Planar Separation Effects: Pacemakers and Wireless
Phones
By F. Hank Grant and Robert E. Schlegel
Center for the Study of Wireless EMC, University
of Oklahoma (Norman, OK)
This article presents the results of experimental
work undertaken to determine the minimum separation
distance required to eliminate electromagnetic interference
(EMI) between wireless phones and cardiac pacemakers.
Previous studies on interaction between these devices
have reported maximum interaction distances on the
order of 6 in. or greater.1 However,
these figures tend, in part, to reflect the methodology
used to measure separation distance. The present
study employed pacemaker height testing, an alternative
measurement method that yielded data suggesting
maximum interaction distances are significantly
lower than 6 in.
The testing described below represents the second
phase of a research project begun earlier by the
Center for the Study of Wireless Electromagnetic
Compatibility at the University of Oklahoma (Norman,
OK). The objectives of the previous phase of testing
were to verify reported EMI between wireless phones
and pacemakers, to more precisely classify the modes
of pacemaker interaction, and to identify the wireless-phone
operating modes that produce the interaction.2
All cases examined in the follow-up study had been
shown to exhibit interaction during the first phase
of testing.
Testing was conducted in a closed, electromagnetically
shielded room at the Lucent Technologies Inc. test
facility in Oklahoma City, OK. Functioning of the
heart and pacemaker system was simulated through
the use of a torso simulator and various electronic
equipment that generated and monitored electrical
signals. The test equipment consisted of the following:
-
A torso simulator with saline bath and supports
for the pacemaker, pacemaker leads, and wireless
phones.
-
Signal-monitoring equipment for acquiring the
waveforms from the pacemaker and the electrocardiographic
(ECG) signal when injected.
-
ECG signal injection equipment.
-
-
-
One wireless phone base-station
simulator.
Testing was conducted under worst-case conditions.
Phones were set at their highest power, and pacemakers
were set to the greatest sensitivity permitted for
each unit. In all tests, the pacemaker case and
the phone keypad were in horizontal parallel planes,
with the pacemaker 0.5 cm below the surface of the
saline solution. Figure 1 shows the overall test
setup.
 |
| Figure 1. Experimental apparatus. |
The test factors were divided into three categories:
pacemaker variables, wireless phone variables, and
separation distance. The pacemaker variables consisted
of the pacemaker model, pacemaker mode, lead polarity
configuration, and the presence or absence of an
injected ECG signal. Of the six pacemakers tested,
two were single-chamber units and four were dual-chamber
units.
The phone variables consisted of the phone technology
or model, phone test mode, and phone orientation
with respect to pacemaker lead alignment. Again,
only the phone technologies that interacted with
pacemakers in the first phase of the study were
tested in the second phase. The eight digital wireless
phones that were used in the testing are shown in
Table 1. Five U.S. manufacturers provided the phones.
|
Name
|
Technology Standard
|
Number Tested
|
|
CDMA
|
Spread Spectrum, IS-95
|
1
|
|
TDMA-50 Hz
|
Dual-mode digital/analog, IS-54/55
|
4
|
|
PCS 1900
|
TDMA-217 Hz, J-STD-007
|
2
|
|
TDMA-22 Hz
|
IDEN
|
1
|
|
TDMA-11Hz
|
8A/XB
|
1
|
| Table I. Wireless phone technologies
tested. |
Except for TDMA-50 Hz phones, all phones were tested
in an open-loop transmit mode. This test mode produced
the typical digital pulsing format of the phone
technology without communicating with an active
cell site or base-station simulator. For some models,
the phone's hardware was modified. For other models,
keypad programming was used to configure the phone
in the full-power transmit mode. TDMA-50 Hz phones
were used while communicating with an NADC base-station
simulator (HP8920A RF Communications Test Set with
an HP83201A dual-mode cellular adapter). The TDMA-50
Hz phones were used in both ringing and talk-back
communication modes. The ringing mode is achieved
by registration of the phone followed by paging.
Once initiated, the ringing continues for more than
a minute while pacemaker testing proceeds. In the
talk-back (loop-back) mode, a short message (approximately
4 seconds) is spoken into the handset. The digitally
encoded form of the message is transmitted to the
base station, which retransmits it continuously
to the handset.
The study examined four relative orientations of
the pacemaker and phone. The pacemaker was oriented
in the tank such that the leads exited the header
to the right (east) when viewed from above, with
the long axis of the tank going from left to right
(west to east). The orientation of the phone antenna
from base to tip was either south to north (90°),
east to west (180°), northeast to southwest
(225°), or southeast to northwest (135°).
Four additional orientations (0°, 45°,
270°, 315°) were examined during the first
phase, but produced redundant results. Therefore,
the orientations of 0°, 45°, 270°,
and 315° were eliminated from future tests.
Various combinations of these factors defined a
single test run (for example, pacemaker 01, unipolar,
VVT [ventricle sensing, ventricle pacing, ventricle
pulse delivered after sensing], with injected ECG
signal, in conjunction with phone Y, talk-back mode,
or at a 90° orientation). The run itself consisted
of individual tests at appropriate grid points.
All runs in which interaction was observed in the
first phase of the study were tested.
To determine the distance at which interaction
between the wireless phone and the pacemaker no
longer occurs, the phone was raised in discrete
steps of approximately 1/8
in. starting from the lowest height, until the interaction
stopped. Testing then proceeded by reducing the
height by one discrete step. This height was the
maximum vertical planar separation distance at which
interaction occurred. Planar separation distance
is the distance between the following two planes:
the plane representing the top of the saline solution,
approximately 1 cm above the pacemaker; and the
parallel test plane defined by the top of a plastic
grid upon which the phone is suspended and over
which it is moved to test interaction. This parallel
test plane is separated from the saline solution
at various increasing distances until interaction
ceases. The surrounding grid where interaction occurred
in the first phase was tested thoroughly, and new
interaction points were identified.
A second interaction distance measure, called the
Euclidean distance, is shown in Figure 1. This distance
was obtained by identifying the farthest x-y grid
point from the pacemaker header and the base of
the phone antenna at which an interaction event
was observed. This x-y grid point was converted
to a Euclidean distance and used as a measure of
interaction distance susceptibility. This distance
measure was used in the first phase.
Figure 2 shows the overall results of the second
phase of the study. Interaction was minimal at planar
separation distances less than 50% of the AAMI-recommended
separation distance of 6 in. Additionally, half
of the pacemakers tested did not interact at more
than the minimal separation distance.
 |
| Figure 2. Height separation effects: maximum
planar distance between the pacemaker header
and B antenna of the wireless phone. The height
of the bars indicates the distance. The two-dimensional
rectangles indicate that there was no interaction
beyond the zero separation distance. |
This study was conducted with each phone operating
at its highest power level and each pacemaker programmed
to its maximum sensitivity setting. All tests were
conducted with the phone in close proximity to the
pacemaker, representing a phone being carried in
clothing pockets or held adjacent to the body, in
the vicinity of an implanted pacemaker. Caution
must be exercised in using these results to directly
contrast one phone technology with another due to
differences in the frequency bands used and possible
differences in the implementation of these technologies.
The following general conclusions can be drawn
from the study. The planar separation distance measured
in this study is significantly less than the Euclidean
distance measured in the first phase of the study.
The maximum planar separation distance obtained
was 3 in., as opposed to a maximum Euclidean distance
of 7.6 in. The study also showed that the maximum
field strength may be concentrated over the case,
the antenna of the phone, or both. Approximately
50% of the phones tested had the maximum field strength
concentrated over the phone body.
Generally, the study results are encouraging when
compared with the recommended guideline of a 6-in.
separation between wireless phones and pacemakers.
Additional security can be derived from the fact
that many medical device firms that make pacemakers
now install feed-through capacitors to filter radio-frequency
emissions that might affect the pacemaker's operation.
Such filtering virtually eliminates EMI.
1. V Barbaro, et al., "GSM Cellular Phone Interference
with Implantable Pacemakers: In Vitro and In Vivo
Observations," (presented at the BEMS conference,
Stockholm, 1994).