An
Accurate Validation Procedure for Component Testing Chambers
Alexander
Kriz and Wolfgang Müllner
A modified validation technique for absorber-lined shielded enclosures
improves repeatability and avoids impedance problems.
The
results of radiated emissions measurements depend on the wave
propagation characteristic of the test site. This characteristic
must be defined to ensure good reproducibility and to obtain similar
results from different test facilities.
An absorber-lined shielded enclosure (ALSE) can be used for automotive
component testing as long as some requirements are fulfilled.
All standards currently prescribe a comparison between an open-area
test site (OATS) and an ALSE. Unfortunately, there is neither
a procedure nor a tolerance of the correlation that is defined
in ISO 13766, ISO 14892, Council Directive 95/54/EC, and Council
Directive 97/24/EC.14 A detailed procedure and
a limit are defined in CISPR 25 and SAE J1113/41.5,6
The current ALSE validation technique described in CISPR 25 induces
some problems. This article suggests a modified ALSE validation
procedure with small conical dipole antennas instead of the test
harness as described in the standard. This procedure improves
repeatability and avoids the impedance problems of the artificial
network and the noise source.
Current Validation Technique
Calibration of an ALSE is described in Annex B of CISPR 25.
Figure 1 shows the test setup. This setup uses a noise source
rather than the equipment under test to generate an electric field.
For this investigation, a RefRad comb generator from ARC Seibersdorf
was connected via a simple wire to the artificial network. The
electric field (E) of this test harness was measured with three
antennas: a monopole antenna at frequencies of 150 kHz to 30 MHz,
a biconical antenna from 30 to 200 MHz, and a log-periodic antenna
from 200 MHz to 1 GHz.
Two
field-strength measurements were performed: The first field strength
was measured on an OATS as reference, and the second was the measurement
in the ALSE. The difference between the two measured field strengths
is the subject of investigation.
A chamber is assumed to be compliant if the deviation does not
exceed ±6 dB in the frequency range of 70 MHz to 1 GHz. No
limits are given for other frequency ranges.
Problems
with the Current Technique
The standardized procedure induces several problems above 100
MHz:
-
-
No
defined impedance of the artificial network
-
No
defined impedance of the noise source.
-
No defined grounding of artificial network and noise source.
Some
of the technique's problems have been described by Swanson and
by Miller.7,8
The
radiation characteristic of the wire causes the poor repeatability.
The length of the wire is 1.5 m, which corresponds to a wavelength
of 1x
for a frequency of 200 MHz. So for higher frequencies, the wire
acts as a Beverage antenna. At a frequency of 1 GHz, the wire
is 5x
long, and the directional pattern shows many lobes. These lobes
are very sensitive in both direction and amplitude to the position
of the wire. If the position of the wire is changed by 5 mm, the
field strength changes by several decibels (see Figures 2 and
3). Below a frequency of 100 MHz, the traces are within 2 dB.
Above this frequency, the situation becomes worse. Table I summarizes
the maximum deviation of the positioning experiment.
|
Antenna
Type
|
Horizontal
Polarization
|
Vertical
Polarization
|
|
Biconical
|
21.7
dB
|
9.0 dB
|
|
Log
periodic
|
11.7 dB
|
18.0
dB
|
|
Table
I. Maximum deviation due to wire position sensitivity.
|
 |
| Figure
2. Sensitivity to wire position, biconical antenna: a) horizontal
polarization and b) vertical polarization. |
 |
| Figure
3. Sensitivity to wire position, log-periodic antenna: a)
horizontal polarization and b) vertical polarization. |
 |
| Figure
4. Impedance characteristics of an artificial network. |
The
impedance for an artificial network is provided for up to 108
MHz in the standard. Annex F of CISPR 25 shows the schematic for
an artificial network. For frequencies above 100 MHz, the 0.1-µF
capacitor can be ignored; however, the inductivity of the cable
to the test harness connector and of the cable to ground become
important. So, the impedance increases above a frequency of ~70
MHz (see Figure 4). A strong resonance can be observed at ~400
MHz, where the impedance reaches nearly 400 W.
The radiation characteristic of the wire antenna will change due
to a standing wave on the wire, which depends on the source impedance.
Therefore, it is essential to use a well-matched 50-W source
such as the RefRad. Alternatively, matching can be improved by
using a 10-dB attenuator at the output of the source.
In general, it is difficult to take measurements over a metallic
table because of the low-impedance connection to ground. The artificial
network and the noise source can be connected to ground in several
ways. The best way is to use wires that are as short as possible,
which then decreases the inductivity.
A new technique modifies the current method for the frequency
range from 30 MHz to 1 GHz. Instead of using the noise source,
the wire, and the artificial network to define field strength,
the new technique uses a small antenna to generate a well-defined
field. For the investigation, this small antenna was placed at
five positions on the table near the former wire position (see
Figure 5a). Table II shows the maximum deviation due to antenna
position sensitivity.
|
Antenna
Type
|
Horizontal
Polarization
|
Vertical
Polarization
|
|
Biconical
|
1.04
dB
|
0.54 dB
|
|
Log
periodic
|
0.58 dB
|
1.12
dB
|
|
Table
II. Maximum deviation due to antenna position sensitivity
measured on antenna P3 as shown in Figure 5.
|
The
transmit antenna can be fed by a signal or by a tracking generator.
A network analyzer also can be used. By measuring the level of
the signal source, this method reduces the drift of the test receiver.
It is not necessary to use the same test receiver for the measurements
on the OATS and in the ALSE. A network analyzer provides very
good accuracy. Attenuators should be used on the feed points of
both antennas to reduce the influence of standing waves and to
further improve the accuracy. The height of the antenna above
the ground plane should be 150 mm to allow measurements in vertical
polarization (see Figure 5b). A precision conical dipole antenna
such as the PCD 3100 (ARC Seibersdorf Research) can be used for
this purpose (see Figure 6).
 |
| Figure
6. PCD 3100 (ARC Seibersdorf Research, Seibersdorf, Austria).
|
The
site attenuation (SA) measurement procedure requires two different
measurements of the voltage received. The first reading (VDIRECT)
is taken with the two coaxial cables disconnected from the two
antennas and connected to each other. The second reading (VSITE)
is taken with the coaxial cables reconnected to the antennas.
The principle of the comparison between the OATS and the ALSE
is the same as described in CISPR 25. Therefore, two site attenuation
measurements should be performed: SAOATS on
the OATS and SAALSE in the ALSE.
(3)
In
Equation 3, the positions of OATS and ALSE change from those
in Equation 1. The reason for this change is the different
sign of the site attenuation. The site attenuation measurements
and a comparison of different sites are extensively described
by Müllner.9
 |
Figure
7. Sensitivity to antenna position using a new technique,
biconical antenna: a) horizontal polarization and b) vertical
polarization. |
 |
| Figure
8. Sensitivity to antenna position using a new technique,
log-periodic antenna: a) horizontal polarization and b) vertical
polarization. |
Measurements
have shown that the problem of poor repeatability is solved by
using the new validation technique. Figures 7 and 8 show the site
attenuation measurements when the position of the small conical
dipole antenna is changed slightly. The antenna was moved 10 mm
to the right, left, front, and back from its original position.
The suggestion for a modified validation method using a small
conical dipole avoids the problems that arise using the CISPR
25 procedure: poor repeatability, undefined impedance of the artificial
network, and unmatched noise source. For optimum accuracy, a network
analyzer can be used for the ALSE validation measurement. This
new technique has been used successfully for an ALSE validation.
This procedure was carried out to get an Automotive EMC Laboratory
Recognition Program (AEMCLRP) accreditation.
1. ISO/DIS 13766, "Earth-Moving MachineryElectromagnetic
Compatibility," Draft International Standard, International Standardization
Organization, Geneva, July 1996.
2. ISO/CD 14892, "Agricultural and Forestry MachinesElectromagnetic
CompatibilityTest Methods and Acceptance Criteria," Final Committee
Draft ISO/TC23/SC2, International Standardization Organization,
Geneva, March 1996.
3. Commission Directive 95/54/EC, "Automotive EMC Directive,"
Official Journal of the European Communities, October 21,
1995.
4. Council Directive 97/24/EC, "Two- or Three-Wheeled Motor
Vehicles: Components and Characteristics," June 17, 1997.
5. CISPR 25, "Limits and Methods of Measurement of Radio
Disturbance Characteristics for Protection of Receivers Used On
Board Vehicles," First edition 1995-11, International Electrotechnical
Commission, Geneva.
6. SAE J1113/41: "Limits and Methods of Measurement of Radio
Disturbance Characteristics of Components and Modules for the
Protection of Receivers Used On Board Vehicles," Society of Automotive
Engineers, Warrendale, PA, July 1995.
7. D Swanson, "Investigation of the Calibration Procedure
from CISPR 25, Annex B, for Use with Vehicle Component Testing,"
in Proceedings of the IEEE International Symposium on EMC,
(Denver: IEEE EMC Society, 1998).
8. P Miller, "A Comparison of Radiated Emissions Testing
to European Directive 95/54/EC Using an Open Field Test Site and
a Semianechoic Chamber," in Proceedings of the IEEE International
Symposium on EMC, (Seattle: IEEE EMC Society, 1999).
9. W Müllner and H Garn: "From NSA to Site-Reference
Method for EMC Test Site Validation," in Proceedings of the
IEEE International Symposium on EMC, (Montreal: IEEE EMC Society,
2001).
Alexander Kriz is a research engineer for Seibersdorf Research
GmbH (Seibersdorf, Austria). Wolfgang Müllner is head of
the business unit for radio-frequency engineering for Seibersdorf
Research. He can be reached at wolfgang.muellner@arcs.ac.at
or via http://itr.arcs.ac.at.
|