Leo
G. Henry
A
comparison of two HBM standards identifies the technical differences
so that a better choice can be made as to which to use for ESD
stress testing.
 |
| Leo
G. Henry |
In
an earlier article, this author made reference to the fact that
there are several standards bodies that have published electrostatic
discharge (ESD) standards for the ESD stress testing and qualification
of integrated circuits (ICs) at the component level.1
The existing ESD standards simulate the ESD events associated
with the human body model (HBM), the machine model (MM), and the
charged device model (CDM), respectively.2–4 These
standards bodies are the ESD Association (ESDA/ANSI), the Automotive
Electronics Council (AEC), JEDEC/EIA (Electronics Industries Association),
and RCJ/EIAJ (Reliability Center for Electronics Components of
Japan).2,5–7 Of course, those familiar with ESD
testing will also be aware that the older 1989 HBM military standard,
MIL-883E-Method 3015.7, is still being used today.8
Existing
ESD Standards
The
ESDA/ANSI combination simply means that the American National
Standard Institute (ANSI) approves and publishes the ESDA standards
instead of writing them itself. ESDA is an accredited ANSI standards
developer, and the ANSI designation gives the ESDA standards instant
recognition and credibility worldwide. Several questions come
to mind. Which standard should we use, and which group do we believe?
Why do we have so many bodies attempting to guide and influence
the user as to what to do to qualify a product? Why isn't there
just one standard for each ESD event type? In any standard, we
should look for at least three things: guidance, specifications,
and direction so that reproducibility and compatibility are maintained.
With these parameters, the standard provides a level playing field.
Let's consider the several existing HBM standards relative to
what they are trying to accomplish.2,5–8 To better
understand the similarities and the differences between these
documents, we will perform a Rosetta Stone analysis, which compares
the specifications and procedures in the two documents.9
Standards for MM, CDM, and transmission line pulse (TLP) will
be addressed in a future column.3,4,10
It
is not too far-fetched to say that the most important event in
ESD stress testing is that a pulse or waveform from the ESD tester
will be delivered to the device under test (DUT). This waveform
must adhere to specific parameters, and the tester must be able
to deliver this pulse consistently for the duration of the test.
Therefore the standard must specify limits for both the waveform
and the tester.
The
Rosetta Stone Analysis
The
Rosetta Stone process is used here to point to the similarities
and differences between the 2001 ESDA HBM standard and the 1989
MIL standard for HBM (see
Table I). These two standards were chosen simply because they
both represent the most recently updated and the oldest existing
HBM standards respectively. In addition, in both standards, the
equipment types (ESD tester, scope, current probe, etc.) are somewhat
similar. In some sections of the standards, however, the specifications
are quite different, which leads to different tester parameters.
This is most apparent in the observed waveforms and, subsequently,
in the different failure levels for the same devices tested on
a minimum of two different testers, each of which meets a different
standard.11,12
The
Oscilloscope
Both
standards bodies agree that an oscilloscope with a minimum 350
MHz single-shot bandwidth (BW) is the appropriate tool to use
to collect and display the waveform. This is good, but the waveform
rise time (Tr) quoted in each standard is not the same.
For the ESDA standard, the specified HBM event rise time range
is 2–10 nanoseconds. The MIL standard quotes <10 nanoseconds
(see
line 13 of Table I). According to Hewlett-Packard, the 350
MHz is equivalent to a 1.0-nanosecond rise time scope measurement
capability, where Tr (nanosecond) = 350/BW (MHz).13This
is because the scope bandwidth should always exceed (i.e., be
faster than) the bandwidth of the predominant frequencies of the
pulse or signal being measured.The measuring probe should also
always have a bandwidth greater (i.e., faster) than the scope
by a factor of 3. In either case, the higher the bandwidth of
the scope and probe, the less likely there will be error in the
measurement.13
Current
Probe and Cable Length
The
HBM MIL standard is aligned with the ESDA HBM standard with respect
to specifying 350 MHz as the minimum bandwidth capability of the
current probe (CT-1, for example). However, the MIL standard does
not specify an equivalent probe, and it does not provide a rise-time
specification or a minimum current capability if indeed it becomes
necessary to use an equivalent-type current probe. This is significant
because of the relatively fast rise time (2–10 nanoseconds)
and peak-current requirements for the waveform (pulse) in the
HBM standard. Note, however, that the difference in the cable
lengths of the current probes (see
line 7 in Table I) of 1 m maximum for the ESDA standard and
0.91 m for the MIL standard is not significant relative to the
measurement capability of the scope.
Waveforms
The general shape of the input pulse or waveform is a double exponential
(see Figure 1) with a relatively fast rise time (2 nanoseconds)
and a relatively long decay time out to 600 nanoseconds, which
is about four time constants (¥ = RC) using the general decay
equation I = Io Exp (–t/RC). I is the
final current, Io is the initial current,
t is the time, R is the resistance, and C is the capacitance.
The good news is that both standards specify the same double exponential
waveform shape.
 |
| Figure
1. Graph depicts the HBM double-exponential waveform through
a shorting wire.2 |
Rise
Time
Two
of the more important waveform parameters are the rise time, Tr,
and the peak current, Ip. The general rule
of thumb is that rise time turns on device structures and peak
current fails devices. The ESDA standard specifies a rise-time
range of 2–10 nanoseconds for all voltage levels. The MIL
standard, however, specifies simply <10 nanoseconds without
setting a lower limit. This latter condition of no minimum can
be problematic because there are published and unpublished data
that show that the failure levels for devices tested using a simulator
or tester with the lower rise-time range (2–5 nanoseconds)
are different from the failure levels for those same devices tested
on a simulator that produces the higher rise-time range (6–10
nanoseconds).14 This difference points to the need
for a thorough rise-time specification. If rise time is not specified
(as is the case with the MIL standard), then there will be an
uncertainty with respect to device correlation between the testers.
This miscorrelation was published for testers that had met the
ESDA HBM standard and were then compared with data from testers
that met the MIL standard. 11
Peak
Current and Ringing Current
For
the peak current (Ip), a short wire (zero
ohm) is used for current measurement at each voltage during the
calibration, verification, or qualification of the ESD tester.
However, the two standards use different methods for the actual
Ip measurement. The ESDA method extrapolates
the peak current from the exponential decaying portion of the
waveform. In contrast, the MIL standard uses a peak position measurement
method. Both measurement methods have drawbacks. Depending on
the level of the tester parasitics, the resulting distorted waveforms
do not always provide for an exact peak position. Further, too
much overshoot (on Ip) and too much ringing
(the higher-frequency oscillations superimposed on the normal
waveform) not only prevent the exact location of a peak position
but also cause the extrapolation to the peak current to become
a small mathematical nightmare. Both standards, however, address
the issue of excessive ringing by specifying that the ringing
current, Ir, must be <15% of Ip,
and the decrease in the ringing must be such that there is no
ringing from the 100-nanosecond position and beyond.
Decay
Time for the Zero-Ohm Test
The
decay time, Td, is another parameter that is extracted
from the waveform. The significance of the decay-time measurements
for both the zero-ohm (short wire) and the 500-W
resistor can be misleading. For the zero-ohm measurement, both
standards (see
line 15 in Table I) agree on the established one-time-constant
measurement position of 1/e, i.e., at 36.8% of Ip using
the short wire. This value is 150± 20 nanoseconds and C =
100 pF (see Figure 3). Both standards agree that the measurement
should start at Ip, which is the nonzero time position
(x axis). Both also agree that the measurement should be done
at all the stress levels for calibration and qualification of
the tester. This is one of the few occasions when both standards
are in total harmony for measurement of a significant waveform
parameter.
500-W
Resistance Measurement
During
tester calibration and qualification, it is important that the
tester provide the measured parameters that are as specified in
the standards. The 500-W-load test
is called an open load test because of its relative higher resistance
value to the overall impedances (5–15-W)
of the devices being tested. This test is called out in the ESDA
standard because it shows how the parasitics in the tester affect
the waveform produced. However, the MIL standard does not support
this test. Therefore, equipment that follows the specifications
in the MIL standard is questionable because the tester parasitics
distort the specified waveform.
To
justify the need to reduce tester parasitics, several research
groups attempted to explain the HBM tester behavior in terms of
rise time (Tr), decay time (Td),
ringing in the current (Ir), and peak current
overshoot by using a simple equivalent-lumped circuit element
(LEM) to model the elements in the HBM tester (see Figure 2).11,12,15–17
 |
| Figure
2. The parasitic capacitances in the HBM ESD tester equivalent
circuit. |
Two
published works showed that the 500-W-load
test waveform had distortions when including the three acknowledged
tester parasitics of shunt capacitor across the HBM resistor,
the series inductance of the discharge path, and the test board
capacitance (see Figure 2).15,16
A
third published work applied the complete 4th-order LEM circuit,
which specifically addresses the influence of the tester board
capacitances in addition to the socket and the socket adapter
board capacitances (see Figure 2).11 They used the
raw data from the testers (nine total) to successfully extract
the tester parameters, including the test board capacitances that
measured between 12 and 95 pF for the nine different testers.
The researchers established that the higher the tester board capacitance,
the earlier the device fails. Therefore, the tester board capacitance
must not only be specified, but it must also be limited to some
maximum value. The MIL standard does not specify this parameter;
therefore, testers that do not conform to these limits would then
be producing questionable data.
A
fourth published work studied the transient behavior of protection
devices during HBM stress and the interaction between the HBM
tester and the ESD protection devices.12 Correlation
(HBM to HBM) problems resulted because of the complex interaction
between the protection devices and the testers. They then compared
HBM and TLP data from two different testers using the same 4th-order
LEM extractions. They indicated that the TLP data provided better
correlation between the experiment and the simulation. Hence,
they concluded that the HBM failure thresholds for several devices
using several testers were highly influenced and dominated by
the tester board capacitance.
The
fifth group investigated the impact of the actual increase in
the effective tester-board capacitance in the same tester type
using two different test boards (motherboards) and the 500-W
load.17 The effect of the increase was tested because
future simulators designed to test products beyond 1024 pins will
likely have built-in increased stray tester parasitics, including
the tester board capacitances and the associated tester relay
of the switching matrix.12 Note here that the parasitics
for the socket adapter board, the shunt capacitor across the HBM
resistor, and the series inductance of the discharge path all
remained the same for this experiment. One of the motherboards
measured 14.5 nanoseconds for the 500-W rise time, which
met the older 1991 ESDA/ANSI standard specification range of 5–20
nanoseconds. The second motherboard was modified by adding capacitance
to meet the suggested increased higher-pin-count-tester specification
of 5–25 nanoseconds (see Figure 3). The measured rise time
was 24.6 nanoseconds, a 69% increase in the measured value (14.5
nanoseconds) of the first motherboard. This increase is quite
significant.
 |
| Figure
3. The addition of copper foil to motherboard; connecting
a current probe for a complex SMT socket. |
The
authors demonstrated that for the 500-W-load measurement,
the increased stray capacitance affects the rise time by slowing
it down (increased value). Hence, the slower the open-load measured
rise time, the larger the tester board capacitance. It is clear
then that because the tester board capacitance is shunted during
the short test, only the 500-W open-load test can provide
extracted value for the capacitance.
To
further emphasize the importance of tester board capacitance and
the need for the 500-W measurement, these authors focused
on the buildup of the voltage across two pins as the HBM pulse
is applied.11,12,15,18,19 They tracked the voltage
buildup simultaneously (from the same HBM pulse) across the parallel
parasitic tester-board capacitance. This tester-board capacitance
voltage is subsequently discharged into the same protection circuit.
This additional pulse from the tester board capacitance causes
more stress to the protection circuit. It has been shown by some
of these same researchers that this additional pulse can cause
added failures at lower voltages than would have occurred if the
tester-board capacitance had been lower. It follows also that
by not specifying the 500-W-load test, the data from using
such a tester is questionable, and, therefore, the standard becomes
highly suspect. The MIL standard does not specify the use of the
500-W-load test.
The
(Ip Zero Ohm)/(Ip 500-W) Current Ratio
A
consequence of the 500-W open-load test is the ratio calculation
for Ip0 and Ip500,
which is not addressed in the MIL standard but is specified in
the ESDA standard. This peak current is defined for 1000 V rather
than 2000 V, and further specifies that Ip500
must not be less than 63% of Ip0 for the
same stress level, i.e., Ip500/Ip0
> 63%.
In
work published in 1996, the authors were able to show that if
the ratio is specified and limits are set, then the limits reflect
the corresponding limits to the total system tester-board capacitance,
which should be below 46 pF.18 Noting that the MIL
standard does not address this issue, both parameters are unspecified
and can result in noncorrelatable data between testers with the
lower capacitance and those with allowable higher tester-board
capacitance.
Device
Cool-Down Time
The
two standards differ in their specification of cool-down time
between pulses. The MIL standard allows 1 second (1000 milliseconds),
whereas the ESDA standard specifies a reduced time of 300 milliseconds.
This may appear to be significant, but published data show that
there were no changes in the device behavior (i.e., failure threshold,
silicon heating, and cumulative damage) between 10-millisecond
and up to 1-second intervals between pulses.18 The
shorter time between pulses simply allows for faster throughput
in stress testing the devices.
No-Connect
(Nonwired) Pins
The
MIL standard 883E-Method 3015.7, 1989, specifically states that
the no-connect pins are not to be tested. However, the ESDA HBM
2001 standard requires that all pins on a device be tested, even
the pins listed as no-connects.
The reasoning is that some pins that are labeled no-connect could
actually be connected. Even though a simple short test could verify
this condition, there are other reasons for wanting to stress
the pin. Published data show that no-connect pins—even though
they are not stressed—could lead to arcing within the insulative
packages.20 Discharging through the package can occur
from one no-connect pin to an adjacent wired pin. Their explanation
is that the package material around the no-connect pin becomes
charged, and the sudden spark discharge currents will flow into
the adjacent wired pins, which could cause ESD failure if the
combined current is high enough. So, adjacent pins are more susceptible,
even though the authors showed that the failure mode was always
the same.
Conclusion
It was clearly shown that the Rosetta Stone analysis can provide
meaningful comparison between documents from different organizations—documents
that were developed to provide the procedure for ESD stress testing
of the same type. Even though the many differences were not discussed,
specific attention was paid to the ones that were the most significant.
These included differences such as the waveform peak current,
the rise time, the decay time, the short test, and the 500-W
open-load test. The 500-W-load resistor test was shown to
be the most significant difference. Because it directly measures
the tester board parasitic capacitance, it therefore has the largest
impact on the failure thresholds of devices being stress tested.
This
review article also clearly demonstrated that if certain parameters
are not specified in the document, then the test results could
be questionable. Although pin-combination testing is quite important,
it was not addressed here for several reasons. Currently, the
ESDA working group is addressing these issues in the HBM standard,
with the goal of upgrading the existing HBM standard.
Many
standards organizations have developed standard procedures for
testing devices to the HBM ESD event. However, a recent popular
saying applies here: "all ESD testing standards are not created
equal," and, therefore, you must have a thorough understanding
of what is in a standard before it is quoted and used for stress
testing. The ESDA HBM standard has been improved, but the HBM
MIL standard still has major deficiencies. In its 1989 revised
format, the MIL standard is not recommended for HBM sensitivity
testing to qualify product.
References
1. Leo G Henry, "All Types of ESD Testing Are Not Created
Equal—Part 1," Compliance Engineering 20, no. 2
(2003): 22–27.
2. ESD
STM-5.1-HBM, 2001, "Standard Test Method for Electrostatic Discharge
(ESD) Sensitivity Testing, Human Body Model—Component Level,"
ESD Association, Rome, NY.
3. ESD
STM-5.2-MM, 2001, "Standard Test Method for Electrostatic Discharge
(ESD) Sensitivity Testing, Machine Model—Component Level," ESD
Association, Rome, NY.
4. ESD
STM-5.3.1-CDM, 1999, "Standard Test Method for Electrostatic Discharge
(ESD) Sensitivity Testing, Charged Device Model—Component Level,"
ESD Association, Rome, NY.
5. AEC–Q101-001,
Rev-B, July 2000, "Human Body Model (HBM) Electrostatic Discharge
(ESD) Test-Discrete Component," Automotive Electronics Council,
Component Technical Committee, July 2000.
6. EIA/JESD
22-114B, "Electrostatic Discharge (ESD) Sensitivity Testing Human
Body Model (HBM)," 2000. JEDEC Solid State Technology Association,
Arlington VA.
7. J/RJC-Std.
ED-4701-1-C111A, 1994, "The HBM Sensitivity Test," Reliability
Center for Electronic Components for Japan, Tokyo.
8. MIL-883E-Method
3015.7, "Electrostatic Discharge Sensitivity Classification—The
Human Body Model, U.S. Military Standard, 1989.
9. Leo
G Henry, "Who or What Is Rosetta Stone?" ESD Threshold
9, no. 2, (2003): 9.
10.
ESDA DSP 5.5 TLP, 2003, "Draft Standard Practice for Electrostatic
Discharge (ESD) Sensitivity Testing, Transmission Line Pulse—Component
Level," ESD Association, Rome, NY.
11.
K Verhaege et al., "Analysis of HBM ESD Testers and Specifications
using the 4th Order Lumped Element Model" in Proceedings of
the EOS/ESD Symposium (Lake Buena Vista, FL: ESD Association,
1993), 129.
12.
C Russ, H Gieser, and K Verhaege, "ESD Protection Elements During
HBM Stress Tests—Further Numerical and Experimental Results,"
in Proceedings of the EOS/ESD Symposium (Las Vegas: ESD
Association, 1994), 96.
13.
Bandwidth and Sampling Rate in Digitizing Oscilloscopes,
Product Notice #54720A-1, Hewlett Packard Publication, 1992.
14.
Leo G Henry, "AMD Internal Technical Report on ESD Testing, Reliability
Engineering," AMD, Sunnyvale, CA, 1997.
15.
L van Roozandaal et al., "Standard ESD Testing of Integrated Circuits,"
in Proceedings of the EOS/ESD Symposium (Lake Buena Vista,
FL: ESD Association, 1990), 119.
16.
A Amerasekera and J Kerweij, "ESD in Integrated Circuits," Quality
and Reliability International 8, (1992): 259–272.
17.
F Khosropour, Leo G Henry, and I Morgan, "Comparison of Failure
for HBM ESD Testers Meeting ANSI/ESDA S-5.1 or the New ESDA Standard
Test Method and JEDEC Standard," in Proceedings of the 24th ISTFA
(Dallas: Electronic Device Failure Analysis Society, 1998), 337.
18.
K Verhaege et al., "Recommendations to Further Improvements of
HBM ESD Component Level Test Specifications," in Proceedings
of the EOS/ESD Symposium (Orlando, FL: ESD Association, 1996),
40.
19.
K Verhaege, "Component Level ESD Testing," Microelectronics
Reliability 38, no. 1, (1998): 115–128.
20.
M Matsumoto, M Ura, K Miyamoto, "New Failure Mechanism Due to
Non-Wired Pin ESD Stressing," in Proceedings of the EOS/ESD
Symposium (Las Vegas: ESD Association, 1994), 90.
Leo
G. Henry is chief engineer for ESD-EMI-TLP Engineering Consultants
(Fremont, CA). He can be reached at 510-708-5252 or leogesd@ieee.org.