|
 |
 |
Methods for Verifying ESD Simulator
Compliance
Greg Senko
Ensuring simulator performance within the
proper limits begins
with calculating EMC uncertainty.
ESD simulator pulses are notoriously difficult
to measure because of their fast rise time
and shot-to-shot variations in waveform. When
measuring ESD simulators, one must consider
all possible error sources and compute the
uncertaintythe estimated bounds of the deviation
of a measured quantity from its true value.
Then a statement of uncertainty should be
made to reflect the quality or accuracy of
a measured result as compared with the true
value,1 which is and will usually
remain unknown. The statement of uncertainty
is accompanied by a statement of confidence
that can be placed in the value of uncertainty.
A calculation of uncertainty should be performed
for each measured parameter of a calibration.
Without these statements of uncertainty and
confidence, one cannot be sure that a given
simulator's performance falls within acceptable
limits.
Calculating uncertainty requires an
uncertainty budget for each measured parameter.
This is a list of the probable sources of error
with an estimation of their uncertainty limits,
probability distribution, and their influence
on the test result. The standard deviations
are calculated for each of the sources in this
list. The stated uncertainty is then calculated
from the root sum of squares of the standard
deviations of the individual uncertainty components
multiplied by a coverage factor, k. The
definitions, procedures, and formulae for calculating
uncertainty can be found in the National Association
for Measurement and Sampling (NAMAS) publication
NIS81, The Treatment of Uncertainty in EMC
Measurements.
Establishing the Uncertainty Budget
Listing all the measured quantities and creating
a table for each one is the first step in
creating an uncertainty budget. The next step
is listing all the probable sources of error
for each quantity, including random sources.
Each error source must then be evaluated for
its bounds and distribution. Sources of information
for estimating the uncertainty include manufacturers'
specifications, calibration reports, estimates
based on experience with the simulator's behavior,
and a series of recorded observations (for
random sources). If a quantity is measured
at more than one point or over a range of
points, it is necessary to decide if the uncertainty
varies over the range. If desired, the calculation
may be simplified by using the worst-case
uncertainty for the entire range.
The following sections give the methods for
estimating the standard deviation for random
and systematic effects. Each error source
must be evaluated with one of the following
methods before calculating the combined and
expanded uncertainty.
Random Components. Random effects
result in errors that vary in an unpredictable
way while the measurement is either being made
or repeated under the same conditions. The uncertainty
associated with these conditions can be evaluated
by statistical techniques from a series of repeated
measurements. An estimate of the standard deviation,
s(qk), of a series
of n readings, qk,
is obtained from
(1)

where q is the mean value of n
measurements.
The estimation of s(qk)
for a random component is part of the process
of creating an uncertainty budget. Typically,
it is performed just once for each random
component. The estimation is only valid if
the EUT measurement is made under exactly
the same conditions. Establishing a value
of standard deviation for each ESD simulator
model to be tested may be necessary, as they
may behave differently.
When testing the EUT, making repeated measurements
instead of just a single measurement will
reduce the random component of uncertainty.
The standard uncertainty, u(xi),
for random contributions is calculated from
the standard deviation of the mean, s(q),
as given by
(2)

If only one EUT measurement is performed,
n = 1 then u(xi)
= s(qk), the
original value. If two or more measurements
are made, n > 1, then the quantity will
be smaller than s(qk)
and thereby the uncertainty is reduced. It
is advisable to make multiple measurements
whenever the measured result is close to the
specified limit, as this will reduce the measurement
uncertainty. The average value of the repeated
measurements is used as the measured result.
Normal Distribution. Normal distribution
is used when there is a high probability of
the true value lying near to the center of
the range, and a low probability of the value
being near the limits of the range (a Gaussian
distribution or bell curve). Normal distribution
is assigned to uncertainties derived from
multiple contributionsfor example, when
a NAMAS- or otherwise-certified calibration
laboratory provides a total uncertainty value
with a stated confidence level for a particular
instrument parameter.
(3)

where uncertainty is the semirange limit
value (± uncertainty).
If the uncertainty is not symmetrical,
+error  error, then
(4)

Rectangular Distribution. Rectangular distribution
is used where there is an equal probability
of the true value lying anywhere within the
given range. A rectangular distribution should
be assigned where a manufacturer's specification
is used as the uncertainty. If there is a statement
of confidence associated with the specification,
normal distribution may be used instead.
If the uncertainty is not symmetrical,
+error  error, then
(4)

(5)

where ai is
the semirange limit value (± ai).
U-Shaped Distribution. A U-shaped
distribution applies to mismatch uncertainty.
Mismatch uncertainty applies to continuous
signals, which is not the case with ESD pulses.
It is included here, however, for completeness.
The value of the limit for the mismatch uncertainty
is M, associated with the power transfer
at a junction, and is obtained from 20 log10
(1|Gg|
|Gl|) dB where
Gg and Gl
are the reflection coefficients for the source
and load:
(6)

Combined Uncertainty
Once the uncertainty budget has been
completed and a standard deviation value has
been assigned to each error source, the combined
uncertainty may be calculated. The combined
standard uncertainty, uc(y),
is calculated by taking the square root of the
sum of the squares of the individual standard
uncertainties. If any of the standard uncertainties
are not already in the same units as the measured
result, they must be converted by the appropriate
function, ci:
ui(y)
= ciu(xi)
(7)
Any contributions with known or suspected
adverse correlation are added together first,
before being entered into the following formula
as ui. The combined
uncertainty for m contributions is
(8)

Expanded Uncertainty
The expanded uncertainty, U, is derived
from the combined uncertainty and defines
an interval around the measured result that
encompasses the true value with a specified
level of confidence, p%:
U=k
uc(y)
(9)
A value of k = 2 approximates a 95%
confidence level (recommended by NAMAS for
EMC testing); a value of k = 3 approximates
a 97.5% confidence level. These are the most-frequently
used values of k. If a significant
portion of the uncertainty is from random
sources,
uc(y)/s( )
3,
then k will be increased. The value
for k is obtained from the t-distribution
based on the effective degrees of freedom, veff,
of the combined uncertainty uc(y)
and the required level of confidence. The effective
degrees of freedom are given by the following
equation:
(10)

For random effects, the degrees of freedom,
vi, are given
by vi = n
1, where n is the number of
readings used to calculate s(qk).
The effective degrees of freedom of the standard
uncertainties based on systematic effects
can be assumed to be infinite in most cases.
Therefore, they will be reduced to 0,

and can be ignored, leaving only the random
contributions. Table I lists the t-distribution
values for a 95% confidence level.
Reporting of Result
The result of the measurement is y and may
be reported as follows: yA ± UA
for a confidence level of 95%, (k =
2).
If the measured result is compared
to a minimum or maximum limit (single-sided
limit), and the result including the uncertainty
falls within the limit, the confidence in a
passing result can be stated as 97.5%. The reason
is that the 5% probability of the error lying
outside the uncertainty range is symmetric,
so half, 2.5%, will also fall within the acceptable
range.
Applying Uncertainty Calculations
The data in this section are given to show
how uncertainty calculations are applied.
Where possible, the data come from actual
data sheets and laboratory measurements. Some
of the data were created either to illustrate
a particular point or to provide information
where none was available. Unless these data
exactly match a given EUT's measurement system
and simulator performance, they cannot be
used for its uncertainty calculations.
Simulator Tip Voltage. The following
is an example of uncertainty calculation for
ESD simulator tip voltage. Creating a table
listing all the probable sources of error
is the first step. An electrostatic voltmeter
measures the tip voltage. Although the electrostatic
voltmeter is the only measuring instrument
used here, it is not the only probable error
source. The resolution of the simulator voltage
display is another probable error source that
should be included in the table. It has also
been observed that in repeating these measurements
under the same conditions, the measured voltage
changes by a few volts each time the measurement
is taken. This variation will be included
in the table as a random component of uncertainty
(see Table II).
|
Contribution
|
Distribution
|
Uncertainty
|
Std. Deviation
|
Notes
|
| Electrostatic
voltmeter accuracy |
|
|
|
|
| Accuracy
of simulator voltage setting |
|
|
|
|
| Measurement
repeatibility |
|
|
|
|
| Table II. Table design
for tracking contributions and simulator
tip-voltage uncertainty at 2000 V. |
One can estimate the electrostatic voltmeter's
uncertainty range by first looking at the
manufacturer's data sheet. It has a rated
accuracy of ±1% over its entire measuring
range. According to the previous definitions,
rectangular distribution is applied to a manufacturer's
specification. Because the measured quantity
is in volts, one converts 1% to 20 V. The
standard deviation, defined as the semirange
limit value divided by the square root of
three, yields a value of .
The meter has recently been calibrated, and
the uncertainty value could possibly be reduced
if the report gives a lower value of uncertainty
or states a confidence factor in the reported
uncertainty (normal distribution divides the
uncertainty by 2 instead of (check 3). The
calibration report measures the accuracy as
±1%, and no confidence factor has been
stated. Therefore, the uncertainty and standard
deviation values remain 20/ .
Next, one must assess the uncertainty of
the simulator's voltage setting, which has
a continuously variable setting. The resolution
of the simulator's voltage display is 10 V,
which allows the tip voltage to vary as much
as 10 V for the same displayed voltage value.
Therefore, the semirange limit value is the
error range divided by 2 = ±5 V. Because
there is an equal probability of the actual
tip voltage being anywhere within this range,
rectangular distribution is applied. The resolution
of the display is the same over the entire
voltage range. Therefore, the uncertainty,
because of the display, will be highest at
low voltages. The lowest IEC test level is
2000 V and has been chosen for our calculation.
The last error source to evaluate is measurement
repeatability. While using the same instrumentation
and general setup will be necessary to minimize
random components, small variations in the
placement of the simulator relative to the
target and in the positioning of its grounding
cable relative to the target plane can still
cause random variations. The series of measurements
to establish the random component must include
these variations in positioning. Setting up
the simulator once and taking the series of
measurements will not take these variations
into account.
The setup should be taken apart and reestablished
between each measurement to include these
possible variations. However, the uncertainty
of the simulator's voltage setting should
not be counted twice. If the voltage setting
uncertainty has already been accounted for
elsewhere in the table, the setting should
not be changed between tests. A series of
10 measurements is taken, and the recorded
values are 2000, 2010, 2020, 2000, 1990, 1990,
2000, 1980, 2000, and 2010. The standard deviation
is calculated using the formula given in the
Random Components section, and its value is
11.6 V. Repeatability should also be checked
at other voltage levels; however, this step
has been skipped here. Table III shows all
of the uncertainty and standard deviation
values.
|
Contribution
|
Distribution
|
Uncertainty
|
Std. Deviation
|
Notes
|
| Electrostatic
voltmeter accuracy |
Rectangular
|
+1%
|
=20/  V
|
From manufacturer's
data sheet or calibration report
|
| Accuracy
of simulator voltage setting |
Rectangular
|
10 V or +0.25% at
2 kV
|
=5/
|
Can be ignored at
higher voltages
|
| Measurement
repeatibility |
Standard deviation
|
11.6 V
|
11.6 V
|
Calculated from a
series of 10 tests
|
| Table III. Simulator
tip-voltage uncertainty and estimated
uncertainty at 2000 V. |
Next, the combined uncertainty is calculated
by taking the square root of the sum of squares
of the individual uncertainty values. The
resulting combined uncertainty value is 17.3
V. One must now evaluate whether a significant
portion of the uncertainty is caused by random
components, as determined by whether or not
One finds that 16.6/11.6 is only 1.43, and
therefore we cannot simply use k =
2. It will be necessary to calculate the effective
degrees of freedom so that a value of k
based on the t-distribution may be determined.
Even if it were decided to make three measurements
when testing the simulator, thereby reducing
the standard deviation for the measurement
repeatability, the value would be increased
to only 2.48, which is still too low (see
Table IV).
|
Contribution
|
Distribution
|
Uncertainty
|
Std. Deviation
|
Notes
|
| Electrostatic
voltmeter accuracy |
Rectangular
|
+1%
|
=20/  V
|
From manufacturer's
data sheet or calibration report
|
| Accuracy
of simulator voltage setting |
Rectangular
|
10 V or +0.25% at
2 kV
|
=5/
|
Can be ignored at
higher voltages
|
| Measurement
repeatibility |
Standard deviation
|
11.6 V
|
11.6 V
|
Calculated from a
series of 10 tests
|
| Combined
uncertainty |
Normal
|
16.6 V
|
16.6 V
|
Combining the standard
deviation values above
|
| Expanded
uncertainty |
Normal
|
34.5 V= 2.08 x 16.6 V
|
N/A
|
(k=2) if uc
(y)=1.49, must calculate
veff
|
| Table IV. Simulator tip-voltage
uncertainty, expanded certainty at 2000
V. |
Using the values of combined uncertainty
and the standard deviation of the measurement
repeatability, the value of veff
is 37.7. The value of k is then determined
to be 2.08 by linear interpolation between
the veff
values of 20 and 50 found in Table I.
When reporting the measured result of simulator
tip voltage at 2000 V, the result y
can be stated as y V ± 34.5 V
with a confidence level of 95%. By converting
the voltage back into a percentage, the uncertainty
could also be stated as ±1.7% with a
95% confidence level.
Simulator Rise Time (Contact Mode).
The next step is to calculate the uncertainty
for contact-mode simulator rise-time measurements.
The uncertainty components for air-discharge
measurements are identical. However, the uncertainty
due to the simulator will be greatly increased
because of the effects of the arc. Obvious
sources of error include the oscilloscope's
time measurement accuracy and the simulator
repeatability. The rise time uncertainty,
however, is much more difficult to estimate
than the first example because of some special
considerations.
First, the simulator's rise time is very
fast (0.71 ns for IEC 1000-4-2compliant
simulators). If the measurement system's rise
time is not much greater than the simulator's
rise time (<100 ps is required for a 1%
error when measuring 0.7 ns), the measured
result can be strongly influenced by the measurement
system's rise time. Storage oscilloscopes
with rise times <100 ps are uncommon and
quite expensive. Most digital oscilloscopes
used for ESD measurements have a rise time
in the neighborhood of 270350 ps, and
it is the oscilloscope's rise time that usually
dominates the system's rise time. The influence
of the system's rise time must be included
as a probable source of error.
Second, it can be seen from the preceding
discussion that the influence of the measurement
system rise time will vary over the range
of measured results. To simplify the calculation,
take the worst-case situation for the range
that is at the lower limit as set by the IEC
simulator standard, 0.7 ns, where the measurement
system rise time has the greatest influence.
Third, ESD rise time is the time required
for the leading edge of the current waveform
to increase from 10 to 90% of the peak amplitude.
Because these two points are referenced to
the peak amplitude, we are not concerned with
amplitude errors as long as they remain invariant
with frequency. Steps must be taken to minimize
any reflections or errors that do vary with
frequency, as they can influence the observed
rise time. The oscilloscope, attenuators,
cables, and target are all probable sources
of this type of error. A tutorial on how to
measure ESD pulses is beyond the scope of
this article; however, such resources are
available.24
The data sheet rating for the 1-GHzbandwidth,
5GS/s, oscilloscope's time measurement accuracy
is <50 ps. A series of 10 rise-time measurements
is taken to assess simulator repeatability,
and the recorded values are 804, 721, 768,
778, 772, 753, 697, 731, 759, and 742 ps.
The standard deviation is calculated using
the formula given in the Random Components
section, and its value is 30.9 ps.
The influence of the system's rise time is
the most difficult quantity to assess. As
previously stated, the oscilloscope is usually
the most limiting factor in the system's rise
time. In a system with a Gaussian impulse
response (a smooth roll-off of 20 dB/decade
above the 3dB point), the rise time
is given by rise_time = 0.35/bandwidth.
The system's rise time is composed
of the individual rise times of all the components
in the system. In a Gaussian system with m
components, the system rise time is given by
(11)

where rise time (n) is the rise time
of each of the system components such as oscilloscope
rise time, target rise time, attenuator rise
time, etc.
For a 1-GHz-bandwidthrated oscilloscope,
the calculated rise time is 350 ps. Most oscilloscopes,
however, only approximate this behavior and
exhibit a pulse response with a faster rise
time and some overshoot. The observed rise
time typically approximates a 30% higher bandwidth.
In our example of a 1-GHz oscilloscope, this
would mean a rise time of 269 ps.
The ideal situation for determining the uncertainty
would be to measure the actual rise time of
the entire target/attenuators/ cable/oscilloscope
chain. This is accomplished by driving the
measurement system with a pulse of known rise
time (100 ps or less is ideal as it will have
a minimal influence on the measured result).
The measured rise time can be corrected for
the contributions of the entire measurement
chain, thereby reducing the uncertainty to
the accuracy of the system's rise-time measurement.
The corrected rise time of the measured result
is given by the following equation:
(12)

Driving a fast-rise-time pulse into the ESD
measurement target requires a special coaxial
target adapter. For this reason it may not
be possible to measure the entire chain's
rise time. Instead, the rated rise time of
the target can be used, and the rise time
of the attenuators/cable/oscilloscope measured.
In this case, the uncertainty of the measurement
chain will comprise two components: the tolerance
on the target rise time and the uncertainty
of the attenuators/cable/oscilloscope chain's
rise-time measurement. Rise-time error sources
contribute to the measurement error as the
square root of the sum of squares.
Our example has neither of these specialized
instruments and attempts to estimate the uncertainty
without the benefit of these measurements.
First, let's estimate the worst-case system
rise time. We'll assume that our 1-GHz oscilloscope
minimally meets its rise-time specification
of <350 ps. In all other respects, the
measurement system uses high-bandwidth/fast-rise-time
components. The cable is manufactured and
tested for >20 GHz use. The two attenuators
have an 18-GHz bandwidth, and the target has
a very flat bandwidth (the target is the new
EOS/ESD design, not the IEC target) with a
rated rise time of <35 ps. An estimated
worst-case contribution of 20 ps for the cable
is included:
(13)

Next, the influence of the worst-case system
rise time on the observed rise time must be
calculated. For our ESD simulator with 700-ps
rise time, this would produce
(14)

Now let's look at the best case. It is expected
that the bandwidth of our 1-GHz oscilloscope
is nominally 30% high. Let's estimate that
it is actually 30% above that, with a rise
time of 207 ps. To simplify matters, assume
the other components are zero as they have
only a small influence on the measured result.
The observed rise time for the 700-ps simulator
is now as follows:
(15)

The error range is 784 730
= 54 ps so the semirange limit value is 27 ps.
It turns out that the uncertainty due to the
estimated range of system rise times is the
smallest of the three contributions to the combined
uncertainty, and that it is not unreasonable
to estimate the uncertainty of the measurement
system rise time in this way. The rise-time
measured value must be corrected for the nominal
system rise time by the following formula:
(16)

The uncertainty and standard deviation values
are shown in Table V.
The combined uncertainty is 45.1 ps. Now
one must evaluate whether a significant portion
of the uncertainty is caused by random components,
as determined by whether or not

We find that 45.1/30.9 is only 1.46, well
below 3; therefore, we cannot simply use k
= 2. It is necessary to calculate the effective
degrees of freedom so that a value of k
based on the t-distribution may be determined.
Using the values of combined uncertainty and
the standard deviation of the measurement
repeatability, the value of veff
is 40.8. The value of k is then determined
to be 2.07 by linear interpolation between
the veff values
of 20 and 50 found in Table I.
When reporting the measured result of simulator
rise time, the result y can be stated
as y ps ± 45 ps with a confidence
level of 95%. The midrange value of the estimated
system rise time is used to correct the observed
result for the system rise time and produce
the reported value.
The examples provided in this section
have shown how to estimate the uncertainty for
just two of the many parameters normally measured
in evaluating a simulator's operation. The general
procedure for estimating uncertainty is the
same for any measured parameter. The method
for creating a table, listing the contributions,
and calculating the standard deviation, combined
uncertainty, and expanded uncertainty are the
same. All that can change is how the uncertainty
is estimated for each of the contributions,
where some contributions are much more difficult
to estimate than others. The tables are created
and the calculations are performed as in the
examples here.
Uncertainty ContributionsAll Parameters
Table VI lists the simulator parameters and
possible contributions to each parameter.
The contributions are divided into categories
of tip voltage, target-chain dc attenuation,
time domain, and random contributions. The
following sections explain some of the issues
that may be encountered in estimating each
of these contributions.
Tip Voltage
Amplitude-based measurements are affected
by the simulator's tip voltage. Possible contributions
to the tip-voltage uncertainty are voltmeter
accuracy and simulator voltage-setting resolution,
as well as tip-voltage repeatability, which
is discussed later in the Random Effects section.
Voltmeter Accuracy. Simulator tip
voltage is measured either with an electrostatic
voltmeter or with a voltmeter plus a high-voltage
attenuator. Electrostatic voltmeters typically
have several ranges, and the accuracy may
not be the same for all of them. If a voltmeter
and an attenuator are used, the accuracy of
both must be considered. The uncertainty can
be taken from the manufacturer's specification
or from a calibration report. With high-voltage
attenuators, it is necessary to consider the
voltage coefficient of resistance in addition
to the normal tolerances. The uncertainty
values for the scales used during simulator
measurements should be selected. If the manufacturer's
specification is used, it is considered rectangular
distribution and the semirange limit value
is divided by. But if a measured value is
taken from a calibration report, it is considered
normal distribution and the semirange limit
value is divided by 2 (or a larger number
depending on the stated level of confidence).
Division by 2 in the second instance is most
likely to produce a lower value of standard
deviation.
Simulator Voltage Setting Resolution.
The uncertainty of the simulator's voltage setting
may be derived from one of two sources: the
resolution of the display or the resolution
of the voltage setting. If the voltage setting
is continuously variable, the resolution of
the simulator's voltage display will dictate
the uncertainty. For example, if the resolution
of the simulator's voltage display is 10 V,
this allows the setting to be high or low by
as much as 10 V for the same displayed voltage
value. Therefore, the accuracy of the simulator
voltage setting is ±10 V. Because there
is an equal probability of the actual tip voltage
being anywhere within this range, rectangular
distribution is applied. If the resolution of
the voltage setting is incremental and is smaller
than the resolution of the display, the uncertainty
will be the resolution of the voltage setting.
Rectangular distribution is applied because
there is an equal probability of the actual
tip voltage being anywhere within the resolution
of the setting.
Target- and Measurement-Chain Dc Attenuation
A dc current source is used to measure the
combined dc attenuation of the target and
measurement chain (see Figures 1 and 2). Measuring
the entire chain simplifies the task of estimating
the uncertainty as compared with evaluating
the target, attenuators, cables, and oscilloscope
separately. It is important that the current
applied to the target does not exceed its
average power dissipation rating since damage
to the target may result. The uncertainty
of the current source output contributes directly
to the attenuation uncertainty, as well as
to the oscilloscope's vertical accuracy. If
the dc attenuation measurement is made with
the same vertical scale as the simulator measurement,
the oscilloscope accuracy has been taken into
account and only the vertical resolution must
be considered.
 |
|
Figure 1. Block diagam of measurement
chain.
|
 |
| Figure 2. Measuring target-chain
dc attenuation. |
Current Source Accuracy. The uncertainty
of the current source output may be derived
from an internal meter, an external measurement,
or the current setting resolution. The uncertainty
can be taken from the manufacturer's specification
or from a calibration report. If the manufacturer's
specification is used, it is considered rectangular
distribution and the semirange limit value
is divided by . If a measured value is taken
from a calibration report, it is considered
normal distribution and the semirange limit
value is divided by 2 (or a larger number
depending on the stated level of confidence).
Division by 2 in the second instance will
most likely produce a lower value of standard
deviation.
Oscilloscope Vertical Accuracy. The
oscilloscope's vertical measurement uncertainty
can be taken from the manufacturer's specification
or from a calibration report. The uncertainty
values for the scales used during simulator
measurements should be selected. Additionally,
the manufacturer's specification is usually
valid for full-scale readings and the uncertainty
will be higher for waveforms that encompass
a smaller portion of the vertical scale. If
the manufacturer's specification is used, it
is considered rectangular distribution and the
semirange limit value is divided by  . If a measured
value is taken from a calibration report, it
is considered normal distribution and the semirange
limit value is divided by 2 (or a larger number
depending on the stated level of confidence).
The measured values may vary since the settings
(i.e., V/various factors) will be different.
The measured value will probably produce a lower
value of standard deviation.
Time Domain
Time domain uncertainty contributions include
the oscilloscope's time measurement accuracy
and the measurement chain's rise time. The
time domain response of the measurement chain
is usually dominated by the oscilloscope.
Other elements, such as the target, attenuators,
and cables, typically make only small contributions
to the system response if they have been selected
properly. A target with a flat frequency response
and high-bandwidth attenuators are required.
Long cables, poor attenuators, and the IEC-specified
target can contribute significant errors.
Recommended performance specifications for
these components are available.5
Oscilloscope Time Measurement Accuracy.
The oscilloscope time measurement uncertainty
can be taken from the manufacturer's specification
or from a calibration report. For digital
oscilloscopes, the time measurement accuracy
is usually a fraction of the sampling interval.
If the manufacturer's specification is used,
it is considered rectangular distribution
and the semirange limit value is divided by .
If a measured value is taken from a calibration
report, it is considered normal distribution
and the semirange limit value is divided by
2 (or a larger number depending on the stated
level of confidence). The measured value will
probably produce a lower value of standard
deviation.
Measurement System Rise Time and Bandwidth.
The measurement system rise time, a function
of bandwidth, contributes to amplitude loss
and reduced rise time or di/dt in the observed
waveform. Rise time and di/dt measurements
can be corrected for the system rise time
(measured or estimated) as outlined in the
Simulator Rise Time section. Amplitude losses
are more difficult to estimate, because the
actual simulator waveform is complex, with
a narrow peak, rather than a square pulse.
It varies from shot to shot and remains unknown,
as we can only observe the measured waveform.
The amplitude losses can be estimated
by mathematical simulation using idealized simulator
waveforms and a Gaussian filter to represent
the system bandwidth. The difference between
the idealized waveform and the output of the
filter can then be measured to obtain the loss
in peak current. A Gaussian filter has no phase
shift and an amplitude response of
|H(j )|
= exp[0.3466( / c)2]
(17)
The Gaussian filter is a good approximation
of oscilloscope performance. Equations for
the simulator waveforms are provided.6
These idealized waveforms have smooth, rounded
peaks and no oscillations. Simulators that
have sharper peaks will exhibit higher losses.
Overall, the amplitude losses are very low,
approximately 1% or even less for the 1-GHzrated
oscilloscope (see Table VII).
|
System
Bandwidth
|
Idealized IEC 1000-4-2
Simulator Waveform:
760 ps rise time, 3.75 A/kV
|
Proposed Faster-Rise
Simulator Wavefrom:
500 ps rise time, 4.75 A/kV
|
| 1.3 GHz |
99.25% |
98.95% |
| 2.6 GHz |
99.86% |
99.79% |
| 3.9 GHz |
99.94% |
99.90% |
|
Actual
oscilloscope performance is typically
30% higher than rating. Percentage of
true peak.
|
| Table VII. Peak amplitude
loss as a function of system bandwidth.
|
Random Effects
Possible random effects include tip-voltage
repeatability, dc amplitude measurement repeatability,
rise-time repeatability, and peak amplitude
repeatability. The uncertainty associated
with random effects is evaluated by statistical
techniques from a series of repeated measurements,
as explained in the Random Components section.
The estimation is only valid if the EUT measurements
are made under exactly the same conditions.
Using the same instrumentation and general
setup will be necessary to minimize random
components. However, small variations in the
placement of the simulator relative to the
target and in the positioning of its grounding
cable relative to the target plane can still
cause random variations.
The series of measurements to establish the
random component must include these variations
in positioning. Setting up the simulator once
and taking the series of measurements will
not take these variations into account. The
setup should be taken apart and reestablished
between each measurement to include these
possible variations. The uncertainty of the
simulator's voltage setting, however, should
not be counted twice. If the voltage setting
uncertainty has been accounted for elsewhere
in the table, this setting should not be changed
between tests.
Shot-to-shot variations in ESD simulators
will most likely be the largest contribution
to the measurement uncertainty, especially
for air discharge. The standard deviation
of the variations, even in contact mode, can
be several percentage points or more, and
can vary considerably across the voltage range
because of the operation of the contact-mode
relay. These shot-to-shot variations can lead
to high values of expanded uncertainty that
may approach the specification limit (see
Table VIII).
According to NIS81, making a statement of compliance
with a standard requires that the measured result
plus or minus the expanded uncertainty value
falls within the specified limits. This has
a dramatic effect on the compliance range. For
example, the IEC 1000-4-2 specification for
peak amplitude is ±10%. If the expanded
uncertainty is 4.82%, as in the example, the
compliance range is ±10 ±4.82
= ±5.18%, effectively cutting the compliance
range in half. Actual values of expanded uncertainty
could be higher, leading to an even smaller
compliance range.
Conclusion
The amplitude measurement uncertainty
for ESD simulators can be very significant.
The values chosen for this example are low and
represent best-case performance. Actual values
may be much greater, especially at low-voltage
levels where the contact-mode relay contributes
significantly to the shot-to-shot variation.
It is essential to identify and minimize each
contribution to the measurement uncertainty
to obtain low levels of expanded uncertainty
and to make a statement of compliance with the
simulator standard.
References
-
NAMAS NIS81,
"The Treatment of Uncertainty in EMC Measurements,"
1st ed., May 1994.
-
EOS/ESD Association
WG14, "Metrology & Methodology of System
Level ESD Testing," June 1998.
-
EOS/ESD Association
WG14, "System Level Electrostatic (ESD)
Simulator Verification Standardpart 1Discharge
Current," ESD-WIP DS14.0-1998.
-
David Pommerenke, "Current
Target and Contact Mode Simulator Specifications,"
EOS/ESD Working Group 14, April 12, 1999.
-
"IEC 1000-4-2, 1st edition:
1995, "Electromagnetic Compatibility (EMC)
part 4: Testing and Measurement Techniquessect.
2: Electrostatic Discharge Immunity Test
Basic EMC Publication."
-
"IEEE Standards Reference
Infobase on CD-ROM," SE105, 1-55937-933-2,
1997.
Greg Senko is sales manager at Schaffner
EMC (Kingston, NH). He can be contacted by
e-mail at gsenko@schaffner.com.
|
|