Medical
Devices: Diagnosing the New EMC Standard
A significantly expanded electromagnetic compatibility (EMC) standard
for medical devices highlights the need for risk analysis.
 |
| Photo
courtesy of CKC Laboratories (Mariposa, CA) |
In
September 2001, the International Electrotechnical Commission
(IEC) published the second edition of IEC 60601-1-2, "General
Requirements for SafetyCollateral Standard: Electromagnetic
CompatibilityRequirements and Tests." This new document
significantly expanded the original 15-page standard to an extensive
93-page set of electromagnetic compatibility requirements for
medical devices.
In addition to being published as an IEC standard, the second
edition of IEC 60601-1-2 has been adopted as a European Norm
(EN 60601-1-2). Although the standard is published by two different
organizations, the requirements of both the European and IEC
documents are identical.
Whether manufacturers use the IEC or the EN publication, the
standard plays a crucial role in demonstrating that an electrically
powered medical device complies with regulatory requirements
(as did its predecessor). In the United States, the Food and
Drug Administration (FDA) has officially "recognized" the new
standard, and the countries of the European Union (EU) have
"harmonized" it for use in demonstrating regulatory compliance.
In both the United States and Europe, standards such as 60601-1-2
(harmonized or recognized standards) are specified only as "a
way" of proving that a device meets regulatory requirements.
Use of such standards, therefore, is not required. However,
when alternative approaches are used, regulators will frequently
use these standards as a benchmark.
Neither FDA nor the EU expects instantaneous upgrading of devices
to the new standard. In fact, the EU will effectively allow
use of the old standard until November 2004. However, the extent
of the effects on products (including accompanying documents
and labeling) suggests that manufacturers should begin identifying
necessary changes to bring products currently in production
into compliance. Manufacturers should also begin defining a
strategy for implementing changes to these products as soon
as possible. Manufacturers should also evaluate products currently
in the development phase of the product life cycle and determine
whether they should be designed to the new standard. To provide
a starting point for this process, this article provides an
overview of the new standard.
The requirements of IEC/EN 60601-1-2 cover three basic aspects
of EMC in electrically operated medical devices. The first is
the general requirements that relate to the overall device and
do not directly relate to physical testing. With only four subclauses,
the requirements in this category are the least extensive but
were not addressed in the first edition.
The second category addresses labeling and accompanying documents.
Although the first edition did contain some requirements applicable
to these items, the second edition has significantly expanded
the amount of documentation required, as well as the level of
detail that must be provided to equipment users.
The third category addresses the technical and performance requirements
to which devices will be tested. As in the first edition, this
category has two subcategories: limitations on the emissions
of the device and immunity of the device to electromagnetic
interference (EMI). The significance of the changes in this
area varies greatly, depending on the specific device in question.
In all cases, the information in this article is intended only
as an overview. To identify actions required for a specific
device, manufacturers must do so based on the standard itself.
General Requirements (Clause 3)
Before addressing the new requirements, it is critical to understand
a fundamental principle that applies to both the first and second
editions of 60601-1-2. This principle is crucial in properly
evaluating equipment, especially in terms of the technical requirements
for a device to be immune to the effects of electromagnetic
radiation.
Although IEC/EN 60601-1-2 is an extensive document, it is not
intended to be used by itself. It is, in fact, part of an entire
family of standards (covering general safety, specific technologies
implemented in the equipment, and specific issues related to
particular devices). When used in combination, these standards
are intended to ensure the overall safety of electrically operated
medical equipment. The parent document (IEC/EN 60601-1) of the
entire family (including 60601-1-2) establishes the baseline
for applying the requirements of all standards within the family.
The current (second) edition of the 60601-1 parent document
(or general standard) establishes the scope of application of
all standards within the 60601 family by stating that it applies
to the safety of medical equipment. This scope is reiterated
in the general requirements of the standard, which states that
equipment that does not comply with the letter of the standard
is considered to be acceptable if "an equivalent level of safety
is provided." The general standard is currently being revised,
but this principle remains crucial when applying the general
standard or other documents in the family.
The principle of equivalent safety has only minimal application
in terms of the emission of electromagnetic energy. This is
because the basis of the emission requirements is to prevent
interference with the safe operation of other medical equipment
and to meet internationally accepted legal requirements not
to interfere with broadcast as described in Clause 3.201.1 of
the EMC standard. However, the concept of equivalent safety
is fundamental in determining which functions of the equipment
being tested are subject to immunity requirements. Only equipment
failures that present an unacceptable risk to persons, animals,
or the environment are considered a violation of the requirements
in the 60601 family of standards.
Closely related to this basic principle of safety is the concept
of essential performance in Clause 3.201.2 of IEC 60601-1-2.
Essential performance is any functional aspect of a device that,
if not performed as intended, would result in an unacceptable
risk (a concept introduced in the first draft of the third edition
of the general standard). Some of the more obvious examples
of essential performance include operation of a ventilator or
defibrillator. Failure of a ventilator could result in suffocation
of the patient. Because a defibrillator is used only in emergency
situations, failure to operate could easily result in death
as well. Less-obvious examples include the failure of some types
of diagnostic equipment in which a failure to provide the appropriate
treatment (due to incorrect information) could result in injury.
It is, therefore, critical that the first step in determining
compliance with the EMC standard (or any standard in the 60601
family) is to perform a thorough risk analysis. Ideally, such
an analysis should be done as part of an overall risk management
process as defined in IEC/ISO 14971. It is important to note
that compliance with IEC/ISO 14971 will be a requirement of
the third edition of the IEC 60601-1 general standard scheduled
for publication in 2005.
In testing equipment to the second edition of 60601-1-2, certification
bodies will now need to make sure that the manufacturer has
performed a risk analysis to (at minimum) identify the safety-related
aspects of the device; otherwise, the certifier will assume
that all characteristics of the equipment are safety related.
Although the wording of IEC 60601-1-2 implies a responsibility
on the part of certifiers to ensure that a safety-related risk
analysis was performed, certifiers are not to determine the
validity of the judgments made in evaluating those risks. The
manufacturer remains the authority on their device and its safe
use.
In Clause 3.201.4, the new EMC standard also introduces (in
terms of electromagnetic compatibility) the concept of medical
systems that utilize or connect to other medical devices or
other nonmedical equipment, such as personal computers, during
operation. Such configurations have become increasingly common
in recent years. The IEC 60601-1-2 standard states that such
equipment does not need to be tested for EMC compliance if it
is unlikely that the nonmedical equipment will affect the essential
performance of the medical equipment, or is unlikely to cause
the system to exceed acceptable emissions limits.
For equipment in which this is asserted to be the case, manufacturers
must provide supporting documentation. Such documentation can
include the manufacturer's risk analysis or proof that the nonmedical
equipment complies with the appropriate international EMC standards.
Whether medical or nonmedical, equipment intended to be connected
to another medical device by supplying power, interchange of
data, or even functional mechanical connections constitutes
a medical system when applying the standards in the IEC 60601
family. Whether a single device or a medical system, equipment
must remain safe. In the case of the EMC standard, equipment
must also remain compliant with emissions limits during normal
use and under any reasonably foreseeable single-fault condition.
Identification, Marking, and Documentation (Clause 6)
Increasing the amount of information to be provided to users
of medical equipment presents no technical challenges to manufacturers.
It does, however, place additional burdens on the design process
and adds costs to the development and maintenance of a device.
It is important to note that when applying the EMC standard,
disclosing information on the electromagnetic characteristics
of a device or system can allow significant flexibility in the
physical design. Such flexibility is possible when the characteristics
of a device make literal compliance with the requirements physically
impossible or financially unfeasible.
Clause 6 of IEC 60601-1-2 identifies the information that must
be disclosed. It addresses both information requirements for
users or operators, and for installation and maintenance personnel.
It covers information that must be provided through labeling
(marking on the outside of the equipment) or by placement in
a device's accompanying documents (instructions for use or technical
description).
The requirement to use the appropriate symbol on equipment incorporating
intentional radiators (transmitters or equipment intended to
radiate to achieve clinical results) has been retained (in Clause
6.1.201.1) from the first edition. In addition, the standard
now requires (in Clause 6.1.201.3) that a warning label be placed
on equipment when operation in an electromagnetically shielded
location is necessary to meet the requirements of the standard.
In the technical requirements of Clause 36.2.202.2, the standard
has added the possibility of a technical exemption from requirements
to perform electrostatic discharge (ESD) testing on input and
output ports. When this exemption is exercised, a "no touch"
symbol must be placed in proximity to the connector in question
(per Clause 6.1.201.2).
Information provided in instructions, which is targeted to clinicians
or even patients (for home-use equipment), is generally not
highly technical. Clause 6.8.2.201 requires that manufacturers
provide the following information in the instructions for use:
-
Electrically
operated medical devices require special care (in terms
of EMC) when being installed. (A reference to the location
of such information must be included in the technical
description.)
-
A
warning that portable and mobile radio-frequency (RF)
communications equipment can interfere with the equipment's
operation.
The
standard also requires that when equipment does not provide an
option to manually adjust for sensitivity (or gain) of physiological
signals, the instructions must specify minimum input levels and
warn that failure to provide the minimum level could result in
inaccurate results. The wording of this requirement can be misleading.
It is important to remember that this standard addresses only
EMC and is, therefore, applicable only to equipment using electromagnetic
energy, such as RF, for transmission of such physiological data
from the patient to the equipment. Although the scope of IEC 60601-1-2
effectively limits the application of this requirement to wireless
telemetry, it would be wise to apply the principle to wired systems
where appropriate.
The EMC standard (or any other IEC 60601 standard) requires
manufacturers to provide a technical description that includes
a significant amount of information related to installing and
maintaining the equipment. The information can be provided in
one of four ways:
-
-
As
a part of the instructions for use (when no service manual
is provided).
-
As
part of the setup, installation, and maintenance documentation.
-
In an appropriate combination of such documents.
It is important to keep in mind that the target audience for
the technical description is the installation, service, and
maintenance personnel. This includes technicians who install,
maintain, and in many cases repair the equipment, and those
who need the information throughout the product life cycle to
maintain compliance.
To ensure ongoing compliance in terms of equipment EMC, technical
personnel may require information such as identification of
all cables and their lengths (where incorrect replacement could
change the emissions characteristics or sensitivity to electromagnetic
energy). They may also need to know specific technical characteristics
of system components, accessories, or transducers that could
affect EMC compliance. The extent of the information provided
may be greatly reduced when specific written instructions indicate
that only the manufacturer can service the equipment or provide
replacement components and accessories. When determining what
information will need to be provided and where it will be presented,
it is imperative to focus on the purpose of the requirement
(the audience, what do they need to know, and when do they need
to know it).
Other information and warnings required in the technical description
include:
-
A warning should state that stacking or placing equipment
adjacent to other devices is not recommended, and that where
such configurations are necessary, all equipment in question
should be carefully observed to ensure that EMI does not degrade
performance.
-
When manufacturers intend other equipment they provide (but
which is not part of a system) to be stacked or placed in
proximity to the equipment (being certified to the standard),
manufacturers should describe all testing that purchasers
must perform to ensure that no risks are associated with the
anticipated configuration.
-
If test levels lower than those specified by the standard
for immunity are used (which is allowed under certain conditions),
manufacturers must indicate the levels used and an explanation
as to why such reduced levels were necessary.
-
For equipment intended for use in shielded locations, manufacturers
must list other equipment allowed and prohibited from use
within the shielded area.
-
If the equipment generates RF energy to produce clinical effects
(diagnosis or treatment), instructions for avoiding or alleviating
disturbances to other medical equipment must be provided.
-
Where the equipment receives electromagnetic energy in order
to perform its intended function (such as wireless telemetry),
each frequency or reception band used and, where applicable,
the preferred frequency or band must be identified. A warning
must indicate that emissions from other equipment may prevent
correct operation of the device.
-
For devices that incorporate RF transmitters, the frequency
or frequencies and band or bands of transmission must be identified.
In addition, the characteristics of the modulation and effective
radiated power must be given.
-
For equipment or systems in which the risk analysis has determined
that the failure to perform any intended function or the degradation
of such a function does not constitute an unacceptable risk
(i.e., there are no essential performance characteristics)
and in which possible failure modes (caused by electromagnetic
disturbances) will not result in an unacceptable risk, a statement
must be included indicating that testing for immunity has
not been performed. This statement will replace Tables 202
through 208 described below.
Most standards in the IEC 60601 family generally avoid defining
the format of information. The EMC standard, however, does specify
the format of tables that are to be included (where appropriate)
in the technical description. These tables define:
-
Emissions characteristics of the equipment or system (Table
201).
-
Equipment or system sensitivity to ESD (Table 202).
-
RF immunity of life-support equipment or system (Table 203).
-
RF immunity of non-life-support equipment or system (Table
204).
-
Recommended separation distances between life-support equipment
or system and portable or mobile RF communications equipment
(Table 205).
-
Recommended separation distances between non-life-support
equipment or system and portable or mobile RF communications
equipment (Table 206).
-
RF immunity of life-support equipment or system that is specified
for use in shielded locations (Table 207).
-
RF immunity of non-life-support equipment or system that is
specified for use in shielded locations (Table 208).
In
each case, the standard provides both text and flowcharts that
define how the information required in the tables is determined.
For equipment or systems that do not provide essential performance
and where no unacceptable risk is associated with possible failure
modes, Tables 202208 need not be provided.
Technical Requirements (Clause 36)
The technical requirements category includes two subcategories.
The first addresses limits on the emissions (for electromagnetic
energy radiated through the air or conducted through the power
or mains connection) of electromagnetic energy by the equipment
or system. The second subcategory sets requirements related
to how the equipment or system reacts to the electromagnetic
environment (immunity).
Emissions (36.201). For the most part, the allowable emissions
levels under the second edition of IEC 60601-1-2 have been left
unchanged from those of the first edition. This is because the
emissions limitations are primarily based on requirements written
by the International Special Committee on Radio Interference
(CISPR). The requirements are implemented in the national law
of participating countries based on international treaty.
Under both editions of the medical device EMC standard, most
medical device emissions are required to be in compliance with
CISPR 11 (emissions limits for industrial, scientific, and medical
equipment). However, under the second edition, simple medical
equipment (not systems) that includes no circuitry operating
at greater than 9 kHz may be tested under CISPR 14 (emissions
limits for household appliances and tools).
Medical lighting equipment may be classified and tested according
to CISPR 15 (lighting), and information technology equipment
(ITE) that is either intended to be connected to the device
or is part of a system may be classified and tested to CISPR
22 (ITE emission requirements). For equipment or systems specified
for use in electromagnetically shielded environments, the appropriate
limits of CISPR 11 may be increased (made less stringent), but
when this is done, the levels used must be disclosed in the
technical description (as described previously).
For equipment being classified and tested to CISPR 11 or 22,
the most important determination is whether the equipment falls
under Class A (industrial environments) or Class B (domestic
environments). The higher (less stringent) limits of Class A
are intended to be used for equipment that will be operated
in environments such as hospitals, where the likelihood of negative
effects to public broadcasts are unlikely and where the mains
supply (ac wall voltage) is likely to be isolated from the public
distribution network. Class B limits are more restrictive because
the levels of electromagnetic isolation in such areas are unreliable.
In addition to the requirements of the identified CISPR standards,
IEC 60601-1-2 adds two requirements addressing issues related
to conducted emissions for devices that are rated to draw ¾16
A per phase and are intended to be connected to the public mains
power distribution network. Neither of these requirements was
included in the first edition.
The requirements in question address harmonic distortion, voltage
fluctuations, and flicker generated by the device on the mains
power connection (power cord). These requirements can be of
particular concern for manufacturers incorporating switch-mode
power supplies and some types of ac motors. Such systems can
cause distortion in the ac waveform or very-short-term voltage
drops that are reflected back into the power lines. These reflections
can cause interference in some radios and televisions that are
also connected to the same mains network. Addressing harmonics
in medical equipment used outside of hospitals and similar locations
where isolation from the public distribution network does not
exist requires a delicate balancing of harmonics and leakage
current limits (established in the IEC 60601-1 general standard)
if filter capacitors (to earth) are used to reduce the distortion.
Immunity (36.202). The most important question related to immunity
testing relates to determining which aspects of the medical
device or system are subject to the requirements. As explained
in the discussion of the general requirements, this determination
is made by evaluating the equipment to identify which functions
and aspects of the design could give rise to hazards that could
result in unacceptable risks. Only risk-related aspects of the
device or system identified during the risk analysis are required
to be immune from the EMI identified in Clause 36.202.
Another aspect of the IEC 60601-1-2 standard that should be
carefully considered is that the levels of interference to which
it requires immunity are based on a typical healthcare environment.
This condition means that when the equipment is intended for
use in other environments, it may be appropriate to increase
or decrease the levels of interference used to test immunity.
Annex EEE of the standard attempts to provide insight into levels
found in some of these other environments. Additionally, Clause
36.202.1 provides guidance on many aspects of performing testing
for immunity.
During immunity testing, equipment is exposed to several types
and levels of EMI, including ESD; RF; electrical fast transients
(EFTs) and bursts; power-line surges; conducted RF disturbances;
voltage dips, interruptions, and variations on power input lines;
and power-frequency magnetic fields.
ESD. The first edition of the standard included requirements
for ESD testing, but the second edition has raised the bar quite
a bit. Where the old standard required that the equipment be
exposed to air discharges of 8000 V and contact discharges of
3000 V (both positive and negative polarity), the new version
requires air discharges of 2000, 4000, and 8000 V (both polarities)
and contact discharges of 2000, 4000, and 6000 V (both polarities).
As mentioned during the review of Clause 6, it is possible for
some connectors to be exempted from ESD testing if appropriate
labeling and disclosure requirements are met.
RF. The second edition expands RF immunity requirements
in three areas. First, the highest frequencies used are increased
from 1.0 to 2.5 GHz. The second important change is the modulation
of the signal to which the equipment is exposed. The first edition
required that the signal be amplitude modulated at 1000 Hz.
The second edition requires equipment that controls or monitors
physiological parameters (e.g., heart rate) be modulated at
2 Hz (closer to the frequencies of such biological parameters).
Equipment that does not fall into this category is tested at
a modulation frequency of 1000 Hz. The first edition did not
specify the modulation level; however, the new standard sets
it at 80%. Finally, equipment such as ventilators or other life-support
equipment must now be tested for immunity to RF at a field strength
of 10 V/m; all other equipment is still tested at 3 V/m. As
with ESD, the second edition allows the RF field-strength levels
used for testing to be eased under certain conditions (e.g.,
when the device is to be used only in shielded locations).
EFTs and Bursts. The second edition of the standard has
increased the test levels for EFTs and bursts from 1 kV for
equipment with power cords and plugs, and 2 kV for permanently
installed equipment, to 2 kV for all mains connections (including
power cordconnected equipment). Test levels for interconnecting
cables greater than 3 m in length have also increased from 0.5
to 1 kV.
The new standard has also added significant detail in terms
of test configurations for equipment with patient connections,
that is, any conductive connection to the patient, although
they are now specifically exempted from the application of EFT
either conductively or through the capacitive clamp. Although
EFT has not been a significant source of failures for most equipment,
the doubling of the interference for all but power connections
of permanently installed equipment may cause problems for sensitive
equipment that has many functions deemed as essential performance.
Power-Line Surges. The most notable change related to
surge testing in the new revision of IEC 60601-1-2 was to actually
make a common practice for EMC certification bodies under the
old standard a requirement in the second edition. That practice
was to test not only at 2000 V between power (input) lines and
the equipment's ground connection but also at intervals of 500
and 1000 V as now specified in the second edition. Surge tests
between power (input) conductors are now also conducted (as
was previously practiced, but not strictly required) at 500
and 1000 V.
Conducted RF Disturbances. The second edition introduces
requirements and tests for these conditions, which were identified
only as "under consideration" in the first edition. As with
radiated RF fields, the conducted RF immunity tests require
a signal strength of 10 V/m for life-support equipment and 3
V/m for all other categories. The test signal is swept from
150 kHz (for all but some types of battery-powered equipment)
through 80 MHz.
Voltage Dips, Interruptions, and Variations on Power Input
Lines. As is the case for many other types of interference,
the second edition adds these requirements where the first only
referred to them as "under consideration." This section requires
that equipment remain safe during and after voltage dips (in
mains voltage) of 95% (dropping mains down to only 5% of nominal)
for a duration of 1/2 cycle, 60% (mains
at 40% of nominal) for 5 cycles, and 30% (mains at 70% of nominal)
for 25 cycles. Equipment is also expected to remain safe when
mains voltage is dropped to 5% of nominal and then restored
after 5 seconds.
The second edition also requires that systems with battery backup
return to operation from mains after the tests. However, if
this last requirement is viewed in light of risk, it is clear
that the requirement is valid only if 1) the user were not made
aware of the fact that the system was operating on battery power
and 2) the available charge might not allow clinicians to take
reasonable actions to prevent any unacceptable risk.
Power-Frequency Magnetic Fields. The first edition of
IEC 60601-1-2 contained no specific requirements for magnetic-field
immunity. The second edition, however, requires that the equipment
remain safe when exposed to magnetic field strengths of 3 A/m
at both 50 and 60 Hz (even if powered internally) unless the
equipment is intended for use only where one of the two power
frequencies is available (e.g., equipment intended only for
sale in Europe would be tested only at 50 Hz.). The standard
specifically allows testing to be performed with the equipment
powered at any allowable mains voltage level.
Although the second edition of the IEC 60601-1-2 standard undoubtedly
provides significantly more information (both in the normative
text and annexes) than its predecessor, the way the information
is presented can easily become confusing. The new standard correctly
points out that only safety-related failures of the equipment
are of interest in some of the immunity clauses. Other clauses,
however, seem to indicate that any deviation (regardless of
the relation to safety) from normal operation is a failure.
This complexity and lack of clarity increases the possibility
that those testing a product could incorrectly identify failures
if adequate documentation is not provided before testing.
The importance of the risk analysis in defining the acceptance
criteria for the testing cannot be overstated. In preparing
to have a product tested, the first step should be to identify
which aspects of the equipment's operation qualify
as essential (safety-related) performance and which EMC-initiated
failures could result in unacceptable risk. Stated succinctly,
manufacturers must determine which equipment operations must
be maintained and which cannot self-actuate in order to ensure
safety.
It is also crucialboth in terms of efficient testing and
safety of the equipmentfor a manufacturer to understand
the electromagnetic environment in which the equipment or system
will be used. Annex EEE provides some insight into this issue
and can prove quite helpful in understanding it. In fact, the
standard provides a great deal of background information that
is helpful in applying it to equipment. Annex AAA attempts to
explain the rationale for nearly every requirement in the standard.
Annex BBB contains examples of the tables that are required
in the technical description. The intended audience of Annex
DDD is actually the writers of device-specific standards (called
"particular" standards); however, where no such standard exists,
this section may provide some help to manufacturers in applying
the EMC standard.
It is now more critical than ever that medical device manufacturers
thoroughly think through and document their strategies for EMC
compliance. In particular, it is essential that they bring the
test house and certification body into the loop early. Failing
to do so will almost certainly mean that EMC testing and certification
will be far more costly and time-consuming than it needs to
be. Taking the time to document the information needed before
contacting an EMC test house will prove to be a valuable investment.
Such information includes essential performance characteristics,
electromagnetic environment, typical configurations of the equipment,
and block diagrams identifying the operating frequencies of
subsystems. Providing this information to the test house as
early as possible in the design process will prove equally valuable.
(Many test houses even provide EMC design review services.)
This documentation process has always been important, but the
second edition of IEC 60601-1-2 makes it critical.
Mike Schmidt is secretary of IEC subcommittee 62D, cochair
of the AAMI Electrical Safety Committee, and the committee liaison
to the AAMI EMC Committee. Schmidt is also the senior standards
associate for Ethicon Endo-Surgery (Cincinnati, OH). He can
be reached at mschmid6@eesus.jnj.com.
|