|
|
|
|
|
|
|
Medical Electronics: Ensuring Compliance with Product
Safety Tests
Jim
Richards
Performing
the required safety tests for medical electronics is complex,
but the process can be simplified with automatic testing.
 |
|
Illustration
by TAISHA PAYTON
|
Designing
and manufacturing medical electronic equipment for compliance
with product safety standards is no easy task. The bestor
onlyway to be certain that a product is in full compliance
with its associated governing standard is for a third-party
test house or recognized test lab to perform type testing
on sample products after the design phase and before the
product is launched on the market. Once a newly designed
product has undergone proper compliance testing, the dielectric
withstand or hipot test is one of the primary safety tests
routinely performed as part of the final production process.
This
test is one that has long been required on medical electronic
products, as well as on most other electrical devices
and appliances before they exit a manufacturer's production
floor. The intention of the test is to stress a product's
insulation beyond what it would encounter in normal use.
The
end goal is to ensure that patients or caregivers never
serve as a current path to ground due to faulty insulation
or faulty grounding within the product. The test is usually
conducted by applying a test voltage that is two times
the normal operating voltage plus 1000 V (12501500
V ac, depending on whether the product is to be operated
from 115 or 240 V).
The
test voltage is applied between the shorted-together high
(hot) and neutral conductors, and the power line ground
as illustrated in Figure 1. This is a test of the basic
insulation looking back into the power cord, which is
different from the reinforced insulation that is tested
at much higher voltages during type testing. Leakage current
is monitored by a measuring instrument, commonly known
as a hipot tester, with the idea that the insulation withstand
a test voltage for a specified time without excessive
current flowing across it.
 |
|
Figure
1. Hipot test.
|
Today's
product safety testing of medical products goes far beyond
the traditional withstand test. The requirements for electrical
safety testing of medical electronic devices are much
more stringent than those for other electronic devices.
Increased precaution is needed because a patient may be
connected to several devices at the same time, contact
with the device may be directly to internal tissue, or
a patient may have reduced immune system functionality
making the person more susceptible to small leakage currents.
Many
standards are used to determine how medical products are
to be built and tested. The most widely accepted are IEC
60601-1 (the International Electrotechnical Commission's
electrical safety standard for medical electronic equipment)
and the U.S. derivation, UL 2601-1. These standards are
intended to ensure that safety considerations are taken
into account during the design phase of a product. Although
most of IEC 60601-1 and UL 2601-1 are applicable to type
testing, in the final analysis, safety testing at the
production line is the only way manufacturers can be sure
that they are shipping safe products. IEC 60601-1 has
been used as the basis for many national medical standards
worldwide.
In
addition to concerns of insulation failure, the integrity
of a product's ground connection is equally important.
The ground-bond test checks the connection from any user-exposed
or user-accessible metal parts to the ground reference
on the product's power line cord by measuring the resistance
of this connection. The object is to determine whether
sufficient current will flow to ground through this connection
rather than through the operator in the event that the
product should fail, putting the operator into contact
with live voltage.
On
electrically powered medical products, the test is performed
with the application of 25 A of current or 1.5 times the
rated current of the product (whichever is greater). This
test checks the integrity of the connectionnot just
the presence of the connection, as is done for many other
types of products in which a low-current continuity test
is adequate. This bond test is illustrated in Figure 2.
 |
|
Figure
2. Ground-bond test.
|
Leakage
Tests
Leakage
tests are performed under quite different conditions than
the hipot or ground-bond tests. For leakage tests, the
product is actually powered up under operating conditions.
The leakage test is the most-often misunderstood safety
test, primarily because it is not a common production
test for most electrical products, with the exception
of medical products.
There
are several ways to perform leakage tests, with the basic
difference simply being how or where a person comes into
contact with a product. Leakage current is the
current that could flow from the point where a person
makes contact with a product, through that person's body,
and back to ground (or some other point). Acceptable levels
of leakage current are generally outlined in the appropriate
governing standard for that product. This article examines
four tests commonly performed as a final production test
on medical products. Leakage current is a general
term, which can be further defined by the more specific
current types listed below.
-
Earth
leakage current: The current flowing from the product
mains power supply, across the product's insulation,
and back through the ground conductor in the line cord.
-
Enclosure
leakage current: The current flowing from any part
of the product enclosure back to ground (also referred
to as touch/chassis leakage).
-
Patient
leakage current: The current flowing from an applied
part back to ground (the applied part being the part
of a medical product that in normal use comes in contact
with the patient).
-
Patient
auxiliary current: The unwanted parasitic current
flowing between two separate patient circuits or connections.
Each
type of leakage-current test is performed with a measuring
device specified by various product standards to simulate
the impedance of the human body. Conducting the leakage
tests in this fashion provides a good indication of how
much leakage current a person could be exposed to in real-world
conditions.
Production
leakage-current tests must be done with the product at
110% of the highest-rated supply voltage and under a variety
of normal and single-fault conditions. Normal conditions
are electrical conditions that might occur daily and thus
are not considered to be a problem. Single faults are
essentially problem conditions that could occur. Because
it is unlikely that two faults could occur at the same
time, a product does not need to be tested with two or
more faults. Examples of normal and single-fault conditions
that are often used in a production-test environment include
normal power applied (high and neutral), reverse power
applied (high and neutral reversed), single fault/normal
(neutral open), and single fault/reverse (neutral open,
high and neutral reversed). Note that a reversed ac line
is considered to be a normal condition, not a fault condition.
Earth
Leakage. The earth line leakage test is conducted
with a tester circuit similar to that shown in Figure
3, where MD represents the measuring device. This earth
leakage is essentially a sum of all leakages in the product
under test, or basically the current flowing back to earth
ground through the ground conductor of the line cord.
This test is only applicable to protective earth products
(Class I) that have a three-prong power cord, which applies
to most electronic medical products. Class II products,
for which other leakage tests are applicable, are those
products that employ a two-prong power cord and typically
include reinforced insulation.
 |
|
Figure
3. Earth leakage test (from IEC 60601-1).
|
For
this test, as well as other leakage tests, an isolation
transformer (T1) is recommended
to represent worst-case conditions in which the secondary
circuit is connected to ground reference of the measuring
circuit (r). This test is made under normal conditions
(S1 closed, S5
to normal), reversed line (S5 reverse,
S1 closed), single-fault normal
(S1 open, S5
to normal), and single-fault reverse (S1
open, S5 to reverse). The allowable
values of leakage current specified in IEC 60601-1 at
normal and at single-fault conditions are listed in Table
I for this and other leakage tests.1 Always
refer to the standard pertinent to the product being tested.
| Current
|
Normal
|
Single
Fault
|
| Earth
leakage |
5
mA |
10
mA
|
| Touch/chassis
leakage |
100
µA |
500
µA
|
| Patient
leakage |
100
µA |
500
µA
|
| Patient
auxiliary leakage |
100
µA |
500
µA
|
|
|
Table
I. Leakage current limits (from IEC 60601-1).
|
Enclosure
Leakage. Enclosure leakage (or touch/chassis leakage)
is essentially the leakage to ground that a person would
be subjected to if he or she were to touch non-earth-protected
exposed parts of the device. This test is shown in Figure
4, where MD1 represents the measuring device. Exposed
parts could be connectors, knobs, screws, or any exterior
metal parts that are not at earth ground.
 |
|
Figure
4. Enclosure leakage test (from IEC 60601-1).
|
When
an enclosure is made completely of insulating material,
a piece of conductive foil should be placed in contact
with the enclosure surface. This test is also done under
the same four conditions discussed in earth leakage measurements:
normal, reverse, single-fault normal, and single-fault
reverse. In addition, all four tests are often performed
with an open ground-fault condition (S7
open). The allowable values of enclosure leakage current
are listed in Table I. Also note in Figure 4 that leakage
tests are sometimes made with an external voltage on a
signal input or output connection to the product under
test. For safety reasons, a series limiting resistor is
placed in the connection to prevent current in excess
of 5 mA.
Patient
Leakage. The patient leakage current test measures
the current from parts normally in contact with the patient
(applied parts) to earth as shown in Figure 5, where MD
represents the measuring device. This test can be one
of the most critical tests for medical instrumentation
because patient leads are often in direct contact with
the skin of the patient and sometimes under the skin,
where body resistance is the lowest.
 |
|
Figure
5. Patient leakage test (from IEC 60601-1).
|
Testing
of applied parts can be a bit more complex. Medical standards
usually specify that the leakage current be less than
other tests because a device may have several patient
connections. This test measures the leakage current patients
would be subjected to if they were in contact with a patient
lead and happened to make contact with a grounded object.
Like the previous tests, this one is also done under the
same four conditions: normal, reverse, single-fault normal,
single-fault
reverse (with or without an open-ground connection). Allowable
values of patient leakage current are listed in Table
I, but the table does not reflect the lower values called
for on some types of applied parts.
Patient
Auxiliary Leakage. The patient auxiliary current is
the current that flows between one patient connection
and another as shown in Figure 6, where MD represents
the measuring device. This test measures the leakage current
that a patient would be subjected to between two patient
leads that are in contact with the patient. This test
is also done under the same four conditions, with or without
an open ground connection. Typical limits for this test
are also listed in Table I.
 |
|
Figure
6. Patient auxiliary leakage test (from IEC 60601-1).
|
The
Testing Process
All
of this testing can be complex, especially in view of
the number of test conditions to consider: normal conditions,
fault conditions, switches to open, switches to close,
connections to the power cord, and connection to patient
leads. How can this be done in a cost-effective and timely
fashion in a production environment? Single-function testers
can perform the hipot, ground bond, or leakage test, but
individual testers do not present an efficient method
for electrical safety testing of medical products.
Through
the use of microprocessor control and relay switching,
it is possible to perform these functions from a single
unit. Some commercially available instruments are specifically
designed to perform all the required tests automatically.
To fulfill this function, a tester must be capable of
running sequential tests, and it must automatically configure
the connections for each test. This process can be simplified
so that it is nearly invisible to the operator. Test conditions
and configurations can be easily selected through front-panel
menu
programming or remotely by external computer. In either
case, incorporating this type of automatic electrical
safety testing goes a long way toward ensuring the delivery
of safe medical electronic products. It is equally critical
to retain test results at the end of the manufacturing
process to help protect the manufacturer from liability
issues.
Reference
1. IEC
60601-1, Medical Electrical EquipmentPart 1: General
Requirements for Safety and Essential Performance, International
Electrotechnical Commission, Geneva, 1995.
Jim
Richards is marketing engineer for QuadTech (Maynard,
MA). He can be reached at jrichards@quadtech.com.
|
|