Using IEC 60601-1-2 for Testing Medical Devices
Instrument Specialties Company, Inc. (Delaware
Water Gap, PA)
As the electromagnetic spectrum becomes increasingly
congested and electronic devices proliferate, ensuring
electromagnetic compatibility (EMC) in electrical
and electronic equipment becomes a critical issue.
Ensuring EMC is a vital issue for companies producing
electrical and electronic medical devices. These
types of devices must perform as intended and not
interfere with other equipment, or the results could
be catastrophic. For medical manufacturers, making
sure that devices meet EMC standards is not only
a marketing necessity, but also a societal concern.
There are several EMC specifications, but none
of them alone will provide appropriate testing guidelines
for specialized medical devices. To reduce confusion
and ensure that devices will be properly tested,
medical manufacturers should use an existing standard
as a guide for developing a customized testing plan.
The International Electrotechnical Commission (IEC)
is a worldwide body that promotes international
standardization in electronics. In 1993 it released
the 60601-1-2 standard, "Medical Electrical EquipmentPart
1: General Requirements for Safety, Amendment No.
2. Collateral Standard: Electromagnetic Compatibility
Requirements and Tests."1
The IEC 60601 standard offers a solid basis for
medical device testing. Although they are relatively
new, the IEC 60601-1-2 requirements have quickly
become recognized throughout the world and are instrumental
in testing to the European Medical Devices Directive.
Organizations such as the American National Standards
Institute (ANSI) use the IEC 60601 standard as a
basis for their own requirements.
This document specifies acceptable levels for immunity
and refers to other documents to specify emission
levels. However, these levels may not be strict
enough to ensure that equipment will operate as
intended. Manufacturers should use the IEC specifications
as a guide but tailor them to produce product-specific
limits.
Test Specifications and Limits
The IEC 60601-1-2 standard specifies test limits
for emissions, immunity, electrostatic discharge
(ESD), radiated RF electromagnetic fields, bursts,
and surges.
Emissions. Equipment should comply with
the conducted and radiated emissions requirements
of the International Special Committee on Radio
Interference (CISPR). Classification of equipment
for this purpose is based on intended use and determined
by the manufacturer.
Equipment may be tested for emissions at a standard
test site, which would include a turntable and ground
plane, and have known attenuation curves. Equipment
may also be tested after it has been installed on
the users' premises. It is recognized that medical
equipment may have unique installation considerations
and that type-testing of the installation is the
only practical solution to demonstrate compliance
to the requirements.
Manufacturers should refer to CISPR 11 for the
appropriate requirements and amplitude levels once
the class of equipment and test location have been
determined.2
Currently, there are no requirements for low-frequency
emissions, harmonic distortion, and voltage fluctuations,
but some equipment that operate in an intermittent
mode will have to meet specific variations of the
CISPR 14 Click requirements.3
CISPR 11 covers a frequency range from 150 kHz
to 18 GHz. Conducted emissions for low- and medium-voltage
power mains (100415 V) are performed from
150 kHz to 30 MHz. The frequency range for radiated
emissions is from 150 kHz to 18 GHz. Depending on
the class and use of the equipment, various frequency
ranges may be defined. Only the magnetic component
of the radiated field is measured from 150 kHz to
30 MHz. Above 30 MHz, both the vertical and horizontal
components of the field must be measured.
Amplitude limits in general are established to
protect the public broadcast services, not for equipment
that may have to operate in close proximity to sensitive
medical equipment.
The specification also refers to frequencies designated
by the International Telecommunication Union: 2450
MHz for industrial, 5800 MHz for scientific, and
24,125 MHz for medical equipment.
Immunity. General immunity requirements
are specified in IEC 60601-1-2. Test levels are
given and test methods are based on the IEC 801
series of immunity requirements. If lower limits
are justified, accompanying documents should explain
this and describe any action that will, as a consequence,
be taken by the installer or user.
Accompanying documents should include guidelines
for avoiding or identifying and resolving adverse
electromagnetic effects. If the use of the equipment
is restricted because of its electromagnetic characteristics,
relevant restrictions should be described in the
accompanying documents.
Compliance with the requirements should be checked
by verifying that the equipment continues to perform
its intended functions as specified by the manufacturer
or fails without creating a safety hazard.
ESD. Equipment should comply with the current
edition of IEC 801-2.4 A limit of 3 kV
applies for direct contact discharge to all conductive
accessible parts and coupling planes. A limit of
8 kV applies for air discharge to nonconductive
accessible parts.
Radiated Radio-frequency Electromagnetic Fields.
Equipment should comply with the IEC 801-3 requirements,
which are being updated.5 A limit of
3 V/m should be used over a frequency range of 26
MHz to 1 GHz. Other levels apply to equipment used
in shielded locations, such as x-ray and MRI facilities.
The 3-V/m requirement is decreased in proportion
to the increasing shielding effectiveness of the
location.
There are provisions for amplitude modulation of
the signal, depending upon the passband of the equipment
under test (EUT). If the EUT does not have a passband,
the signal should be amplitude modulated at 1 kHz.
Bursts. Test methods and instruments specified
in IEC 801-4 should be followed.6 A 1-kV
level applies to equipment connected to the power
line with a plug. For permanently installed equipment,
a level of 2 kV applies. Interconnecting lines longer
than 3 m should be able to withstand a 0.5-kV surge.
Surges. Test methods and instruments specified
in IEC 801-5, which is currently still under consideration,
should be followed.7 Power lines should
meet levels of 1 kV for differential mode and 2
kV for common mode. Signal lines need not be tested,
and telecom lines are covered by other standards.
Ring wave and damped sinusoid tests are not applicable.
There are future provisions for voltage dips, short
interruptions, and voltage variations on power lines,
as well as for conducted immunity above 9 kHz and
magnetic field immunity.
Manufacturers of electrical and electronic equipment
for any use are recognizing the need for specifications
that ensure compatibility among equipment. Medical
electronics manufacturers also recognize that such
generic standards are not necessarily appropriate;
they may be too severe, or, even worse, not severe
enough to protect their products. To lessen confusion
and to ensure that test specifications will be appropriate,
medical manufacturers should use an existing document
such as IEC 60601-1-2 as a basis for creating their
own product-specific standards.
1."Medical Electrical Equipment Part 1: General
Requirements for Safety, Amendment No. 2. Collateral
Standard: Electromagnetic CompatibilityRequirements
and Tests," Geneva, IEC, Bureau Central de la Commission
Electrotechnique, 1st ed, 1993.
2.International Special Committee on Radio
Interference, CISPR Publication 11, "Limits and
Methods of Measurement of Radio Interference Characteristics
of Industrial, Scientific and Medical (ISM) Radio
Frequency Equipment (Excluding Surgical Diathermy
Apparatus)," Geneva, IEC, 2nd ed, 1990.
3.International Special Committee on Radio
Interference, CISPR Publication 14, "Limits and
Methods of Measurements of Radio Interference Characteristics
of Household Electrical Appliances, Portable Tools
and Similar Electrical Apparatus," Geneva, IEC,
2nd ed, 1985.
4.IEC 801-2, "Electromagnetic Compatibility
for Industrial-Process Measurement and Control Equipment,
Part 2: Electrostatic Discharge Requirements," Geneva,
IEC, 2nd ed, 1991.
5.IEC 801-3, "Electromagnetic Compatibility
for Industrial-Process Measurement and Control Equipment,
Part 3: Radiated Electromagnetic Field Requirements,"
Geneva, IEC, 1st ed, 1984, 3rd impression, 1991.
6.IEC 801-4, "Electromagnetic Compatibility
for Industrial-Process Measurement and Control Equipment,
Part 4: Electrical Fast Transient/Burst Requirements,"
Geneva, IEC, 1st ed, 1988.
7.IEC 801-5, "Electromagnetic Compatibility
for Industrial-Process Measurement and Control Equipment,
Part 5: Surge Immunity Requirements," draft, Geneva,
IEC, July 1992.