voltage is a very important calibration measurement for performing
the ESD immunity test. In the method discussed, the maximum reading
on the oscillogram is adopted as the quantity with which to appraise
the discharge voltage levels.
Figure
7 shows two typical oscillograms. Except for the discharge voltage,
all test parameters are the same for both. The simple link between
the discharge voltage and maximum voltage thus is evident.
Detection
of Malfunctioning of the Bleeder Resistor Circuit. The
value of the bleeder resistor between the HCP and the ground plane–940
k‡–is established by IEC 61000-4-2 and is one of the
parameters that ensures the proper performance of ESD test stations.1
Because of this, the bleeder resistors should be checked day by
day along with the ESD simulators.
The
principle on the basis of which this method detects any malfunction
of the bleeder resistor is this: the rate at which the quasi-electrostatic
field of the HCP decreases depends on the value of the bleeder resistor.
When the bleeder resistor circuit is open, the quasi-electrostatic
field of the HCP decreases significantly more slowly.
The
oscillograms in Figure 8 display
that difference. The point at the right of the last graticule on
the oscillogram may be chosen for this measurement. In the method
of the experimenters, the last time point is 28 microseconds. A
comparison of the voltage readings at 28 microseconds reveals any
possible malfunctioning of the bleeder resistor circuit. In the
oscillograms in Figure 8, all parameters are the same except the
value of the bleeder resistor.
When
the equipment under test (EUT) is internally powered or contains
circuitry isolated from protective ground, the applicable standard
mandates that a measure shall be taken to prevent appreciable charge
retention between individual test discharges.2 For that purpose,
test stations have attached a ground-plane wire, the bleeder wire,
for temporary grounding of the EUT through two connected-in-series
470-k‡ resistors. The bleeder wire might become a bypass circuit
for the bleeder resistor if the open end of the wire accidentally
touches the HCP. This diagnostic method will detect any reduction
in resistance to ground, which would indicate a faulty test setup.
A
pair of Excel spreadsheets were developed to implement this method.
A statistical sheet was designated for computing arithmetical means
and standard deviations of maximum voltage, voltage an interval
(28 microseconds) after maximum, and rise time.The statistical information
was to be obtained immediately after an ESD simulator was returned
from calibration (within a period of 10 days).
The
six numbers from the first sheet–three arithmetical means
and three standard deviations–have to be transferred to a
second diagnostic sheet, where they represent the baseline measurement.
The results of daily diagnostic discharge measurements then are
entered into the second sheet. If measurements are out of tolerance,
the diagnostic sheet points to the most likely source. The statistical
analysis of data obtained using this method is published.3
Conclusion
The
method discussed here exposes any changes in the indicated discharge
voltage or the HCP bleeder resistor impedance. Near-field measurement
with a duration of about 30 microseconds after discharge to the
HCP provides enough information to detect the disrepair of the discharge
network of the ESD simulator and any fault in the condition of the
bleeder resistor circuit. Consequently, that measurement has potential
for use in quickly checking the performance capability of ESD equipment.
This
verification method was developed to enable daily checking of ESD
simulators and test stations before performing ESD immunity tests.
The method is less time-consuming than standard methods for the
verification of ESD simulators and involves less-expensive apparatus.
It discovers deviations from baseline measurements that are taken
immediately after the standard calibration. EMC test houses might
use this method as a convenient and economically effective tool
for managing their ESD test equipment.
References
1.
IEC 61000-4-2, "Electromagnetic compatibility (EMC)–Part
4-2: Testing and measurement techniques–Electrostatic discharge
immunity test" (Geneva: International Electrotechnical Commission,
2001).
2.
IEC 60601-1-2, 2nd ed., "Medical electrical equipment–Part
1-2: General requirement for safety–Collateral standard: Electromagnetic
compatibility–Requirement and tests" (Geneva: International
Electrotechnical Commission, 2001).
3.
V Kocharyan and D Tolman, "An Express Diagnostic Method for
ESD Simulators and Standardized ESD Test Stations," in Proceedings
of the 2003 IEEE International EMC Symposium (Boston: Institute
of Electrical and Electronics Engineers, 2003):708-712.
Varuzhan Kocharyan is EMC engineer and Dave Tolman is quality
manager for Northwest EMC Inc. (Hillsboro, OR). They can be reached
via e-mail at vkocharyan@nwemc.com
and dtolman@nwemc.com, respectively.
|