|
TLP
ESD Testing of Electronic Components
Leo
G. Henry
A
review of the scope requirements in the recently published standard
practice document raises questions.
Since
the release of the ESD Association's Transmission Line Pulse (TLP)
ESD Standard Practice (SP) document in September 2003, the questions
being asked about TLP testing have shown a simplistic trend.1 Design,
test, and product engineers, as well as failure analysts and technicians
are asking some basic questions. One such question revolves around
the oscilloscope requirements for not only recording the waveforms
from the TLP tester (initial pulse verification), but also for recording
the electrical parameters from the device under test (DUT).
This
standard practice document (ESD SP5.5.1-2003) provides some guidance
for the scope requirements. The document, "Electrostatic Discharge
Sensitivity Testing Transmission Line Pulse (TLP) Component Level,"
defines the TLP as a rectangular current pulse formed by discharging
a transmission line. It further defines the TLP test system as one
that applies a rectangular pulse to a DUT and allows measurement
of the device's electrical characteristics during a pulsed state.
These electrical characteristics include the current, voltage, and
leakage (IVL) data from the DUT at the various pulsed voltages.2
In
an earlier article, TLP is referred to as an engineering characterization
tool or a development tool.3 It can be shown to represent or simulate
most, if not all, of the many ESD events and models like human body
model (HBM), machine model (MM), and charged device model (CDM).
TLP testing of integrated circuits (ICs) first appeared in the literature
when Maloney and Khurana presented their unique approach to testing
the ESD susceptibility of the IC core's protection structures.4
Their two-pin stress-testing technique was similar in some ways
to that of the existing test method in the HBM standard. However,
the way the HBM test was developed limited it to a pass-or-fail
test. The data obtained from TLP stress are more useful to design
and product engineers for development and design purposes. TLP uses
rectangular-pulse testing (RPT) or square-pulse testing (SPT) to
simulate the energy in an exponential HBM test pulse.
The Specifications
The
measurement requirements in the SP document are as follows: The
oscilloscope (o-scope) must have a minimum single-shot bandwidth
of 500 MHz. The current probe must have a minimum bandwidth of 1.0
GHz. The pulse width is specified as 100 nanoseconds wide, and the
TLP tester rise time range is from 0.20 to 10 nanoseconds for both
the voltage (open load) measurements and current (short load) measurements.
The document also specifies that a sampling rate of 5 GSa/sec for
the 500-MHz scope is sufficient because 100 data points exists in
a 20-second measurement window of the pulse. In this case, the document
is referring to the flattop area of the current or voltage pulses
from the DUT. This flattop area is used to collect the I-V data.
The sampling rate must be such that it captures enough data points
in the measurement window to secure accuracy. If the sampling rate
is insufficient, aliasing will occur; that is, the number of data
points sampled is too low for the waveform to be faithfully reproduced.5
Bandwidth
The
bandwidth specification is important because the scopes measure
the rise time, the voltage (and/or current), and the pulse width.
If the bandwidth of the scope is not adequate to properly reproduce
the pulse applied to the input of the scope, errors will be introduced.
These errors will occur not only in the time domain, but also in
the voltage and current measurement of the pulse height.6
 |
| Figure
1. Operating bandwidth region and -3-dB point of the o-scope. |
The
bandwidth specification tells the frequency range that the scope
is capable of measuring accurately. The bandwidth tells the frequency
at which the displayed signal reduces to 70.7% of the applied signal.
This 70.7% point is the well-known -3-dB point referenced in many
texts (see Figure 1). The amplitude of the signal is attenuated
by approximately 30% at this frequency (BW) point. As the frequency
of the input pulse increases above this frequency point, the ability
of the scope to accurately respond decreases as is shown by the
rolloff in Figure 1. It follows then that an o-scope cannot accurately
display input pulses with rise times faster than the specified rise
time of that oscilloscope.
The
Rise Time
Rise
time is a parameter that reflects a storage instrument's ability
to accurately record fast pulses. It is another way of describing
the useful frequency range of an oscilloscope. The rise time, Tr,
of the scope, if a Gaussian response is assumed, is determined as
follows:
or
So, a 500 MHz BW o-scope will have a theoretical rise time of 700
picoseconds (0.700 nanoseconds), and a 3.5 GHz BW o-scope will have
a theoretical rise time of 100 picoseconds (0.100 nanoseconds).
A simple, but well-established equation is used to define what the
o-scope actually measures, and what is observed on the screen or
window of the scope.
It
is important to note that the oscilloscope rise time includes the
rise time of the current or voltage probe used.6 The probe should
ideally be nonintrusive and should not introduce resistive, capacitive,
or inductive loading to the circuit being measured.7 This probe's
Tr must be considered because it can be a source of large error.
The general rule of thumb is that the probe's Tr must be faster
than the Tr of the o-scope.8
Error Calculations
An
example of the error associated with the o-scope measurement is
as follows. Consider the 2001 ESDA HBM standard test method (STM)
document, ESDA-HBM-STM-5.1.9 This STM document specifies that the
fastest HBM rise time to be measured from the tester is 2 nanoseconds
and the slowest is 10 nanoseconds. Using Equation 3, the measured
value is 2.83 nanoseconds.
 |
where
S is the input signal pulse and O is the o-scope. If a 175 MHz (2
nanoseconds Tr) o-scope is used to measure the 2-nanoseconds input
pulse, the result is a 41% error (0.83/2.0) in the rise time. However,
if the higher bandwidth 500 MHz (0.700 nanosecond) o-scope is used
to measure the same 2.0 nanoseconds Tr, then the measured Tr = 2.12
nanoseconds, and the error is reduced to 6%. If a 10-nanoseconds
input pulse is then measured using the same 500 MHz o-scope, the
measured Tr = 10.02 nanoseconds, and the error is further reduced
to 0.25%. The trend is obvious and expected. The slower the input
signal Tr, the lower the error in the measurement.
Table
I shows that 500 MHz scope meets the 2-10 nanoseconds rise time
specification per the 2003 ESD DSP5.5.1 document. The error in the
measurement is 6%, and it is lower for the whole range. However,
below this 2-nanoseconds minimum, the errors (22%) are huge as is
seen for the 1-nanosecond Tr. The scope's 264% error at the SP's
0.20 nanosecond Tr should not even be considered. The limit of the
scope is clearly demonstrated.
Table
I compares the errors associated with the rise times and relevant
scope used to measure the same input signal. It is clear that the
350 MHz and the 175 MHz o-scopes are not the best suited for the
2-10 Tr measurements and probably should not be used to measure
the voltage and current from the DUT. The 1.0 GHz BW o-scope is
also adequate for the 2-10 nsec range specified in the document,
but even this apparently high bandwidth o-scope has a huge measurement
error (102%) at the 0.2-nanosecond rise time.
The
last section of Table I shows
the change (decrease) in error as the rise time of the scope chosen
gets faster. It appears that the higher bandwidth 5.0 GHz o-scope
is the one ideal for measuring the rise time of the 200-picosecond
input signal. It is helpful to create a table for the range (10
to 0.20 nanoseconds) specified in the new TLP document. The acceptable
level of error is dependent on the needs of the individual.
 |
| Table
II. The errors associated with the 2-nanosecond measurement. |
Table
II shows the error associated with the measurement of the 2-nanosecond
input signal. Column 5 shows the actual error, and column 4 shows
the general rule of thumb for the ratio of the o-scope rise time
to the signal rise time. The table clearly shows that the 500 MHz
scope provides the lower limit for acceptable rise time measurement
accuracy.
Rule of Thumb Considerations
In
the literature as well as in manufacturers' (e.g., TEK, HP and LeCroy)
technical notes, the rule of thumb ratio ranges from 3:1 to 5:1;
that is, the Tr of the measuring system or scope is three to five
times faster than the Tr of the event being measured (the input
signal).10 The 5:1 rule has a maximum allowable error of 2% and
avoids most of the rolloff region shown in Figure 1. The 3:1 rule
has a maximum allowable error of 5%, but includes some of the rolloff
regions described in Figure 1.
 |
| Table
III. The bandwidth required for the 5:1 (2%) rule-of-thumb ratio. |
Another
way to present the same idea is as follows: for a 5% error, the
Tr system is <1/3 Tr of the signal being measured, and for a
2% error, the Tr of the measuring system is <1/5 Tr of the signal
being measured. Table III indicates the best (theoretically) o-scope
to use to measure each rise time in the range as specified in the
ESD SP5.1.1 document.
Table IV shows the best o-scope to use to measure each rise time
in the range of rise times as specified in the 2003 document.
The
rule of thumb changes depending on the technical note or published
article being cited. When the 5:1 ratio with the 2% error is recommended
as the minimum, the 3:1 ratio with the 5% error is the recommended
maximum error. In fact, Tables III and IV are only of major concern
if the intention is to calibrate the TLP tester rise times in addition
to verification of the SOLZ (SORZ). When possible, it is best to
leave the calibration to the manufacturer.
Furthermore,
some of the scopes mentioned in Tables III and IV do not actually
exist commercially. They are included in the tables simply to provide
the theoretical rule-of-thumb relationship. Therefore, the closest
bandwidth would have to be used (e.g., an existing 500 MHz for the
theoretical 525 MHz).
 |
| Table
IV. The bandwidth required for the 3:1 (5%) rule-of-thumb ratio. |
The
SOLZ/SORZ acronyms represent the short-open-load/short-open-resistor-zener
elements to be tested. These elements are tested so that corrections
can be identified for the errors associated with the parasitic resistances
in the TLP system or tester. 11 The zener diode is used specifically
to verify the correctness of the TLP system voltage error, and the
load-resistor is used specifically to verify the system's current
error.
Conclusion
This
article presents some guidance for selecting an appropriate o-scope
for TLP measurements. It also points out that the apparently simple
question as to which o-scope is most suitable does not have a simple
answer.12 There are three distinct TLP measurements: the measurement
of the DUT's I-V and leakage (IVL) data, the I-V data measurement
for the SOLZ/SORZ verification data, and the waveform data for the
calibration and verification of the TLP tester rise times. The new
TLP document addresses the bandwidth of the scope for IVL data collection
and for SORZ data, but it does not specifically address the rise
time measurement and the associated error issue. Fortunately, the
500 MHz bandwidth is sufficient to measure rise times down to 2
nanoseconds. However, this bandwidth is inadequate to measure any
rise time below 1 nanosecond (see Table I) because the errors at
this level are above the acceptable limits.
The
rise time is important for devices that are sensitive to ?V/?T or
?I/?T changes. The duplicating of HBM behavior requires 2-10 nanoseconds
per the TLP standard, but engineering development may actually require
<2 nanoseconds Tr.13,14 Data do exist that show changes and differences
for Tr <5 nanoseconds for both HBM testing and TLP testing.14,15
Therefore, it is important that the o-scope used to measure the
rise time parameter have sufficient bandwidth to ensure that the
errors in the measurements are at an acceptable minimum level.
References
1.
ESD SP5.5-TLP, 2002, "Standard Practice for Electrostatic Discharge
(ESD) Sensitivity Testing, Transmission Line Pulse (TLP) Testing--Component
Level," ESD Association, Rome, NY.
2.
J Barth et al.,"TLP Calibration, Correlation, Standards and
New Techniques," in Proceedings of the EOS/ESD Symposium
(Rome, NY: ESD Association, 2000), 85-96; and IEEE Transactions
on Electronics Packaging Manufacturing 24, no. 2, (2001): 99-108.
3.
LG Henry, "All Types of ESD Testing Are Not Created Equal,"
Compliance Engineering 20, no. 2 (2003): 22-27.
4.
TJ Maloney and N Khurana, "Transmission Line Pulsing Techniques
for Circuit Modeling," in Proceedings of the EOS/ESD Symposium
(Rome, NY: ESD Association, 1985), 49.
5.
Tektronix Application Note, "An Introduction to Digital Storage,"
1990.
6.
Hewlett-Packard Application Note, "BW and Sampling Rate in
Digitizing O-scopes," 1992.
7.
Tektronix Technical Brief, "The ABCs of Probe," 1998.
8.
Tektronix Application Note, "The XYZs of Oscilloscopes,"
1992.
9.
ESD STM-5.1-HBM, 1999, "Standard Test Method for Electrostatic
Discharge (ESD) Sensitivity Testing, Human Body Model–Component
Level," ESD Association, Rome, NY.
10.
Tektronix Technical Brief, "High BW Transient Capture,"
1991.
11.
Barth Electronics Inc. Application Note, "Evaluation of the
TLP System by the Use of Known SOLZ Elements to Determine the Pulse
Measurement Range, Accuracy, and Resolution."
12.
Hewlett-Packard Technical Note, "Ten Steps to Choosing the
Right Oscilloscope," 2001.
13.
S Voldman et al. (TLP-5.5 Working Group), "Standardization
of the Transmission Line Pulse (TLP) Methodology for Electrostatic
Discharge (ESD)," in Proceedings of the EOS/ESD Symposium
(Rome, NY: ESD Association, 2003), 372.
14.
M Chaine, J Davis, and A Kearney, "TLP Analysis of 0.125 µm
CMOS ESD Input Protection Circuit," in Proceedings of the
EOS/ESD Symposium (Rome, NY: ESD Association, 2003), 70.
15.
LG Henry et al., "Transmission Line Pulse ESD Testing of ICs--A
New Beginning," Compliance Engineering 18, no. 2 (2001):
46-53.
Leo G. Henry, PhD, is chief engineer for ESD-EMI-TLP Consultants
(Fremont, CA). He can be reached at 510-708-5252 or leogesd@ieee.org.
|