Susceptibility Testing Requirements for Medical
Systems
MIKES Product Service GmbH (Strasskirchen, Germany)
To be marketable in Europe, medical electronic
equipment or systems must comply with the Medical
Devices Directive (MDD) regarding electromagnetic
compatibility (EMC) requirements. The product standard
for medical devices, EN 60601-1-2, has been harmonized
with the MDD. EN 60601-1-2 is identical to IEC 60601-1-2
and is the collateral standard to IEC 60601-1, which
is the general safety standard for medical electrical
equipment. IEC 60601-1-2, "Medical Electrical Equipment;
Part 1: General Requirements for Safety; 2: Collateral
Standard; Electromagnetic CompatibilityRequirements
and Tests," was published in April 1993. EN 60601-1-2
was published in November 1993, and it was harmonized
with the EMC Directive in August 1995 and with the
MDD in September 1995. This article discusses the
susceptibility tests required to comply with these
standards. It also discusses modifications expected
for the upcoming revision of IEC 60601-1-2.
A significant difference between the European EMC
standards and the medical device standards (such
as EN 60601-1-2) is the performance criteria. For
example, the immunity criteria for the collateral
standard are based on safety. In contrast, all other
EMC standards require evaluation of performance
as defined by the manufacturer.
According to EN 60601-1-2:1993, the immunity criteria
for medical products are defined as "equipment and/or
system continues to perform its intended function
as specified by the manufacturer or fails without
creating a safety hazard." Safety hazard is defined
by IEC 60601-1 as a "potentially detrimental effect
on the patient, other persons, animals, or the surroundings
arising directly from equipment." This definition
places a great deal of responsibility on the manufacturer
and testing laboratory. For example, it is possible
for a device to fail safely and still create a safety
hazard by its unavailability, such as a diagnostic
system that either fails to start or can be interrupted.
The standard does not require manufacturers to inform
users of such hazards. Although an accredited test
laboratory is likely to identify such a malfunction
in its test report, the manufacturer is not required
to state it in the user manual. Other examples also
illustrate a gap in EN 60601-1-2:1993 for defining
pass/fail criteria (i.e., safety hazard). This problem
has forced Working Group 13 of IEC Technical Committee
62 to define performance criteria more clearly.
EN 60601-1-2 is a generic standard designed to
address many different devices. Some medical products,
such as systems that measure human physiological
signals, require specific immunity levels. Based
on the low levels of these signals, it was necessary
for the standard to allow for lower immunity levels,
provided that the manufacturer instructed users
about any necessary actions to be taken because
of the lower test levels. This will be discussed
in more detail later.
EN 60601-1-2 provides requirements for emissions
and immunity. The required emission test is EN 55011
(CISPR 11) Group I or II, Class A or B. Immunity
test requirements are provided in IEC 801 and EN
61000-4-3, which are the so-called basic standards.
Specific tests are explained in the following standards:
-
IEC 801-2:1991, for electrostatic discharge
(ESD).
-
EN 61000-4-3 and IEC 801-3:1984, for
radiated immunity.
-
IEC 801-4:1984, for bursts and electrical fast
transients (EFTs).
-
IEC 801-5:1984, for surges.
IEC 801-2:1991ESD. This standard requires
contact- and air-discharge testing. EN 60601-1-2
set up a requirement of ±3 kV for contact discharge
and ±8 kV for air discharge since rise times
and spectrums will differ as test levels change.
The reliability of the air discharge presents another
problem: temperature, humidity, and the approach
speed of the loaded ESD tip can all influence the
test. For this reason, contact discharge was included
in IEC 801-2:1991, whereas only air discharge had
been required in the 1984 version.
The test setup requires a horizontal coupling plane
(HCP) of 1.6 x 0.8 m placed on a 0.8-m-high wooden
table. The equipment under test (EUT) and cables
must be isolated from the coupling plane by a 0.5-mm-thick
insulation support. For indirect discharge, a vertical
coupling plane (VCP) of 0.5 x 0.5 m must be used.
The VCP must be parallel to the EUT at a distance
of 0.1 m. Discharges must be applied to different
positions of the coupling plane so that the four
faces of the EUT can be completely illuminated.
Where coupling planes are specified, they must be
constructed from the same material (copper or aluminum)
and same thickness (minimum 0.25 mm x 1 m2)
as that of the ground reference plane (GRP). The
HCP and VCP must be connected to the GRP via a cable
with a 470-k
resistor located at each end.
For both indirect and direct discharge tests, 10
discharges of each polarity must be applied. Discharges
must be applied directly only to those points and
surfaces of the EUT accessible to personnel during
normal use. This includes areas accessible to maintenance
personnel. For contact discharges, the discharge
electrode tip must touch the EUT before the discharge
switch is operated. For air discharges, the round
tip of the electrode must approach the EUT as quickly
as possible without causing mechanical damage. After
each discharge, the ESD generator must be removed
from the EUT so that it can be retriggered. This
procedure must be repeated until all discharges
are completed.
EN 61000-4-3Radiated Immunity. This is
the relevant basic standard for simulating the interference
of transmitted electromagnetic waves to test for
radiated immunity. The collateral standard requires
3 V/m from 26 to 1000 MHz, with an 80% amplitude-modulated
signal. The modulation frequency must be within
each functionally significant signal-processing
passband. This often constitutes a severe test,
because many devicessuch as patient monitors with
electrocardiocorder and electrocardiograph optionsare
more susceptible to amplitude modulation that falls
within the range of frequencies used or sensed by
the device for its normal function. For equipment
that does not have a defined passband, the standard
specifies 1-kHz modulation.
This test is usually done in an anechoic chamber.
Special absorbent material (2 m in length) compensates
for standing waves and reflections. The state-of-the-art
standard for radiated immunity, IEC 61000-4-3, requires
a uniform field of 1.5 x 1.5 m at a distance of
3 m, compared with 1 m required by EN 61000-4-3.
The 3-m distance provides a more reliable test result.
IEC 801-4:1984Burst/EFT. During testing,
the bursts or EFTs should be coupled by using a
coupling-decoupling network or by using a coupling
clamp on the EUT cables. The coupling-decoupling
network for an ac/dc main supply circuit allows
the test voltage to be applied nonsymmetrically
to the EUT's power-supply input terminals. The capacitive
coupling clamp enables coupling of the fast transients
to the circuit under test without any galvanic connection
to the circuit terminals, cable shielding, or any
part of the EUT. The clamp should be placed on a
ground plane with a minimum area of 1 m2,
and the ground reference plane should extend beyond
the clamp by at least 0.1 m on all sides. The generator
should be connected to the end of the clamp nearest
to the EUT. EN 60601-1-2 requires test levels of
1 kV for plug-connected equipment, 2 kV for permanently
installed equipment, and 0.5 kV for interconnecting
lines longer than 3 m (i.e., lines using coupling
clamps).
The test should last no less than 1 minute for
each coupling point. Both polarities must be tested.
For this test, high-voltage pulses are not the primary
problem for the EUT. Problems can occur because
of the very fast rise time of a single pulse, which
can create a very high electromagnetic field.
IEC 801-5:1984Surge. Although a surge
is not a very fast pulse, it contains considerable
energy. The surge test, according to IEC 801-5,
is the most severe EMC test for medical device power
supplies. The test simulates an overvoltage caused
by switching and lightning transients, with a pulse
coupled via a coupling-decoupling network on the
power line of the EUT. The test voltage is 1 kV
in differential mode with a 2-
coupling resistor, and 2 kV in common mode with
a 12-
coupling resistor. At least five positive and five
negative discharges must be tested at selected points
of the power supply. The pulse must be repeated
at a rate of at least one per minute, and it is
recommended to increase the test levels from 0.51
kV to 2 kV. The selected points should be 0°,
90°, 180°, and 270° of the sine wave.
Because of the high energy level of this pulse,
great care must be taken when performing this test.
It is important to ensure that the generator is
in a safe mode, especially during counter measurement
or modifications.
A revised version of IEC 60601-1-2 is currently
being developed. This new edition is scheduled to
be published this year but could be delayed depending
on the comments received on the second draft and
subsequent voting results. The changes from the
first edition are expected to be extensive, with
requirements updated to reflect the development
of basic standards during the past 5 years. Performance
criteria will change from being safety based to
being based on continued utility of the device.
Test levels will be increased for some tests, and
additional tests will be required. Significant changes
include a provision allowing emission tests to be
based on CISPR/EN standards, such as EN 55013 for
audio/video products, EN 55022 for information-technology
equipment, or EN 55014 for simple motor-driven devices
(dentistry drills, for example). The requirements
for harmonics distortion and voltage fluctuations
according to EN 61000-3-2 and EN 61000-3-3 are also
included in the draft version.
Major changes are also proposed regarding immunity
testing. The failure criteria will be changed from
its safety hazard focus to defined compliance criteria.
The EUT must be fully functional, must be able to
provide the intended clinical benefit, and must
remain safe. The ESD contact-discharge level will
be increased to 6 kV. Radiated immunity will require
sweeps for all operating modes, and the frequency
range will be 802500 MHz. Life-supporting
equipment will require 10 V/m from 860 to 2500 MHz.
To reduce test time, the modulation frequency will
be either 1 kHz or 2 Hz, depending on the type of
equipment. The differences between plug-connected
and permanently connected equipment will be eliminated.
Test levels for the burst test will be 2 kV for
main cables and 1 kV for input/output (I/O) cables,
and patient cables will be excluded. If the EUT
does not have a noise-protection component (a varistor,
for example) to minimize surge, the test with the
highest levels (1 kV differential mode or 2 kV common
mode) will be sufficient. Conducted RF immunity
testing according to EN 61000-4-6 will be added,
with a level of 3 V from 150 kHz to 80 MHz. A magnetic
immunity test with a level of 10 A/m will also be
added, according to EN 61000-4-8. As required in
EN 61000-4-11, voltage dips and interrupts with
nominal line voltages of 0% Vnominal
for 0.5 periods, 40% Vnominal
for 5 periods, and 70% Vnominal
for 25 periods, and a separate test with restore-to-operation
criteria at 0% Vnominal for
5 seconds will be included. Additional disclosure
requirements and warnings will be required for end-user
documents. Information will be included in the "General
Guidance and Rationale" annex of the revision to
help manufacturers and test laboratories understand
the requirements.
EN 60601-1-2:1993 can be difficult to use, and
therefore has caused some confusion over how to
conduct tests properly to meet the standard's requirements.
A key example is correctly testing patient monitors,
which often measure signals in the microvolt to
millivolt range or the nanoampere to microampere
range. Signals from the patient's body interact
with radio frequencies, making it difficult for
the monitor to discriminate signal interference.
Another problem for performing the tests arises
because of the response time for medical devices,
which mandate a long dwell time for each frequency.
The result is a long testing time.
Used to demonstrate compliance with the EMC Directive
as well as with the MDD, EN 60601-1-2 has been in
place for nearly 6 years. It is hoped that the current
revisions will correct lingering problems like those
described above. Innovative products and new technologies
will compel working groups and committees to continue
to raise the bar for susceptibility test standards.