A Single-Probe Method for Measuring High-Frequency Adjacent
Electromagnetic Fields
Satoshi
Kazama
A
new measurement technique resolves electromagnetic-field
characteristics down to the lead-frame level of an IC
package.
Identifying
the source of electromagnetic radiation and mitigating
electromagnetic interference (EMI) in electronic equipment
is crucial, particularly as operating frequencies increase
and noise margins shrink. The commonly used method for
measuring electric and magnetic fields is limited both
in the frequencies at which the measurements can be made
and in its resolution. This technique requires both a
magnetic probe, such as a shielded loop, and an electronic
probe to measure the distribution of current from the
adjacent magnetic field and voltage from the adjacent
electric field, respectively. The shielded loops used
to measure current distribution typically require a bulky
shielding structure to prevent electric coupling. In addition
to size issues, shielded loops have proven, in certain
cases, to be ineffective for measuring high-frequency
radiation.
This
article introduces a novel method to measure both electric-
and magnetic-field distributions simultaneously using
a single probe. The structure of the probe is simple,
and the size of the probe can be easily reduced. The resulting
technique allows for the characterization of electric
and magnetic fields at the lead-frame level of integrated
circuits (ICs) and at frequencies up to 3 GHz.
Basics
of the Measurement Technique
The
measurement probe, a simple metal loop, is terminated
with the same impedance at each end. When the probe is
in the presence of electric and magnetic fields, current
will be induced in the sensor. The current that results,
from the magnetic coupling of current and the electric
coupling of voltage, flows in the probe. Figure 1 shows
the probe coupling with voltage V and current
I, both of which are parallel with the probe. The
voltage and current are given by
V
= A sin(wt +
qV) (1)
and
I
= B sin(wt +
qC),
(2)
where
A and B are the respective magnitudes of
voltage and current, and qV
and qC
are the respective phase angles.
 |
|
Figure
1. Probe couplings.
|
The
direction of the current resulting from the magnetic coupling
is different at both ends of the probe, but the direction
of the current resulting from the electric coupling is
the same at both ends, thereby providing different outputs
at the two ends of the probe. The current and voltage
are estimated from this difference.
The
outputs at the two ends of the probe are given by
O1
= a1V
+ b1I
(3)
and
O2
= a1V
b1I,
(4)
where
a1 and
b1 are
coupling coefficients that vary with the position of the
ground and the probe.
From
these relationships, the current and voltage are derived
as
(5)
and
(6)
respectively.
Figure
2 illustrates the new method for estimating the current
and voltage distributions. It shows current I with
current angle f and voltage
V at a point with coordinates (x, y) on
the measurement plane. The probe outputs for the four
rectangular directions (Oxf,
Oxr, Oyf,
Oyr) are given by
Oxf
= a2A(x,
y)sin(wt
+ qV(x,
y)) cos(f)b2B(x,
y)sin(wt
+ qC(x,
y)), (7)
Oxr
= a2A(x,
y)sin(wt
+ qV(x,
y)) + cos(f)b2B(x,
y)sin(wt
+ qC(x,
y)), (8)
Oyf
= a2A(x,
y)sin(wt
+ qV(x,
y)) sin(f)b2B(x,
y)sin(wt
+ qC(x,
y)), (9)
and
Oyr
= a2A(x,
y)sin(wt + qV(x,
y)) + sin(f)b2B(x,
y)sin(wt
+ qC(x,
y)), (10)
where
a2 and
b2
are coupling coefficients; A(x, y)
and B(x, y) are the magnitudes
of the voltage and current, respectively; and qV(x,
y) and qC(x,
y) are the respective phase angles.
 |
|
Figure
2. Estimation of current and voltage distributions.
|
Equations
710 can be used to derive the x-direction voltage
(Equation 11), y-direction voltage (Equation 12), x-direction
current (Equation 13), and y-direction current (Equation
14):
(11)
(12)
(13)
and
(14)
For
these relationships to be correct, the x- and y-direction
voltages must be the same.
Component-Level
Field Distributions
Figure
3 shows the structure of the probe used for the measurement
system. It is composed of a bundled pair of semirigid
cables with a characteristic impedance of 50 W.
The sensing component consists of a metal wire connected
to the two center conductors of the cables. The sensing
wire is about 1.0 x
0.5 mm (0.039 x
0.019 in.). The current and voltage of the measurement
plane couple with the metal wire. Both cables are connected
to the inputs of the measuring instrument. This simple
structure makes it easy to reduce the size of the probe.
 |
|
Figure
3. Structure of the probe.
|
Figure
4 shows the setup of the measurement system for testing
a digital IC device. A three-dimensional actuator controlled
by a personal computer drives the scanning movement of
the probe. A fixed probe, which measures the reference
signal, and the two cables of the scanning probe are connected
to a three-input phase-difference measurement system.
The vector outputs from the probe are measured using the
three-input phase-difference measurement system, which
measures both the magnitude and phase difference of the
high-frequency signals between the three channels.
 |
|
Figure
4. Outline of the measurement system.
|
Figure
5 shows an outline of the three-input phase-difference
measurement system. The measured input signal is down
converted to an intermediate frequency and digitized by
an analog-to-digital convertor. The signal magnitude,
frequency, and phase are derived by fast Fourier transform.
All inputs are synchronized by using the same local oscillator
and clock signal, preserving the phase difference between
the channels. The upper frequency limit of the system
is 3 GHz.
 |
|
Figure
5. The three-input phase-difference measurement
system.
|
The
probe is shifted between two rectangular orientations
while it continuously scans above the test device. The
vector outputs at each of these probe positions are measured
and then used to calculate the electric- and magnetic-field
distributions.
Current
and Voltage Distributions in a Digital IC
Figure
6 shows how the current and voltage distributions of a
300-mil small-outline-package complementary metal-oxide
semiconductor digital IC (74HC04) were measured using
this method. The IC was mounted on a printed circuit board
(PCB), and six inverters were connected to the IC and
driven by a 40-MHz signal wave. The source voltage was
5 V. The probe scanned in 0.5-mm steps on a plane approximately
1 mm above the IC package.
 |
|
Figure
6. Measured IC.
|
Figures
7a to 7d show the estimated x-direction current and voltage
distributions in the IC at the third harmonic (120 MHz)
of the signal wave. Because the coupling coefficients
are unknown, the magnitude scales are relative. Using
this technique, estimates of y-direction distribution
produced a complementary set of figures (not shown). The
same technique was used to produce a family of graphics
estimating the second-harmonic distribution (also not
shown).
 |
|
Figure
7a. Current distribution (relative magnitude) in
a digital IC (120 MHz, x direction).
|
| |
 |
|
Figure
7b. Current distribution (relative phase) in a digital
IC.
|
| |
 |
|
Figure
7c. Voltage distribution (relative magnitude) in
a digital IC.
|
| |
 |
|
Figure
7d. Voltage distribution (relative phase) in a digital
IC.
|
To
verify these estimated results, the voltage distributions
(magnitude and phase) in the x and y directions were compared.
Both the magnitudes and phase angles coincided exactly
at the second and third harmonics, indicating that this
measurement system produces reliable results.
The
current magnitude (Im) distributions
were calculated by using
(15)
where
Ix and Iy
are the current magnitudes in the x and y directions,
respectively.
As
shown in Figure 8, the third-harmonic current flowed mainly
in the input-output lead frame of the inverter. A similar
pattern was seen at other odd-numbered harmonics. As shown
in Figure 9, the second-harmonic current flowed mainly
in the source and ground lead frame.
 |
|
Figure
8. Current distribution at third harmonic (120 MHz).
|
| |
 |
|
Figure
9. Current distribution at second harmonic (80 MHz).
|
A
similar pattern was seen at other even-numbered harmonics.
The waveform in the input-output lead frame is therefore
trapezoidal and consists mainly of odd harmonics, whereas
the current in the source and ground lead frame contains
even harmonics.
Probe
Output Characteristics
The
output characteristics of the probe were measured using
a microstrip line. The probe was set parallel to the microstrip
line, as shown in Figure 1. (O1
O2 is referred to
as the current output of the probe, and O1
+ O2 is referred to as the
voltage output.) The microstrip line was formed on a 1.6-mm-thick
PCB with a 50-W characteristic
impedance. One end of the line was connected to a signal
generator with a 10-dBm output, and the other end was
connected to an impedance of 50 W.
The current in the line was about 14.2 mA, and the voltage
was approximately 0.71 V.
Figure
10 shows the current output distribution above the line
at 1 GHz. The probe was set parallel to the microstrip
line, and scanning was done in 0.2-mm steps along the
upper part of a cross section of the line.
 |
|
Figure
10. Current output distribution by probe position.
|
Figure
11 shows the voltage output distribution above the line
at 1 GHz. (The hatched squares in Figures 10 and 11 show
the width of the line.) These distributions show the relationship
between the probe output and the position of the probe.
Figure 12 shows the relationship between the current and
voltage outputs of the probe and the distance from the
center of the microstrip line. The signal frequency was
1 GHz, the current in the line was about 14.2 mA, and
the voltage in the line was about 0.71 V. See Figure 13
for a diagram of the measured microstrip line.
 |
|
Figure
11. Voltage output distribution by probe position.
|
| |
|
|
|
Figure
12. Voltage and current output by probe distance.
|
| |
 |
|
Figure
13. Measured microstrip line.
|
The
current and voltage distributions in the microstrip line
shown in Figure 14 were derived from the relationships
illustrated by Figures 12 and 13. In Figure 12, output
voltage is about 0.011 V at a distance of 1 mm, and the
output current is about 8 mA at a distance of 1 mm. Therefore,
the voltage par output in Figure 13 is

and
the current par output is

The
voltage and current are proportional to their respective
outputs. As shown in Figure 14, these distributions show
characteristics typical of standing waves in a transmission
line, indicating that the current and voltage distributions
can be measured simultaneously by this method.
 |
|
Figure
14. Current and voltage distribution on a microstrip
line.
|
The
output characteristics of the probe varied with the thickness
of the PCB. They are sufficiently different from those
shown in Figure 12; therefore, these relationships must
be measured for each case. The output characteristics
also varied with the frequency, as shown in Figure 15
for a 50-W microstrip line
on 1.6-mm-thick PCB when the probe was set 1.3 mm above
the center of the line.
 |
|
Figure
15. Voltage and current output by frequency (T
= 1.6 mm).
|
EMC
Countermeasure Application
To
demonstrate the effectiveness of the high-frequency adjacent-electromagnetic-distribution
measurement system in the application of EMC countermeasures,
an experiment was performed using a test board, the upper
half of which was connected to supply voltage and the
lower half to ground. A logic IC was connected to a 50-MHz
input signal.
The
purpose of the experiment was not only to demonstrate
the performance of the new measurement technique but also
to show that the careful application of the appropriate
device in the circuit will dramatically reduce the effects
of electromagnetic fields. The basic circuit consisted
of a logic IC, bypass capacitors, power and ground connections,
and a selected countermeasure inductor, a Taiyo Yuden
(Tokyo, Japan) model FBMH. High-impedance devices like
the FBMH are appropriate for applications in which a large
withstand current (≤ 2 A) is needed to mitigate
radiated and conducted noise, particularly at frequencies
≥ 1 GHz. This type of inductor should also have
low dc resistance (Rdc) to
reduce heat generation and power loss in low-voltage/high-current
applications, including cpu VCC
lines. (The 150-W FBMH has
a maximum Rdc value of 0.050
W, whereas the 220-W
part has a rating of 0.070 W.)
Using
the adjacent-electromagnetic-distribution system, the
electric and magnetic fields were measured across the
test platform both before and after the placement of the
inductor. Figures 16 and 17 depict the resulting reduction
in the magnetic and electric fields, respectively.
 |
| |
 |
|
Figures
16 and 17. The levels of radiation from the electromagnetic
field of an IC are compared before and after placing
a high-impedance inductor in the test circuit. The
"after" pictures clearly show a significant reduction
in radiation after installation of the countermeasure.
|
Conclusion
Because
the new measurement technique can resolve electric- and
magnetic-field characteristics down to the lead-frame
level of an IC package, the application of appropriate
countermeasures at the optimum location in the circuit
can be quickly and easily accomplished, thereby speeding
the new-product development process. As operating frequencies
continue to increase and logic voltage levels continue
to decrease, the importance of effective EMC countermeasures
has never been greater. This technique not only speeds
resolution of EMC concerns but also has the potential
to reduce costs by optimizing the use of required components.
References
1.
H Wakuba et al., "Estimation of the RF Current at
IC Power Terminal Using Magnetic Probe with Multilayer
Structure," in Technical Report of IECE, Electromagnetic
Compatibility Japan 98-6: 3943.
2.
A Namba, O Wada, and R Koga, "Measurement of Near-Field
Emission from Printed Circuit Board Using Miniature E-Field
Probe," in Technical Report of IECE, Electromagnetic
Compatibility Japan 98-46: 4146.
3.
Y Kami and T Tobana, "Measurement of Magnetic Nearfield
on Printed Circuit Boards by Using a Magnetic Loop Antenna,"
in Proceedings of International Zurich Symposium on
EMC (Zurich, Switzerland: EMC Zurich, 1997): 591596.
4.
S Yabukami, M Yamaguchi, and KI Arai, "HF-UHF Band
Electromagnetic Measurements Using Multi-Layer Printed
Wiring Board," in Technical Report of IECE, Electromagnetic
Compatibility Japan 97-37: 2126.
5.
T Kurouchi, A Ohneda, and A Hasumi, "Research of
Noise Measurement Technology in Minute Area," in Technical
Report of IECE, Electromagnetic Compatibility Japan
94-12: 3337.
Satoshi
Kazama is a member of the product evaluation team at the
Taiyo Yuden EMC Center (Tokyo) and is a doctoral candidate
in electrical engineering at Tohoku University (Sendai,
Japan).